Rob H Ireland1, Karen E Dyker, David C Barber, Steven M Wood, Michael B Hanney, Wendy B Tindale, Neil Woodhouse, Nigel Hoggard, John Conway, Martin H Robinson.
Abstract
PURPOSE: Head and neck radiotherapy planning with positron emission tomography/computed tomography (PET/CT) requires the images to be reliably registered with treatment planning CT. Acquiring PET/CT in treatment position is problematic, and in practice for some patients it may be beneficial to use diagnostic PET/CT for radiotherapy planning. Therefore, the aim of this study was first to quantify the image registration accuracy of PET/CT to radiotherapy CT and, second, to assess whether PET/CT acquired in diagnostic position can be registered to planning CT. METHODS AND MATERIALS: Positron emission tomography/CT acquired in diagnostic and treatment position for five patients with head and neck cancer was registered to radiotherapy planning CT using both rigid and nonrigid image registration. The root mean squared error for each method was calculated from a set of anatomic landmarks marked by four independent observers.Entities:
Mesh:
Substances:
Year: 2007 PMID: 17445999 PMCID: PMC2713594 DOI: 10.1016/j.ijrobp.2007.02.017
Source DB: PubMed Journal: Int J Radiat Oncol Biol Phys ISSN: 0360-3016 Impact factor: 7.038
Comparison of diagnostic/staging PET and treatment planning PET
| Diagnostic/ staging PET | Treatment planning PET | |
|---|---|---|
| Identification of nodes | Yes | Yes |
| Timing during patient management | Early | Late |
| Time to assess treatment options | Yes | Limited |
| Custom immobilization | No | Yes |
| Flat bed | No | Yes |
Abbreviation: PET = positron emission tomography.
Fig. 1Image registration stages displayed as red-green fused images in which regions of similar intensity are shown in yellow. (a) Original images. (b) Manual translation. (c) Rigid registration. (d) Nonrigid registration. Red = Treatment planning Computed tomography (CT). Green = Attenuation correction Hawkeye computed tomography (CT) acquired in treatment position. Figure appears in color online.
Summary of registration accuracy results
| Rigid (mm) | Nonrigid (mm) | Paired | |
|---|---|---|---|
| Treatment position | 4.96 ± 2.38 | 2.77 ± 0.80 | |
| Diagnostic position | 5.96 ± 1.05 | 3.20 ± 1.22 |