Literature DB >> 28491740

Inhibition of pacing in a dependent patient with an implantable cardioverter-defibrillator and a left ventricular assist device.

Michele Murphy1, Timothy Welch1, Peter W Shaw1, Jamie L W Kennedy1, Kenneth C Bilchick1.   

Abstract

Entities:  

Keywords:  Bradycardia; Cardiac resynchronization therapy; Electromagnetic interference; Implantable cardioverter-defibrillator; Pacemaker; Ventricular assist devices; Ventricular tachycardia

Year:  2016        PMID: 28491740      PMCID: PMC5419973          DOI: 10.1016/j.hrcr.2016.06.008

Source DB:  PubMed          Journal:  HeartRhythm Case Rep        ISSN: 2214-0271


× No keyword cloud information.

Introduction

KEY TEACHING POINTS Interactions between left ventricular assist devices (LVADs) and implantable cardioverter-defibrillators (ICDs) are more likely when the sensing lead in a transvenous system or the device in a subcutaneous system is in close proximity to the LVAD cannula. The likelihood of this electromagnetic interference (EMI) depends on the type of LVAD implanted and can lead to both inappropriate sensing of ventricular tachycardia and inhibition of bradycardia pacing. Typical programming strategies to resolve inappropriate ICD sensing from LVAD-associated EMI, such as adjusting sensitivities, refractory periods, decay delay parameters, or sensing filters interaction, may not be adequate to prevent an invasive device revision procedure. When these usual programming strategies are not effective, a novel algorithm that allows setting separate sensitivities for tachycardia detection and bradycardia pacing can resolve these interactions and prevent the need for additional invasive procedures. Advanced heart failure is the most frequent diagnosis in adult patients admitted to the hospital. The prevalence of heart failure has increased as options for treatment have improved. In addition to pharmacologic therapies, standard implantable cardioverter-defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-D) have decreased mortality and improved quality of life. Left ventricular assist devices (LVAD) are also effective therapeutic devices for patients with advanced heart failure, and concurrent therapy with an LVAD and CRT-D is not uncommon. Inappropriate ICD therapy due to electromagnetic interference (EMI) has been reported with LVADs, and ICDLVAD interactions sometimes require lead repositioning or even device removal. Fortunately, some current ICDs have novel detection and pacing algorithms that can prevent the need for invasive therapy to treat these interactions. The present report adds significantly to the literature on LVADs and cardiac implantable electronic devices by demonstrating how certain ICDs may be programmed to preserve appropriate pacing for bradycardia without compromising the ability of the device to detect and treat ventricular tachycardia and ventricular fibrillation.

Case presentation

A 73-year-old man with ischemic cardiomyopathy, longstanding persistent atrial fibrillation, third-degree atrioventricular block, and advanced heart failure underwent LVAD implantation (HeartMate II; Thoratec, Pleasanton, California) as destination therapy. The patient had undergone CRT-D implantation (St. Jude Medical, St. Paul, Minnesota) a year prior to the LVAD placement. Following the LVAD implant, he had been active at home for more than 2 years. He then presented for hospital admission with progressive low back pain and lower-extremity weakness and was diagnosed with lumbar vertebral compression. Telemetry monitoring demonstrated multiple 2-second pauses despite having the CRT-D programmed DDD with a heart rate of 90 beats per minute (Figure 1A). Evaluation of the CRT-D device (St. Jude Unify Quadra 3249-40) demonstrated atrial fibrillation and normal lead parameters during initial testing. A review of the device electrograms during biventricular pacing and a ventricular sensing amplitude test showed that ventricular pacing was inhibited because of periodic right ventricular (RV) lead sensing of low-amplitude electrograms (Figure 1B). The RV lead impedance and capture threshold were normal. Decreasing the RV pacing sensitivity (increasing the threshold value to a higher number) was not effective in preventing these sensed events because the sensitivity threshold would decay significantly during the time intervals between the previous paced event and low-amplitude signal. What is the most likely cause of these low-amplitude signals? How should the device be reprogrammed to avoid inappropriate inhibition of pacing while preserving appropriate sensing of ventricular fibrillation? Is RV lead repositioning necessary, or is there an appropriate programming change that could fix the problem?
Figure 1

Telemetry and initial interrogation. A: The telemetry strip shows frequent pauses and pacing inhibition. B: Device electrograms before and during a ventricular sensing amplitude test include low-amplitude signals on the ventricular channel that inhibit pacing and are inappropriately sensed as intrinsic beats during the sensing test. The magnified box shows a far-field ventricular electrogram of amplitude 0.7 mV that is inappropriately sensed after a ventricular paced beat. The atrial electrograms (not included) showed atrial fibrillation.

