Abstract

The design of high-resolution retinal prostheses presents many unique engineering and biological challenges. Ever smaller electrodes must inject enough charge to stimulate nerve cells, within electrochemically safe voltage limits. Stimulation sites should be placed within an electrode diameter from the target cells to prevent 'blurring' and minimize current. Signals must be delivered wirelessly from an external source to a large number of electrodes, and visual information should, ideally, maintain its natural link to eye movements. Finally, a good system must have a wide range of stimulation currents, external control of image processing and the option of either anodic-first or cathodic-first pulses. This paper discusses these challenges and presents solutions to them for a system based on a photodiode array implant. Video frames are processed and imaged onto the retinal implant by a head-mounted near-to-eye projection system operating at near-infrared wavelengths. Photodiodes convert light into pulsed electric current, with charge injection maximized by applying a common biphasic bias waveform. The resulting prosthesis will provide stimulation with a frame rate of up to 50 Hz in a central 10 degrees visual field, with a full 30 degrees field accessible via eye movements. Pixel sizes are scalable from 100 to 25 microm, corresponding to 640-10,000 pixels on an implant 3 mm in diameter.

Mesh:

Year:  2007        PMID: 17325419     DOI: 10.1088/1741-2560/4/1/S09

Source DB:  PubMed          Journal:  J Neural Eng        ISSN: 1741-2552            Impact factor:   5.379


  32 in total

1.  Strength-duration relationship for extracellular neural stimulation: numerical and analytical models.

Authors:  David Boinagrov; Jim Loudin; Daniel Palanker
Journal:  J Neurophysiol       Date:  2010-08-11       Impact factor: 2.714

2.  Contrast Sensitivity With a Subretinal Prosthesis and Implications for Efficient Delivery of Visual Information.

Authors:  Georges Goetz; Richard Smith; Xin Lei; Ludwig Galambos; Theodore Kamins; Keith Mathieson; Alexander Sher; Daniel Palanker
Journal:  Invest Ophthalmol Vis Sci       Date:  2015-11       Impact factor: 4.799

3.  Factors affecting perceptual thresholds in epiretinal prostheses.

Authors:  Chloé de Balthasar; Sweta Patel; Arup Roy; Ricardo Freda; Scott Greenwald; Alan Horsager; Manjunatha Mahadevappa; Douglas Yanai; Matthew J McMahon; Mark S Humayun; Robert J Greenberg; James D Weiland; Ione Fine
Journal:  Invest Ophthalmol Vis Sci       Date:  2008-06       Impact factor: 4.799

Review 4.  Polymer optoelectronic structures for retinal prosthesis.

Authors:  Vini Gautam; K S Narayan
Journal:  Organogenesis       Date:  2014-02-24       Impact factor: 2.500

5.  Integration of Perforated Subretinal Prostheses With Retinal Tissue.

Authors:  Adewumi N Adekunle; Alice Adkins; Wei Wang; Henry J Kaplan; Juan Fernandez de Castro; Sang Joon Lee; Philip Huie; Daniel Palanker; Maureen McCall; Machelle T Pardue
Journal:  Transl Vis Sci Technol       Date:  2015-08-14       Impact factor: 3.283

Review 6.  Wireless microstimulators for neural prosthetics.

Authors:  Mesut Sahin; Victor Pikov
Journal:  Crit Rev Biomed Eng       Date:  2011

7.  Selectivity of direct and network-mediated stimulation of the retinal ganglion cells with epi-, sub- and intraretinal electrodes.

Authors:  David Boinagrov; Susanne Pangratz-Fuehrer; Georges Goetz; Daniel Palanker
Journal:  J Neural Eng       Date:  2014-03-10       Impact factor: 5.379

8.  Vertically integrated photo junction-field-effect transistor pixels for retinal prosthesis.

Authors:  Samir Damle; Yu-Hsin Liu; Shaurya Arya; Nicholas W Oesch; Yu-Hwa Lo
Journal:  Biomed Opt Express       Date:  2019-12-04       Impact factor: 3.732

9.  Photovoltaic Restoration of Central Vision in Atrophic Age-Related Macular Degeneration.

Authors:  Daniel Palanker; Yannick Le Mer; Saddek Mohand-Said; Mahiul Muqit; Jose A Sahel
Journal:  Ophthalmology       Date:  2020-02-25       Impact factor: 12.079

10.  Toward a wide-field retinal prosthesis.

Authors:  Hossein Ameri; Tanapat Ratanapakorn; Stefan Ufer; Helmut Eckhardt; Mark S Humayun; James D Weiland
Journal:  J Neural Eng       Date:  2009-05-20       Impact factor: 5.379