Sony Jacob1, Prasad K Cherian, Waqas S Ghumman, Mithilesh K Das.
Abstract
Patients implanted with left ventricular assist devices (LVAD) may have implantable cardioverter defibrillators (ICD) implanted for sudden cardiac death prevention. This opens the possibility of device-device communication interactions and thus interferences. We present a case of such interaction that led to ICD communication failure following the activation of an LVAD. In this paper, we describe a practical solution to circumvent the communication interference and review the communication links of ICDs and possible mechanisms of ICD-LVAD interactions.Entities:
Mesh:
Year: 2009 PMID: 19636687 DOI: 10.1007/s10840-009-9415-6
Source DB: PubMed Journal: J Interv Card Electrophysiol ISSN: 1383-875X Impact factor: 1.900