| Literature DB >> 28491740 |
Michele Murphy1, Timothy Welch1, Peter W Shaw1, Jamie L W Kennedy1, Kenneth C Bilchick1.
Abstract
Entities:
Keywords: Bradycardia; Cardiac resynchronization therapy; Electromagnetic interference; Implantable cardioverter-defibrillator; Pacemaker; Ventricular assist devices; Ventricular tachycardia
Year: 2016 PMID: 28491740 PMCID: PMC5419973 DOI: 10.1016/j.hrcr.2016.06.008
Source DB: PubMed Journal: HeartRhythm Case Rep ISSN: 2214-0271
Figure 1Telemetry and initial interrogation. A: The telemetry strip shows frequent pauses and pacing inhibition. B: Device electrograms before and during a ventricular sensing amplitude test include low-amplitude signals on the ventricular channel that inhibit pacing and are inappropriately sensed as intrinsic beats during the sensing test. The magnified box shows a far-field ventricular electrogram of amplitude 0.7 mV that is inappropriately sensed after a ventricular paced beat. The atrial electrograms (not included) showed atrial fibrillation.
Figure 2Three-dimensional (3D) imaging of the implantable cardioverter-defibrillator leads and left ventricular assist device. The 3D reconstruction of the chest computed tomography (CT) (A) and the corresponding 2-dimensional CT image in the axial plane (B) demonstrate the close proximity of the right ventricular lead tip to left ventricular apical cannula.
Figure 3Resolution with advanced device programming. A: The algorithm decouples the right ventricular pacing and defibrillation sensitivities. B: The ventricular paced events occur shortly after inappropriately sensed events, which are ignored only for the purposes of pacing.
KEY TEACHING POINTS
Interactions between left ventricular assist devices (LVADs) and implantable cardioverter-defibrillators (ICDs) are more likely when the sensing lead in a transvenous system or the device in a subcutaneous system is in close proximity to the LVAD cannula. The likelihood of this electromagnetic interference (EMI) depends on the type of LVAD implanted and can lead to both inappropriate sensing of ventricular tachycardia and inhibition of bradycardia pacing. Typical programming strategies to resolve inappropriate ICD sensing from LVAD-associated EMI, such as adjusting sensitivities, refractory periods, decay delay parameters, or sensing filters interaction, may not be adequate to prevent an invasive device revision procedure. When these usual programming strategies are not effective, a novel algorithm that allows setting separate sensitivities for tachycardia detection and bradycardia pacing can resolve these interactions and prevent the need for additional invasive procedures. |