Literature DB >> 17370126

Accuracy of a novel approach to measuring arterial thermodilution cardiac output during intra-aortic counterpulsation.

Werner Baulig1, Philipp Schuett, Oliver Goedje, Edith R Schmid.   

Abstract

OBJECTIVE: To assess the agreement between a novel approach of arterial and the pulmonary artery bolus thermodilution for measuring cardiac output in critically ill patients during aortic counterpulsation.
METHODS: Eighteen male patients aged 37-80 years, undergoing preoperative insertion of an intra-aortic balloon pump (IABP) and elective coronary artery bypass grafting. A thin 1.3FG thermistor was introduced through the pressure lumen to the tip of an 8FG IABP catheter, and the pump rate was set at 1:1. After arrival in the intensive care unit cardiac output (CO) was measured under haemodynamic steady-state conditions hourly for 8-11 h, and arterial bolus thermodilution (BCO(iabp)) and pulmonary artery bolus thermodilution (BCO(pulm)) were determined after the patients' admission to the intensive care unit.
RESULTS: A total of 198 data pairs were obtained: 177 with aortic counterpulsation and 21 without. During aortic counterpulsation, median CO was 6.8 l/min for BCO(iabp) and 6.1 l/min for BCO(pulm), without aortic counterpulsation; corresponding values were 7.1 l/min for BCO(iabp) and 6.5 l/min for BCO(pulm) with aortic counterpulsation. Mean bias was +0.77 l/min, limits of agreement ( +/- 2 SD) were -1.27/+2.81 l/min, and mean error (2 SD/[(BCO(iabp )+ BCO(pulm))/2] was 31.4%. Without aortic counterpulsation, corresponding values were +0.43 l/min, -1.03/+1.87 l/min, and 22.4%.
CONCLUSIONS: Agreement between BCO(iabp) and BCO(pulm) was satisfactory for CO values between 2.0 and 10 l/min only without aortic counterpulsation. BCO(iabp) CO measurements during aortic counterpulsation after coronary artery bypass grafting cannot be recommended at the present time.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 17370126     DOI: 10.1007/s10877-007-9068-x

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   1.977


  32 in total

1.  Ultrastructural changes in the lung alveoli after cardiac surgical operations with the use of cardiopulmonary bypass (CPB).

Authors:  M Wasowicz; P Sobczyński; W Biczysko; R Szulc
Journal:  Pol J Pathol       Date:  1999       Impact factor: 1.072

Review 2.  Pulmonary artery blood temperature and the measurement of cardiac output by thermodilution.

Authors:  S F Moise; C J Sinclair; D H T Scott
Journal:  Anaesthesia       Date:  2002-06       Impact factor: 6.955

3.  Comparison between cardiac output measured by the pulmonary arterial thermodilution technique and that measured by the femoral arterial thermodilution technique in a pediatric animal model.

Authors:  M Rupérez; J López-Herce; C García; C Sánchez; E García; D Vigil
Journal:  Pediatr Cardiol       Date:  2003-12-23       Impact factor: 1.655

4.  A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients.

Authors:  James Dean Sandham; Russell Douglas Hull; Rollin Frederick Brant; Linda Knox; Graham Frederick Pineo; Christopher J Doig; Denny P Laporta; Sidney Viner; Louise Passerini; Hugh Devitt; Ann Kirby; Michael Jacka
Journal:  N Engl J Med       Date:  2003-01-02       Impact factor: 91.245

5.  Continuous, less invasive, hemodynamic monitoring in intensive care after cardiac surgery.

Authors:  O Gödje; K Höke; P Lamm; C Schmitz; C Thiel; M Weinert; B Reichart
Journal:  Thorac Cardiovasc Surg       Date:  1998-08       Impact factor: 1.827

6.  [Cardiac output determination with transpulmonary thermodilution. An alternative to pulmonary catheterization?].

Authors:  T von Spiegel; G Wietasch; J Bürsch; A Hoeft
Journal:  Anaesthesist       Date:  1996-11       Impact factor: 1.041

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  The slowing of sinus rhythm during thermodilution cardiac output determination and the effect of altering injectate temperature.

Authors:  A P Harris; C F Miller; C Beattie; G I Rosenfeld; M C Rogers
Journal:  Anesthesiology       Date:  1985-11       Impact factor: 7.892

9.  Beat-to-beat measurement of cardiac output by intravascular pulse contour analysis: a prospective criterion standard study in patients after cardiac surgery.

Authors:  C Zöllner; M Haller; M Weis; K Mörstedt; P Lamm; E Kilger; A E Goetz
Journal:  J Cardiothorac Vasc Anesth       Date:  2000-04       Impact factor: 2.628

10.  Is the placement of a pulmonary artery catheter still justified solely for the measurement of cardiac output?

Authors:  S G Sakka; K Reinhart; K Wegscheider; A Meier-Hellmann
Journal:  J Cardiothorac Vasc Anesth       Date:  2000-04       Impact factor: 2.628

View more
  1 in total

1.  Accuracy and precision of transcardiopulmonary thermodilution in patients with cardiogenic shock.

Authors:  Bonaventura Schmid; Katrin Fink; Manfred Olschewski; Stephan Richter; Tilmann Schwab; Michael Brunner; Hans-Joerg Busch
Journal:  J Clin Monit Comput       Date:  2015-10-01       Impact factor: 2.502

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.