Commentary

Diagnosis

The differential diagnosis in this case includes appropriate sensing or inappropriate sensing from one of the following: myopotentials, ambient EMI, a lead fracture, a loose set screw, mechanical interference from the LVAD, or EMI from the LVAD. A lead fracture was unlikely with the normal impedance, and isometric measures in the pocket did not reproduce the interference. The discrete potentials were also atypical for myopotentials. The 2-dimensional and 3-dimensional computed tomography images (Figure 2) show very close proximity of the LVAD cannula to the RV pacing/defibrillating lead, making it most likely that the signals were caused by EMI from the LVAD or perhaps a mechanical interaction. Furthermore, evaluation of the patient’s HeartMate II LVAD showed normal flows and power. Of note, HeartMate II and HeartWare VAD (HVAD; Thoratec, Pleasanton, CA) are the most commonly used LVADs in clinical practice. Although both generate electromagnetic fields, they operate at different pump speeds: 2400–3200 rpm for the HVAD and 8000–12,000 rpm for the HeartMate II. Interestingly, there is not necessarily a proportional relationship between LVAD pump speeds and EMI, as our institutional experience and that of others has been that the HeartMate II less frequently causes EMI compared with the HVAD, even though it uses a higher pump speed, as discussed below in the section “LVAD interactions with other types of ICDs.”
Figure 2

Three-dimensional (3D) imaging of the implantable cardioverter-defibrillator leads and left ventricular assist device. The 3D reconstruction of the chest computed tomography (CT) (A) and the corresponding 2-dimensional CT image in the axial plane (B) demonstrate the close proximity of the right ventricular lead tip to left ventricular apical cannula.

ICD–LVAD interactions

Previously identified interactions between ICD and LVADs have included EMI from oversensing of extracardiac signals, which can result in inappropriate therapies. Lead revisions and generator replacements frequently have been required to resolve this problem. As a result, it is recommended that patients have their ICDs checked at the time of LVAD surgery to evaluate EMI. An ICDLVAD interaction, less frequently seen now, would result in the inability to communicate with older models of St. Jude ICDs.5, 6 These devices used an 8-kHz telemetry frequency, which is close to that of the HeartMate II pulse width modulator. Metal shielding was used to bypass the pulse width modulator of the LVAD and circumvent this interaction, but generator replacement was often required. Of note, the pulse width modulator of the HeartMate II LVAD is used to reduce the 12-volt battery voltage needed to power the motor. More recent ICD models from St. Jude, as well as those made by Biotronik, Medtronic, and Boston Scientific, use a greater telemetry frequency, which makes this type of interaction much less likely. In our patient, with a newer St. Jude ICD model, telemetry was intact but EMI from the LVAD appeared to be the most likely cause. Initial programming attempts to fix the problem included lengthening the postventricular atrial refractory period, increasing the postventricular atrial blanking period, and reducing the sensitivity, but these interventions were not effective. Because the sensed deflection appeared much later than the prior ventricular event, the sensitivity would decay much sooner and still sense the low-amplitude signal. The other concern about reducing the RV sensitivity was increasing the probability of undersensing of ventricular tachycardia or fibrillation and withholding appropriate therapy for these arrhythmias. Adjusting the decay delay parameter, which determines the amount of time after a sensed or paced event that the sensing threshold remains at the programmed value prior to decreasing, did not resolve the problem because the oversensed electrograms occurred well after the last ventricular event. The high attenuation filter was also considered, but this would have required defibrillation threshold testing and potentially inhibited appropriate sensing of ventricular tachyarrhythmias.

