Comparison between a built-in "dual side" chest imaging device and a standard "single side" CR.

Lucia Riccardi1, Maria Cristina Cauzzo, Roberto Fabbris, Eugenia Tonini, Roberto Righetto.   

Abstract

An integrated readout computed radiography system (Fuji XU-D1) incorporating dual-side imaging plates (ST-55BD) was analyzed in terms of modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) for standard beam qualities RQA 9 and RQA 5. NPS and DQE were assessed using a detector entrance air kerma consistent with clinical practice for chest radiography. Similar investigation was performed on a standard reader (Fuji FCR 5000) using single-side imaging plates (ST-VI). Negligible differences were found between the MTFs of the two imaging systems for RQA 9, whereas for RQA 5 the single-side system exhibited slightly superior MTF. Regarding noise response, the dual-side system turned out to be better performing for both beam qualities over a wide range of frequencies. For RQA 9, at 8 microGy, the DQE of the dual-side system was moderately higher over the whole frequency range, whereas for RQA 5, at 10 microGy, significant improvement was found at low- and midrange frequencies. As an example, at 1 cycle/mm, the following improvements in the DQE of the dual-side system were observed: +22% (RQA 9, at 8 microGy), +50% (RQA 9, at 30 microGy), and +45% (RQA 5, at 10 microGy).

Entities:  

Mesh:

Year:  2007        PMID: 17278497     DOI: 10.1118/1.2400619

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

Review 1.  [Balance of required dose and image quality in digital radiography].

Authors:  M Uffmann; C Schaefer-Prokop; U Neitzel
Journal:  Radiologe       Date:  2008-03       Impact factor: 0.635

2.  Dose reduction in general radiography for adult patients by use of a dual-side-reading photostimulable phosphor plate in a computed radiography system.

Authors:  Norisato Tsuda; Nobukazu Tanaka; Tsutomu Akasaka; Hidetake Yabuuchi; Junji Morishita
Journal:  Radiol Phys Technol       Date:  2014-05-21

3.  Modeling the performance characteristics of computed radiography (CR) systems.

Authors:  Srinivasan Vedantham; Andrew Karellas
Journal:  IEEE Trans Med Imaging       Date:  2010-03       Impact factor: 10.048