| Literature DB >> 19948050 |
Frank Hoffmeyer1, Monika Raulf-Heimsoth, Volker Harth, Jürgen Bünger, Thomas Brüning.
Abstract
BACKGROUND: The collection of exhaled breath condensate (EBC) is a suitable and non-invasive method for evaluation of airway inflammation. Several studies indicate that the composition of the condensate and the recovery of biomarkers are affected by physical characteristics of the condensing device and collecting circumstances. Additionally, there is an apparent influence of the condensing temperature, and often the level of detection of the assay is a limiting factor. The ECoScreen2 device is a new, partly single-use disposable system designed for studying different lung compartments.Entities:
Mesh:
Substances:
Year: 2009 PMID: 19948050 PMCID: PMC2793251 DOI: 10.1186/1471-2466-9-48
Source DB: PubMed Journal: BMC Pulm Med ISSN: 1471-2466 Impact factor: 3.317
Characteristics of the study subjects
| Category | N | Median | Range | |
|---|---|---|---|---|
| Gender | Male | 9 | ||
| Female | 7 | |||
| Age (years) | 39 | 26 - 62 | ||
| Smoking status | Current/Ex-smokers | 0 | ||
| Non-smokers | 16 | |||
| FVC % predicted | 16 | 108.9 | 93.1 - 147.2 | |
| FEV1% predicted | 16 | 102.5 | 88.1 - 139.8 | |
| FEV1/FVC [%] | 16 | 81.7 | 70.1 - 93.9 | |
FVC, forced vital capacity; FEV1, forced expiratory volume in 1 sec.
Results of statistical analysis on repeatability
| EcoScreen | ||||||
|---|---|---|---|---|---|---|
| bias % | 3.5 | 0.2 | +) | 48.9 | 2.1 | +) |
| CV % | 7.7 | 0.9 | 166.3 | 21.4 | ||
| r2 | 0.479 | 0.889 | n.c. | 0.816 | ||
| p | 0.027 | < 0.0001 | n.c. | 0.002 | ||
| ICC | 0.692 | 0.943 | n.c. | 0.903 | ||
| bias % | 0.5 | 0.7 | 9.0 | +) | 3.2 | 11.9 |
| CV % | 8.0 | 1.0 | 18.6 | 11.5 | 34.4 | |
| r2 | 0.618 | 0.915 | 0.656 | 0.626 | 0.704 | |
| p | 0.021 | 0.0002 | 0.015 | 0.017 | 0.018 | |
| ICC | 0.786 | 0.957 | 0.810 | 0.791 | 0.839 | |
Repeated measurements of volume, pH and the mediators in EBC obtained with either the ECoScreen or ECoScreen2 device within 30 minutes. Values of the mean differences of sampling period 1 and 2 (bias), CVs and correlations. CV, coefficient of variation; ICC, intraclass correlation coefficient; n.c., not calculated. +)mediator concentration < LOQ.
Figure 1Comparison of sample volume (A) and EBC-pH (B) obtained with the two devices. Difference between the values (ECoScreen2 -- ECoScreen) plotted against the mean sample volume according to Bland and Altman [20]. The continuous line represents the mean difference and the dashed lines represent the ± SD for the differences.
Figure 2Correlations between pH values obtained with ECoScreen (x--axis) and ECoScreen2 (y-axis). A: Immediate determination; B: determination after argon treatment and -70°C storage.
Concentrations of selected biomarkers in samples collected with ECoScreen and ECoScreen2
| ECoScreen | ECoScreen2 | |
|---|---|---|
| Samples > LOQ/total samples | 8/16 | 0/16 |
| concentration (μM) | 4.3 (3.3 - 7.1) | |
| Samples > LOQ/total samples | 2/16 | 12/16 |
| concentration (pg/mL) | n.c. | 22.8 (8.9 - 36.7) |
| Samples > LOQ/total samples | 0/11 | 13/15 |
| concentration (pg/mL) | 202.7 (55.6 - 356.1) | |
| Samples > LOQ/total samples | 0/8 | 0/9 |
| Samples > LOQ/total samples | 12/12 | 15/15 |
| concentration (pg/mL) | 81.9 (70.2 - 241.5) | 464.4 (223.0 - 661.1) |
| Difference | p < 0.01 | |
NOx, nitrogen oxides; LTB4, leukotriene B4; PGE2, prostaglandin E2; cys-LT, cysteinyl-LT; 8-iso-PGF2α, 8-isoprostane. n.c., not calculated.