Resolution of EMI interaction in the present case

Considering that standard programming changes did not resolve the problem, it is fortunate that this patient’s St. Jude ICD had the SenseAbility algorithm, as it was this algorithm that ultimately resolved the problem. Because many electrophysiologists may not be familiar with this algorithm, we highlight here why it was effective when other measures failed. There are several features of this algorithm that help prevent oversensing of low-amplitude signals. Specifically, this algorithm permits decoupling of ventricular pacing sensitivity and ventricular defibrillator sensitivity. This approach facilitates adjustment of the sensitivity of RV pacing to minimize oversensing without compromising the sensitivity of tachyarrhythmia detection. The device had been programmed at the time of implantation to a nominal ventricular pacing and defibrillation sensitivity of 0.5 mV. To address the inappropriate pacing inhibition during the present admission, the sensitivity for RV pacing was reprogrammed to 2 mV, but the defibrillation sensitivity was maintained at 0.5 mV (Figure 3A), thus allowing for both maintenance of pacing for bradycardia and appropriate detection of ventricular tachyarrhythmias. In other words, the algorithm allows different sensitivities to be programmed for bradycardia pacing and tachyarrhythmia detection. The repeat device interrogation with these new settings revealed similar low-amplitude electrograms on the RV lead; however, pacing immediately followed sensing of these far-field events without inhibition by these signals (Figure 3B). This example of appropriate ventricular pacing following detection of far-field electrograms contrasts with more common algorithms to prevent oversensing that involve blanking of far-field electrograms or marking them as refractory. In summary, this algorithm likely prevented an ICD revision procedure in this patient, as none of the other programming solutions were able to resolve the problem.
Figure 3

Resolution with advanced device programming. A: The algorithm decouples the right ventricular pacing and defibrillation sensitivities. B: The ventricular paced events occur shortly after inappropriately sensed events, which are ignored only for the purposes of pacing.

LVAD interactions with other types of ICDs

With respect to available options with other device manufacturers, Medtronic transvenous ICDs have an RV lead noise discrimination feature, which compares the far-field electrogram signals to the near-field electrograms in order to differentiate RV lead noise from true ventricular arrhythmias. If lead noise is identified when these signals are compared, ventricular arrhythmia detection and therapy are withheld. Other features used for discrimination in Medtronic devices include PR logic, wavelet, T-wave discrimination, onset, and stability, which may have a role in discriminating low-amplitude EMI. In patients with Boston Scientific subcutaneous ICDs (S-ICDs), there have been reported interactions with the HVAD, as the frequency of the signals generated by the electromagnetic field of the HVAD is within the range of the signals detectable by the S-ICD. The relationship between pump speeds and EMI is controversial, as a recent case report suggested that the HeartMate II is actually safe in patients with S-ICDs despite high pump speeds, with no interference detected in both the vectors involving the pulse generator and the alternative vector using the distal lead electrode and proximal ring. However, in another case involving an HVAD, this alternative S-ICD was reported to be associated with less EMI compared with the vectors using the generator (in closer proximity to the LVAD cannula). As a result, in some patients with HVADs, inappropriate therapies may be avoided based on using this alternative S-ICD vector.

Conclusion

In the care of end-stage patients with concurrent LVADs and ICDs, a heightened degree of awareness of potential complications and interactions is needed. In this regard, the present case highlights the complexity of LVADICD interactions and how resolution of a particular LVADICD interaction was achieved using a novel programming algorithm, which helped avoid an invasive LVAD or ICD revision procedure. Specifically, this novel algorithm allowed different sensitivities for bradycardia pacing and ventricular tachyarrhythmia detection, which avoided inhibition of pacing from inappropriately sensed events on the ventricular channel. Application of the SenseAbility algorithm to patients with ICDs, LVADs, and EMI has not previously been reported, and familiarity with this algorithm could help other arrhythmia specialists treat their patients more effectively in the future. Minimizing pacing inhibition in these patients, particularly those who are pacemaker dependent, is of great clinical importance. As cardiac devices continue to evolve, it is critical for electrophysiology providers to understand interactions between cardiac implantable electronic devices and advanced surgical devices for end-stage heart failure, as well as how these problems may be most appropriately addressed.

KEY TEACHING POINTS

Interactions between left ventricular assist devices (LVADs) and implantable cardioverter-defibrillators (ICDs) are more likely when the sensing lead in a transvenous system or the device in a subcutaneous system is in close proximity to the LVAD cannula.

