David Mattes1, David R Haynor, Hubert Vesselle, Thomas K Lewellen, William Eubank.
Abstract
We have implemented and validated an algorithm for three-dimensional positron emission tomography transmission-to-computed tomography registration in the chest, using mutual information as a similarity criterion. Inherent differences in the two imaging protocols produce significant nonrigid motion between the two acquisitions. A rigid body deformation combined with localized cubic B-splines is used to capture this motion. The deformation is defined on a regular grid and is parameterized by potentially several thousand coefficients. Together with a spline-based continuous representation of images and Parzen histogram estimates, our deformation model allows closed-form expressions for the criterion and its gradient. A limited-memory quasi-Newton optimization algorithm is used in a hierarchical multiresolution framework to automatically align the images. To characterize the performance of the method, 27 scans from patients involved in routine lung cancer staging were used in a validation study. The registrations were assessed visually by two expert observers in specific anatomic locations using a split window validation technique. The visually reported errors are in the 0- to 6-mm range and the average computation time is 100 min on a moderate-performance workstation.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12703765 DOI: 10.1109/TMI.2003.809072
Source DB: PubMed Journal: IEEE Trans Med Imaging ISSN: 0278-0062 Impact factor: 10.048