Literature DB >> 17425443

Relationship between interstitial and blood glucose in type 1 diabetes patients: delay and the push-pull phenomenon revisited.

Iris M E Wentholt1, Augustus A M Hart, Joost B L Hoekstra, J Hans Devries.   

Abstract

BACKGROUND: Interpretation of glucose sensor results requires clarification of the relationship between interstitial (IG) and blood (BG) glucose. We examined the delay of IG upon BG change and reinvestigated the push-pull phenomenon in type 1 diabetes patients. The push-pull phenomenon postulates that IG shows a delayed increase but earlier decrease compared to BG. If so, postprandial sensor curves should have narrower peak widths than BG curves.
METHODS: For both sensors a two-point calibration procedure was used. Delay was assessed by shifting combined fitted postprandial glucose sensor curves horizontally. The sensor and BG peak widths of the separately fitted curves were assessed and compared. Peak width was re-assessed for the microdialysis sensor using raw current values to rule out any calibration effect on the shape of the curve. The contribution of instrumental delay to the earlier reported 7.1-min delay of the microdialysis sensor was calculated.
RESULTS: No delay [-2.2 +/- 6.2 (SD) min] was seen for the needle-type sensor. Instrumental delay was >6.2 min for the microdialysis sensor, accounting for more than 87% of the total reported delay of 7.1 +/- 5.5 min. Mean peak width for the BG curves was 100.8 +/- 25.0 min, for the needle-type sensor curves 110.0 +/- 20.5 min, and for the microdialysis sensor curves 104.6 +/- 21.7 min (P = 0.052 and P = 0.11 vs. BG, respectively). Mean peak width for the uncalibrated microdialysis current values was 105.0 +/- 23.1 min, which was not different from the peak width of the BG curves (P = 0.347).
CONCLUSIONS: IG-BG delay may be smaller than previously postulated. The sensor curves tended to have broader peaks than the BG curves, in contrast to the expected narrower peaks predicted by the push-pull phenomenon. This argues against the existence of the push-pull phenomenon.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17425443     DOI: 10.1089/dia.2006.0007

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  26 in total

Review 1.  Technologies for continuous glucose monitoring: current problems and future promises.

Authors:  Santhisagar Vaddiraju; Diane J Burgess; Ioannis Tomazos; Faquir C Jain; Fotios Papadimitrakopoulos
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

2.  Graphical and numerical evaluation of continuous glucose sensing time lag.

Authors:  Boris P Kovatchev; Devin Shields; Marc Breton
Journal:  Diabetes Technol Ther       Date:  2009-03       Impact factor: 6.118

Review 3.  Amperometric glucose sensors: sources of error and potential benefit of redundancy.

Authors:  Jessica R Castle; W Kenneth Ward
Journal:  J Diabetes Sci Technol       Date:  2010-01-01

4.  Combining insulin pumps and continuous glucose monitors; where are we to go from here?

Authors:  J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2008-03

5.  Accuracy and reliability of continuous glucose monitoring systems: a head-to-head comparison.

Authors:  Yoeri M Luijf; Julia K Mader; Werner Doll; Thomas Pieber; Anne Farret; Jerome Place; Eric Renard; Daniela Bruttomesso; Alessio Filippi; Angelo Avogaro; Sabine Arnolds; Carsten Benesch; Lutz Heinemann; J Hans DeVries
Journal:  Diabetes Technol Ther       Date:  2013-05-07       Impact factor: 6.118

Review 6.  Delays in minimally invasive continuous glucose monitoring devices: a review of current technology.

Authors:  D Barry Keenan; John J Mastrototaro; Gayane Voskanyan; Garry M Steil
Journal:  J Diabetes Sci Technol       Date:  2009-09-01

7.  Optimum subcutaneous glucose sampling and fourier analysis of continuous glucose monitors.

Authors:  Marc D Breton; Devin P Shields; Boris P Kovatchev
Journal:  J Diabetes Sci Technol       Date:  2008-05

8.  Analysis of the Accuracy and Performance of a Continuous Glucose Monitoring Sensor Prototype: An In-Silico Study Using the UVA/PADOVA Type 1 Diabetes Simulator.

Authors:  Marc D Breton; Rolf Hinzmann; Enrique Campos-Nañez; Susan Riddle; Michael Schoemaker; Guenther Schmelzeisen-Redeker
Journal:  J Diabetes Sci Technol       Date:  2016-12-13

9.  Sense and nonsense in sensors.

Authors:  J Hermanides; J H DeVries
Journal:  Diabetologia       Date:  2010-01-10       Impact factor: 10.122

10.  Glucose sensing issues for the artificial pancreas.

Authors:  J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2008-07
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.