Literature DB >> 17522880

Preoperative templating of femoral components on plain X-rays. Rotational evaluation with synthetic X-rays on ORTHODOC.

Takehito Hananouchi1, Nobuhiko Sugano, Nobuo Nakamura, Takashi Nishii, Hidenobu Miki, Mitsuyoshi Yamamura, Hideki Yoshikawa.   

Abstract

INTRODUCTION: Preoperative femoral templating for total hip arthroplasty (THA) has been performed by antero-posterior radiographs of both hips which should be taken with the femur rotated internally to reduce the effect of femoral anteversion. However, there is no criterion to evaluate the optimal rotated radiograph. Here, we quantitatively investigated 50 femora whether the thickness of the lesser trochanter (TLT) was useful when evaluating the femoral rotation, and assessed the effect of prosthetic templating on the inappropriate rotated radiographs.
MATERIALS AND METHODS: We assessed 50 femora of 46 consecutive osteoarthritic patients using "synthetic X-ray" based on computer tomography (CT) images, which can be displayed as virtual plain radiographs with any magnification and any projected direction such as plain radiographs. We made four femoral radiograph groups of different rotation prepared (neutral, 15 degrees, 30 degrees, and 45 degrees external rotation), and measured the TLT. We also templated femoral stem on the different rotated synthetic X-ray of each patient, and investigated the effect of inappropriate rotated radiograph by the measurement of the position and the size of the stem.
RESULTS: Seventy-four percent of the neutral rotational group had less than 5 mm of the TLT. While 30 degrees and 45 degrees external rotation group had few cases with less than 5 mm of the TLT. Compared to the neutral group, smaller stem size was selected in more than 80% cases of 30 degrees and 45 degrees external rotational group, and the stem position in these two groups was more than 5 mm proximally.
CONCLUSION: We suggest that when templating femoral stem, this criterion "the TLT is less than 5 mm" reduces some risks by inappropriate rotated radiograph.

Entities:  

Mesh:

Year:  2007        PMID: 17522880     DOI: 10.1007/s00402-007-0349-0

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  9 in total

1.  Preoperative radiographic assessment of limb-length discrepancy in total hip arthroplasty.

Authors:  Geert Meermans; Ahmad Malik; Johan Witt; Fares Haddad
Journal:  Clin Orthop Relat Res       Date:  2010-09-29       Impact factor: 4.176

2.  Reliability of overcoverage parameters with varying morphologic pincer features: comparison of EOS® and radiography.

Authors:  Shafagh Monazzam; Mandar Agashe; Harish S Hosalkar
Journal:  Clin Orthop Relat Res       Date:  2013-05-09       Impact factor: 4.176

3.  Factors related to disagreement in implant size between preoperative CT-based planning and the actual implants used intraoperatively for total hip arthroplasty.

Authors:  Takeshi Ogawa; Masaki Takao; Takashi Sakai; Nobuhiko Sugano
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-12-13       Impact factor: 2.924

Review 4.  Plain Radiography of the Hip: A Review of Radiographic Techniques and Image Features.

Authors:  Seung-Jae Lim; Yoon-Soo Park
Journal:  Hip Pelvis       Date:  2015-09-30

5.  Radiologic measurement of lesser trochanter and its clinical significance in Chinese.

Authors:  Qi Zhang; Huaijun Liu; Wei Chen; Xicheng Li; Zhaohui Song; Jinshe Pan; Yingze Zhang
Journal:  Skeletal Radiol       Date:  2009-03-10       Impact factor: 2.199

6.  Does femoral rotation influence anteroposterior alpha angle, lateral center-edge angle, and medial proximal femoral angle? A pilot study.

Authors:  Shafagh Monazzam; James D Bomar; Mandar Agashe; Harish S Hosalkar
Journal:  Clin Orthop Relat Res       Date:  2012-11-21       Impact factor: 4.176

7.  Independent restoration of femoral and acetabular height reduces limb length discrepancy and improves reported outcome following total hip arthroplasty.

Authors:  Jonathan Warnock; Janet Hill; Lee Humphreys; Nicola Gallagher; Richard Napier; David Beverland
Journal:  J Orthop       Date:  2019-05-28

8.  CORR Insights®: Anatomical Mapping of the External Obturator Footprint: A Study In Cadavers with Implications for Direct Anterior THA.

Authors:  Geert Meermans
Journal:  Clin Orthop Relat Res       Date:  2021-02-01       Impact factor: 4.755

9.  A new digital preoperative planning method for total hip arthroplasties.

Authors:  Hendrikus J A Crooijmans; Armand M R P Laumen; Carola van Pul; Jan B A van Mourik
Journal:  Clin Orthop Relat Res       Date:  2008-09-10       Impact factor: 4.176

  9 in total