Physical characterization of a scanning photon counting digital mammography system based on Si-strip detectors.

Magnus Aslund1, Björn Cederström, Mats Lundqvist, Mats Danielsson.   

Abstract

The physical performance of a scanning multislit full field digital mammography system was determined using basic image quality parameters. The system employs a direct detection detector comprised of linear silicon strip sensors in an edge-on geometry connected to photon counting electronics. The pixel size is 50 microm and the field of view 24 x 26 cm2. The performance was quantified using the presampled modulation transfer function, the normalized noise power spectrum and the detective quantum efficiency (DQE). Compared to conventional DQE methods, the scanning geometry with its intrinsic scatter rejection poses additional requirements on the measurement setup, which are investigated in this work. The DQE of the photon counting system was found to be independent of the dose level to the detector in the 7.6-206 microGy range. The peak DQE was 72% and 73% in the scan and slit direction, respectively, measured with a 28 kV W-0.5 mm Al anode-filter combination with an added 2 mm Al filtration.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17654894     DOI: 10.1118/1.2731032

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  17 in total

1.  Radiation dose reduction in computed tomography: techniques and future perspective.

Authors:  Lifeng Yu; Xin Liu; Shuai Leng; James M Kofler; Juan C Ramirez-Giraldo; Mingliang Qu; Jodie Christner; Joel G Fletcher; Cynthia H McCollough
Journal:  Imaging Med       Date:  2009-10

2.  Quantification of breast density with spectral mammography based on a scanned multi-slit photon-counting detector: a feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Phys Med Biol       Date:  2012-07-06       Impact factor: 3.609

Review 3.  Vision 20/20: Single photon counting x-ray detectors in medical imaging.

Authors:  Katsuyuki Taguchi; Jan S Iwanczyk
Journal:  Med Phys       Date:  2013-10       Impact factor: 4.071

4.  A comparative analysis of OTF, NPS, and DQE in energy integrating and photon counting digital x-ray detectors.

Authors:  Raymond J Acciavatti; Andrew D A Maidment
Journal:  Med Phys       Date:  2010-12       Impact factor: 4.071

5.  Energy weighting improves dose efficiency in clinical practice: implementation on a spectral photon-counting mammography system.

Authors:  Johan Berglund; Henrik Johansson; Mats Lundqvist; Björn Cederström; Erik Fredenberg
Journal:  J Med Imaging (Bellingham)       Date:  2014-08-28

6.  Photon counting performance of amorphous selenium and its dependence on detector structure.

Authors:  Jann Stavro; Amir H Goldan; Wei Zhao
Journal:  J Med Imaging (Bellingham)       Date:  2018-10-30

7.  Breast density evaluation using spectral mammography, radiologist reader assessment, and segmentation techniques: a retrospective study based on left and right breast comparison.

Authors:  Sabee Molloi; Huanjun Ding; Stephen Feig
Journal:  Acad Radiol       Date:  2015-05-29       Impact factor: 3.173

8.  Ideal-observer detectability in photon-counting differential phase-contrast imaging using a linear-systems approach.

Authors:  Erik Fredenberg; Mats Danielsson; J Webster Stayman; Jeffrey H Siewerdsen; Magnus Aslund
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

Review 9.  Breast cancer imaging: a perspective for the next decade.

Authors:  Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

10.  Quantitative contrast-enhanced spectral mammography based on photon-counting detectors: A feasibility study.

Authors:  Huanjun Ding; Sabee Molloi
Journal:  Med Phys       Date:  2017-06-28       Impact factor: 4.071

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.