Interobserver variability in confocal optic nerve analysis (HRT).

Manuel M Hermann1, David F Garway-Heath, Christian P Jonescu-Cuypers, Reinhard O W Burk, Jost B Jonas, Christian Y Mardin, Jens Funk, Michael Diestelhorst.   

Abstract

BACKGROUND/AIMS: To study the observer-related variability of optic nerve head (ONH) measurements using confocal laser scanning tomography (HRT I) in a screening setting.
METHODS: Six experienced glaucoma specialists independently evaluated 50 ONH topographies from 25 adults using HRT software ver. 2.01 in a masked fashion. ONH topographies were obtained from a cohort study of 882 healthy adults and additionally included one patient with glaucomatous eyes. A glaucoma-screening-like setting was intended. The mean interobserver difference was defined as the mean percentual difference between an observer's analysis and the mean of all six observers for all eyes and all observers. The interobserver range was calculated for each eye as the percentual difference between the lowest and highest measurement, with the highest measurement as denominator. Additionally, Kendall's coefficient of rank concordance was assessed for the main HRT parameters.
RESULTS: Mean disc area ranged from 1.83 +/- 0.49 to 2.21 +/- 0.40 mm(2) (mean interobserver difference: 8.3%; interobserver range: 5-50%; rank concordance: 0.86). The lowest mean interobserver differences were found for mean retinal nerve fibre layer thickness (RNFLT; 6.5%), maximum cup depth (2.9%) and cup shape (6.8%). An increased interobserver range was significantly correlated to a low cup to disc area ratio (r=0.64, P<0.0001).
CONCLUSIONS: The observer-dependent diagnostic variability of HRT measurements can lead to divergent diagnostic evaluation of the ONH in a screening setting. Any HRT software relying on a reference database is exposed to relevant observer-related variability of the disc area. For screening purposes, HRT measurements should be completed by other diagnostic methods to compensate for possible diagnostic uncertainty.

Entities:  

Mesh:

Year:  2007        PMID: 17279310     DOI: 10.1007/s10792-006-9022-9

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  22 in total

1.  [Effect of the contour line on cup surface using the Heidelberg Retina Tomograph].

Authors:  K U Bartz-Schmidt; C P Jonescu-Cuypers; G Thumann; J Frucht; G K Krieglstein
Journal:  Klin Monbl Augenheilkd       Date:  1996-11       Impact factor: 0.700

2.  The cumulative normalised rim/disc area ratio curve.

Authors:  K U Bartz-Schmidt; A Sengersdorf; P Esser; P Walter; R D Hilgers; G K Krieglstein
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  1996-04       Impact factor: 3.117

3.  Development of the standard reference plane for the Heidelberg retina tomograph.

Authors:  R O Burk; K Vihanninjoki; T Bartke; A Tuulonen; P J Airaksinen; H E Völcker; J M König
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2000-05       Impact factor: 3.117

4.  Automated segmentation of the optic nerve head for diagnosis of glaucoma.

Authors:  R Chrástek; M Wolf; K Donath; H Niemann; D Paulus; T Hothorn; B Lausen; R Lämmer; C Y Mardin; G Michelson
Journal:  Med Image Anal       Date:  2005-04-08       Impact factor: 8.545

5.  Optic nerve head analyser and Heidelberg retina tomograph: accuracy and reproducibility of topographic measurements in a model eye and in volunteers.

Authors:  P Janknecht; J Funk
Journal:  Br J Ophthalmol       Date:  1994-10       Impact factor: 4.638

6.  Reproducibility of the optic nerve head topography with a new laser tomographic scanning device.

Authors:  K Rohrschneider; R O Burk; F E Kruse; H E Völcker
Journal:  Ophthalmology       Date:  1994-06       Impact factor: 12.079

7.  Inter- and intraobserver variation in the analysis of optic disc images: comparison of the Heidelberg retina tomograph and computer assisted planimetry.

Authors:  D F Garway-Heath; D Poinoosawmy; G Wollstein; A Viswanathan; D Kamal; L Fontana; R A Hitchings
Journal:  Br J Ophthalmol       Date:  1999-06       Impact factor: 4.638

8.  Automated analysis of normal and glaucomatous optic nerve head topography images.

Authors:  N V Swindale; G Stjepanovic; A Chin; F S Mikelberg
Journal:  Invest Ophthalmol Vis Sci       Date:  2000-06       Impact factor: 4.799

9.  Ranked-segment distribution curve for interpretation of optic nerve topography.

Authors:  S Asawaphureekorn; L Zangwill; R N Weinreb
Journal:  J Glaucoma       Date:  1996-04       Impact factor: 2.503

10.  Measurement of optic disc size: equivalence of methods to correct for ocular magnification.

Authors:  D F Garway-Heath; A R Rudnicka; T Lowe; P J Foster; F W Fitzke; R A Hitchings
Journal:  Br J Ophthalmol       Date:  1998-06       Impact factor: 5.908

View more
  4 in total

Review 1.  [Functional glaucoma diagnosis].

Authors:  C Erb; K Göbel
Journal:  Ophthalmologe       Date:  2009-04       Impact factor: 1.059

2.  Intraobserver variability of confocal scanning laser ophthalmoscopy with and without stereo photographs.

Authors:  H T Nguyen; K P Pikey; S K Gardiner; D Gritz; R Krishnadas; G A Cioffi; S L Mansberger
Journal:  Br J Ophthalmol       Date:  2009-11       Impact factor: 4.638

3.  Automated diagnosis of glaucoma using digital fundus images.

Authors:  Jagadish Nayak; Rajendra Acharya U; P Subbanna Bhat; Nakul Shetty; Teik-Cheng Lim
Journal:  J Med Syst       Date:  2009-10       Impact factor: 4.460

4.  Retinal vessel diameter in bilateral glaucoma suspects: comparison between the eye converted to glaucoma and the contralateral non-converted eye.

Authors:  Eunjoo Yoo; Chungkwon Yoo; Tae-Eun Lee; Yong Yeon Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-06-02       Impact factor: 3.117

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.