The likelihood of this electromagnetic interference (EMI) depends on the type of LVAD implanted and can lead to both inappropriate sensing of ventricular tachycardia and inhibition of bradycardia pacing.

Typical programming strategies to resolve inappropriate ICD sensing from LVAD-associated EMI, such as adjusting sensitivities, refractory periods, decay delay parameters, or sensing filters interaction, may not be adequate to prevent an invasive device revision procedure.

When these usual programming strategies are not effective, a novel algorithm that allows setting separate sensitivities for tachycardia detection and bradycardia pacing can resolve these interactions and prevent the need for additional invasive procedures.

  8 in total

1.  Inappropriate implantable cardioverter-defibrillator therapy due to electromagnetic interference in patient with a HeartWare HVAD left ventricular assist device.

Authors:  Abram Mozes; David DeNofrio; Duc Thinh Pham; Munther K Homoud
Journal:  Heart Rhythm       Date:  2010-11-16       Impact factor: 6.343

2.  Subcutaneous implantable cardioverter-defibrillator implantation in a patient with a left ventricular assist device already in place.

Authors:  Ashwani Gupta; Faiz Subzposh; Shelley R Hankins; Steven P Kutalek
Journal:  Tex Heart Inst J       Date:  2015-04-01

3.  Left ventricular assist device in a patient with a concomitant subcutaneous implantable cardioverter defibrillator.

Authors:  Diyar Saeed; Alexander Albert; Ralf Westenfeld; Bujar Maxhera; Hildegard Gramsch-Zabel; Stephen O'Connor; Artur Lichtenberg; Joachim Winter
Journal:  Circ Arrhythm Electrophysiol       Date:  2013-06

4.  Electromagnetic interference of automatic implantable cardioverter defibrillator and HeartWare left ventricular assist device.

Authors:  Mohamed R Labedi; Rami Alharethi; A G Kfoury; Deborah Budge; Reid Bruce; Brad Rasmusson; T Jared Bunch
Journal:  ASAIO J       Date:  2013 Mar-Apr       Impact factor: 2.872

5.  Adverse interaction between a left ventricular assist device and an implantable cardioverter defibrillator.

Authors:  Jennifer Cowger Matthews; Daphne Betley; Fred Morady; Frank Pelosi
Journal:  J Cardiovasc Electrophysiol       Date:  2007-03-31

6.  Advanced heart failure treated with continuous-flow left ventricular assist device.

Authors:  Mark S Slaughter; Joseph G Rogers; Carmelo A Milano; Stuart D Russell; John V Conte; David Feldman; Benjamin Sun; Antone J Tatooles; Reynolds M Delgado; James W Long; Thomas C Wozniak; Waqas Ghumman; David J Farrar; O Howard Frazier
Journal:  N Engl J Med       Date:  2009-11-17       Impact factor: 91.245

7.  Implantable defibrillator with left ventricular assist device compatibility.

Authors:  Hanno Oswald; Gunnar Klein; Martin Strüber; Ajmal Gardiwal
Journal:  Interact Cardiovasc Thorac Surg       Date:  2009-02-17

8.  Simultaneous use of implantable cardioverter-defibrillators and left ventricular assist devices in patients with severe heart failure.

Authors:  Michael Kühne; Michaela Sakumura; Stephen Scott Reich; Jean-Francois Sarrazin; Darryl Wells; Nagib Chalfoun; Thomas Crawford; Warangkna Boonyapisit; Laura Horwood; Aman Chugh; Eric Good; Krit Jongnarangsin; Frank Bogun; Hakan Oral; Fred Morady; Francis Pagani; Frank Pelosi
Journal:  Am J Cardiol       Date:  2009-12-22       Impact factor: 2.778

  8 in total
  1 in total

Review 1.  Electromagnetic interference between implantable cardiac devices and continuous-flow left ventricular assist devices: a review.

Authors:  Jonathan S Gordon; Elizabeth J Maynes; Thomas J O'Malley; Behzad B Pavri; Vakhtang Tchantchaleishvili
Journal:  J Interv Card Electrophysiol       Date:  2021-01-12       Impact factor: 1.900

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.