Assessment of secondary patient motion induced by automated couch movement during on-line 6 dimensional repositioning in prostate cancer treatment.

Nadine Linthout1, Dirk Verellen, Koen Tournel, Truus Reynders, Michael Duchateau, Guy Storme.   

Abstract

BACKGROUND AND
PURPOSE: The purpose of this study is to assess retrospectively secondary patient motion induced by 6D patient setup correction.
MATERIALS AND METHODS: For 104 patients, treated with Novalis, 6D setup correction prior to treatment was performed by ExacTrac5.0/NovalisBody in combination with the Robotic Tilt Module mounted underneath the Exact Couch top. This 6D correction might induce additional setup errors due to patient reaction against the rotations. To evaluate induced secondary motion, the 6D setup correction is verified and evaluated with respect to the tolerance limits.
RESULTS: The majority of measured secondary motions are found within the tolerance limits. Detected secondary motions are mostly found in longitudinal shifts and lateral rotations, and mainly found in only 1 dimension during the same verification. The verifications indicate that the patient population can be divided into a group that hardly moves and a group that moves throughout all 6D setup corrections. The patient's behavior can be predicted by the evaluation of the first five fractions as none of the patients demonstrate a learning curve during the treatment.
CONCLUSIONS: 6D setup correction does not induce secondary motion for the majority of the patients and can therefore be applied for all treatment indications.

Entities:  

Mesh:

Year:  2007        PMID: 17499870     DOI: 10.1016/j.radonc.2007.04.015

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  10 in total

1.  Roll and pitch set-up errors during volumetric modulated arc delivery: can adapting gantry and collimator angles compensate?

Authors:  Nienke A Hoffmans-Holtzer; Daan Hoffmans; Max Dahele; Ben J Slotman; Wilko F A R Verbakel
Journal:  Strahlenther Onkol       Date:  2014-11-28       Impact factor: 3.621

2.  Residual translational and rotational errors after kV X-ray image-guided correction of prostate location using implanted fiducials.

Authors:  Reinhold Graf; Dirk Boehmer; Volker Budach; Peter Wust
Journal:  Strahlenther Onkol       Date:  2010-09-30       Impact factor: 3.621

3.  Measurement of patient imaging dose for real-time kilovoltage x-ray intrafraction tumour position monitoring in prostate patients.

Authors:  James K Crocker; Jin Aun Ng; Paul J Keall; Jeremy T Booth
Journal:  Phys Med Biol       Date:  2012-04-20       Impact factor: 3.609

4.  Evaluation of combining bony anatomy and soft tissue position correction strategies for IMRT prostate cancer patients.

Authors:  Marta Adamczyk; Tomasz Piotrowski; Ewa Adamiak
Journal:  Rep Pract Oncol Radiother       Date:  2012-02-09

5.  Analysis of the Setup Uncertainty and Margin of the Daily ExacTrac 6D Image Guide System for Patients with Brain Tumors.

Authors:  Se An Oh; Ji Woon Yea; Min Kyu Kang; Jae Won Park; Sung Kyu Kim
Journal:  PLoS One       Date:  2016-03-28       Impact factor: 3.240

6.  Prostate rotation detected from implanted markers can affect dose coverage and cannot be simply dismissed.

Authors:  Qingyang Shang; Lawrence J Sheplan Olsen; Kevin Stephans; Rahul Tendulkar; Ping Xia
Journal:  J Appl Clin Med Phys       Date:  2013-05-06       Impact factor: 2.102

Review 7.  Adaptive optimization by 6 DOF robotic couch in prostate volumetric IMRT treatment: rototranslational shift and dosimetric consequences.

Authors:  Silvia Chiesa; Lorenzo Placidi; Luigi Azario; Gian Carlo Mattiucci; Francesca Greco; Andrea Damiani; Giovanna Mantini; Vincenzo Frascino; Angelo Piermattei; Vincenzo Valentini; Mario Balducci
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

8.  eNAL++: a new and effective off-line correction protocol for rotational setup errors when using a robotic couch.

Authors:  Daan Martens; Mark Luesink; Henk Huizenga; Kasper L Pasma
Journal:  J Appl Clin Med Phys       Date:  2015-11-08       Impact factor: 2.102

9.  Setup error assessment based on "Sphere-Mask" Optical Positioning System: Results from a multicenter study.

Authors:  Yan Zhang; Han Zhou; Kaiyue Chu; Chuanfeng Wu; Yun Ge; Guoping Shan; Jundong Zhou; Jing Cai; Jianhua Jin; Weiyu Sun; Ying Chen; Xiaolin Huang
Journal:  Front Oncol       Date:  2022-10-04       Impact factor: 5.738

10.  Study of ExacTrac X-ray 6D IGRT setup uncertainty for marker-based prostate IMRT treatment.

Authors:  Chengyu Shi; Adam Tazi; Deborah Xiangdong Fang; Christopher Iannuzzi
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.