Literature DB >> 18559137

In situ fracture of stents implanted for relief of pulmonary arterial stenosis in patients with congenitally malformed hearts.

Doff B McElhinney1, Lisa Bergersen, Audrey C Marshall.   

Abstract

BACKGROUND: One of the most common uses of stents in patients with congenitally malformed hearts is treatment of pulmonary arterial stenosis. Although there are reports of fractured pulmonary arterial stents, little is known about the risk factors for, and implications of, such fractures.
METHODS: We reviewed angiograms to identify fractures in stents previously inserted to relieve stenoses in pulmonary arteries from 1990 through 2001 in patients who also underwent follow-up catheterization at least 3 years after placement of the stent. We undertook matched cohort analysis, matching a ratio of 2 fractured to 1 unfractured stent.
RESULTS: Overall, 166 stents meeting the criterions of our study had been placed in 120 patients. We identified fractures in 35 stents (21%) in 29 patients. All fractured stents were in the central pulmonary arteries, 24 (69%) in the central part of the right pulmonary artery, and all were complete axial fractures, or complex fractures along at least 2 planes. Stent-related factors associated with increased risk of fracture identified by multivariable logistic regression included placement in close apposition to the ascending aorta (p = 0.001), and a larger expanded diameter (p = 0.002). There was obstruction across 28 of 35 fractured stents, which was severe in 11. We re-stented 21 of the fractured stents, and recurrent fracture was later diagnosed in 3 of these. A fragment of the fractured stent embolized distally in 2 patients, without clinically important effects.
CONCLUSIONS: In situ fracture of pulmonary arterial stents is relatively common, and in most cases is related to compression by the aorta. There is usually recurrent obstruction across the fractured stent, but fractured stents rarely embolize, and are not associated with other significant complications.

Entities:  

Mesh:

Year:  2008        PMID: 18559137      PMCID: PMC4235280          DOI: 10.1017/S1047951108002424

Source DB:  PubMed          Journal:  Cardiol Young        ISSN: 1047-9511            Impact factor:   1.093


  35 in total

1.  The use of balloon-expandable stents in the management of intracranial arterial diseases: a 5-year single-center experience.

Authors:  Iruena Moraes Kessler; Charbel Mounayer; Michel Piotin; Laurent Spelle; Jose Ricardo Vanzin; Jacques Moret
Journal:  AJNR Am J Neuroradiol       Date:  2005-10       Impact factor: 3.825

2.  Fracture of a renal artery stent due to mobile kidney.

Authors:  Sinan Sahin; Ahmet Memiş; Mustafa Parildar; Ismail Oran
Journal:  Cardiovasc Intervent Radiol       Date:  2005 Sep-Oct       Impact factor: 2.740

3.  Two cases of fractured esophageal nitinol stents.

Authors:  R Schoefl; F Winkelbauer; M Haefner; R Poetzi; A Gangl; J Lammer
Journal:  Endoscopy       Date:  1996-08       Impact factor: 10.093

4.  Strength, elasticity, and plasticity of expandable metal stents: in vitro studies with three types of stress.

Authors:  F Flueckiger; H Sternthal; G E Klein; M Aschauer; D Szolar; G Kleinhappl
Journal:  J Vasc Interv Radiol       Date:  1994 Sep-Oct       Impact factor: 3.464

5.  Use of endovascular stents in congenital heart disease.

Authors:  M P O'Laughlin; S B Perry; J E Lock; C E Mullins
Journal:  Circulation       Date:  1991-06       Impact factor: 29.690

6.  Fracture of a carotid stent: an unexpected complication.

Authors:  Arif Rahim Valibhoy; Bibombe Patrice Mwipatayi; Kishore Sieunarine
Journal:  J Vasc Surg       Date:  2007-03       Impact factor: 4.268

7.  Risk stratification, systematic classification, and anticipatory management strategies for stent fracture after percutaneous pulmonary valve implantation.

Authors:  Johannes Nordmeyer; Sachin Khambadkone; Louise Coats; Silvia Schievano; Philipp Lurz; Giovanni Parenzan; Andrew M Taylor; James E Lock; Philipp Bonhoeffer
Journal:  Circulation       Date:  2007-03-05       Impact factor: 29.690

8.  Implantation and intermediate-term follow-up of stents in congenital heart disease.

Authors:  M P O'Laughlin; M C Slack; R G Grifka; S B Perry; J E Lock; C E Mullins
Journal:  Circulation       Date:  1993-08       Impact factor: 29.690

9.  The Genesis stent: A new low-profile stent for use in infants, children, and adults with congenital heart disease.

Authors:  Thomas J Forbes; Edwin Rodriguez-Cruz; Zahid Amin; Lee N Benson; Thomas E Fagan; William E Hellenbrand; Larry A Latson; Phillip Moore; Charles E Mullins; Julie A Vincent
Journal:  Catheter Cardiovasc Interv       Date:  2003-07       Impact factor: 2.692

10.  Longitudinal stent fracture 11 months after implantation in the left pulmonary artery and successful management by a stent-in-stent maneuver.

Authors:  Walter Knirsch; Nikolaus A Haas; Martin A G Lewin; Frank Uhlemann
Journal:  Catheter Cardiovasc Interv       Date:  2003-01       Impact factor: 2.692

View more
  5 in total

Review 1.  Transcatheter Pulmonary Valve Replacement: Current State of Art.

Authors:  Wail Alkashkari; Amani Alsubei; Ziyad M Hijazi
Journal:  Curr Cardiol Rep       Date:  2018-03-15       Impact factor: 2.931

2.  Postmarketing trials and pediatric device approvals.

Authors:  Thomas J Hwang; Aaron S Kesselheim; Florence T Bourgeois
Journal:  Pediatrics       Date:  2014-04-14       Impact factor: 7.124

3.  Clinical and stent-related outcomes after transcatheter or operative placement of bare-metal stents in the ventricular septum or subvalvar systemic outflow tract.

Authors:  Diego Porras; Doff B McElhinney; Pedro Del Nido; James E Lock; Jeffrey Meadows; Audrey C Marshall
Journal:  Circ Cardiovasc Interv       Date:  2012-07-31       Impact factor: 6.546

4.  Interventions after Arterial Switch: A Single Low Case-Volume Center Experience.

Authors:  Karolis Jonas; Virginijus Jakutis; Rita Sudikienė; Virgilijus Lebetkevičius; Virgilijus Tarutis
Journal:  Medicina (Kaunas)       Date:  2021-04-21       Impact factor: 2.430

5.  Comparison of pulmonary artery dimensions in swine obtained from catheter angiography, multi-slice computed tomography, 3D-rotational angiography and phase-contrast magnetic resonance angiography.

Authors:  Ryan Pewowaruk; Klarka Mendrisova; Carolina Larrain; Christopher J Francois; Alejandro Roldán-Alzate; Luke Lamers
Journal:  Int J Cardiovasc Imaging       Date:  2020-10-09       Impact factor: 2.357

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.