Literature DB >> 17403078

Adverse interaction between a left ventricular assist device and an implantable cardioverter defibrillator.

Jennifer Cowger Matthews1, Daphne Betley, Fred Morady, Frank Pelosi.   

Abstract

An increasing number of patients have a coexisting implantable cardioverter defibrillator (ICD) and left ventricular assist device (LVAD) to treat ventricular arrhythmias and refractory heart failure, respectively. To date, there have been no published reports of negative interactions between these devices that have impacted appropriate ICD or LVAD function. In this case report, we describe a patient with an LVAD-ICD interaction that necessitated replacement of the implantable defibrillator.

Entities:  

Mesh:

Year:  2007        PMID: 17403078     DOI: 10.1111/j.1540-8167.2007.00805.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  9 in total

1.  Clinical implications of left ventricular assist device implantation in patients with an implantable cardioverter-defibrillator.

Authors:  Fanny Boudghène-Stambouli; Stéphane Boulé; Céline Goéminne; Edward Botcherby; Christelle Marquié; Claude Kouakam; Laurence Guédon-Moreau; Guillaume Schurtz; Pascal de Groote; Nicolas Lamblin; Marie Fertin; Emmanuel Robin; François Brigadeau; Didier Klug; Dominique Lacroix; Jonathan Meurice; Ludivine Wissocque; André Vincentelli; Salem Kacet
Journal:  J Interv Card Electrophysiol       Date:  2013-12-13       Impact factor: 1.900

Review 2.  Electromagnetic interference between implantable cardiac devices and continuous-flow left ventricular assist devices: a review.

Authors:  Jonathan S Gordon; Elizabeth J Maynes; Thomas J O'Malley; Behzad B Pavri; Vakhtang Tchantchaleishvili
Journal:  J Interv Card Electrophysiol       Date:  2021-01-12       Impact factor: 1.900

3.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

4.  "Pseudo" Faraday cage: a solution for telemetry link interaction between a left ventricular assist device and an implantable cardioverter defibrillator.

Authors:  Sony Jacob; Prasad K Cherian; Waqas S Ghumman; Mithilesh K Das
Journal:  J Interv Card Electrophysiol       Date:  2009-07-28       Impact factor: 1.900

5.  Right ventricular dysfunction following continuous flow left ventricular assist device placement in 51 patients: predicators and outcomes.

Authors:  Siyamek Neragi-Miandoab; Daniel Goldstein; Ricardo Bello; Robert Michler; David D'Alessandro
Journal:  J Cardiothorac Surg       Date:  2012-06-27       Impact factor: 1.637

6.  MediGuide-assisted atrial flutter ablation in a patient with a HeartMate II left ventricular assist device.

Authors:  Hien Kiem Nguyen Thanh; Jason Andrade; Katia Dyrda; Bernard Thibault; Paul Khairy; Laurent Macle
Journal:  HeartRhythm Case Rep       Date:  2015-03-30

7.  Inhibition of pacing in a dependent patient with an implantable cardioverter-defibrillator and a left ventricular assist device.

Authors:  Michele Murphy; Timothy Welch; Peter W Shaw; Jamie L W Kennedy; Kenneth C Bilchick
Journal:  HeartRhythm Case Rep       Date:  2016-07-27

8.  Electromagnetic Interference from Left Ventricular Assist Device (LVAD) Inhibiting the Pacing Function of an Implantable Cardioverter-Defibrillator (ICD) Device.

Authors:  Sebhat Erqou; Robert L Kormos; Norman C Wang; Dennis M McNamara; Raveen Bazaz
Journal:  Case Rep Cardiol       Date:  2018-10-03

9.  Influence of a fully magnetically levitated left ventricular assist device on functional interrogation of implantable cardioverter defibrillators.

Authors:  Christoph Schukro; Thomas Schlöglhofer; Cesar Khazen; Michael Röhrich; Günther Laufer; Daniel Zimpfer; Dominik Wiedemann
Journal:  Clin Cardiol       Date:  2019-07-07       Impact factor: 2.882

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.