Literature DB >> 17254363

Clinical experience with the Bicarbon heart valve prosthesis.

Yoshio Misawa1, Tsutomu Saito, Hiroaki Konishi, Shin-Ichi Oki, Yuichiro Kaminishi, Yasuhito Sakano, Hideki Morita, Kei Aizawa.   

Abstract

BACKGROUND: We have previously reported mid-term results of a study, which ended in January 2000, on the Bicarbon valve. The study concluded that the valve showed excellent clinical results, associated with a low incidence of valve-related complications. In the present study, the same patients were prospectively followed for an additional 5 years.
METHODS: Forty-four patients had aortic valve replacement (AVR), 48 had mitral valve replacement (MVR), and 13 had both aortic and mitral valve replacement (DVR). The mean age of the 105 patients was 61.2 +/- 11.3 years. The mean follow-up was 6.1 +/- 1.9 years with a cumulative follow-up of 616 patient-years.
RESULTS: There were 5 early deaths (4.7%: 4 in the AVR group and 1 in the MVR group) and 21 late deaths (3.4%/patient-year: 5 valve related deaths and 16 valve unrelated deaths). Survival at 8 years was 75.2 +/- 7.0% in the AVR group, 76.6 +/- 6.2% in the MVR group, and 55.4 +/- 16.1% in the DVR group. The linearized incidence of thrombo-embolic complications, hemorrhagic complications, and paravalvular leaks in all patients was 0.65 +/- 1.48%, 0.81 +/- 1.69%, and 0.16 +/- 0.54%/patient-year respectively. No other complications were observed.
CONCLUSION: The Bicarbon prosthetic heart valve has shown excellent long-term clinical results, associated with a low incidence of valve-related complications.

Entities:  

Mesh:

Year:  2007        PMID: 17254363      PMCID: PMC1794238          DOI: 10.1186/1749-8090-2-8

Source DB:  PubMed          Journal:  J Cardiothorac Surg        ISSN: 1749-8090            Impact factor:   1.637


  24 in total

1.  Guidelines for reporting morbidity and mortality after cardiac valvular operations. The American Association for Thoracic Surgery, Ad Hoc Liaison Committee for Standardizing Definitions of Prosthetic Heart Valve Morbidity.

Authors:  L H Edmunds; R E Clark; L H Cohn; G L Grunkemeier; D C Miller; R D Weisel
Journal:  Ann Thorac Surg       Date:  1996-09       Impact factor: 4.330

2.  Single-center experience with the bicarbon bileaflet prosthetic heart valve in Japan.

Authors:  Yoshio Misawa; Tsutomu Saito; Hiroaki Konishi; Shin-ichi Oki; Yuichiro Kaminishi; Yasuhiro Tezuka; Kei Aizawa; Hideki Takahashi; Nobuyuki Hasegawa; Osamu Kamisawa; Morito Kato; Katsuo Fuse
Journal:  Artif Organs       Date:  2002-10       Impact factor: 3.094

3.  Nineteen-millimeter aortic St. Jude Medical heart valve prosthesis: up to sixteen years' follow-up.

Authors:  D Sawant; A K Singh; W C Feng; A A Bert; F Rotenberg
Journal:  Ann Thorac Surg       Date:  1997-04       Impact factor: 4.330

4.  The Bicarbon heart valve prosthesis: short-term results.

Authors:  F Casselman; P Herijgers; B Meyns; W Flameng; W Daenen
Journal:  J Heart Valve Dis       Date:  1997-07

5.  Twenty-year experience with the St Jude Medical mechanical valve prosthesis.

Authors:  John S Ikonomidis; John M Kratz; Arthur J Crumbley; Martha R Stroud; Scott M Bradley; Robert M Sade; Fred A Crawford
Journal:  J Thorac Cardiovasc Surg       Date:  2003-12       Impact factor: 5.209

6.  When and how does nonstructural mechanical prosthetic heart valve dysfunction occur?

Authors:  Yoshio Misawa; Tsutomu Saito; Hiroaki Konishi; Shin-ichi Oki; Yuichiro Kaminishi; Hideki Takahashi; Kei Aizawa; Osamu Kamisawa; Morito Kato; Katsuo Fuse
Journal:  Jpn J Thorac Cardiovasc Surg       Date:  2003-08

7.  A single-center experience of the Sorin Bicarbon heart valve prosthesis: long-term clinical, hematological and hemodynamic results.

Authors:  Donald C Whitaker; Stuart E James; Robin K Walesby
Journal:  J Heart Valve Dis       Date:  2004-01

8.  Sorin Bicarbon bileaflet valve: a 10-year experience.

Authors:  J B Borman; C De Riberolles
Journal:  Eur J Cardiothorac Surg       Date:  2003-01       Impact factor: 4.191

9.  Pivot design in bileaflet valves.

Authors:  F Vallana; S Rinaldi; P M Galletti; A Nguyen; A Piwnica
Journal:  ASAIO J       Date:  1992 Jul-Sep       Impact factor: 2.872

10.  Clinical experience with the Omnicarbon prosthetic heart valve.

Authors:  Y Misawa; T Hasegawa; M Kato
Journal:  J Thorac Cardiovasc Surg       Date:  1993-01       Impact factor: 5.209

View more
  5 in total

Review 1.  Valve-related complications after mechanical heart valve implantation.

Authors:  Yoshio Misawa
Journal:  Surg Today       Date:  2014-12-18       Impact factor: 2.549

2.  Fifteen-year experience with the Bicarbon heart valve prosthesis in a single center.

Authors:  Yoshio Misawa; Arata Muraoka; Shin-ichi Ohki; Kei Aizawa; Koji Kawahito; Tsutomu Saito; Hirotaka Sato; Ippei Takazawa; Soki Kurumisawa; Hirohiko Akutsu; Akira Sugaya
Journal:  J Cardiothorac Surg       Date:  2015-06-28       Impact factor: 1.637

3.  Integrated strategy for in vitro characterization of a bileaflet mechanical aortic valve.

Authors:  Francesca Maria Susin; Stefania Espa; Riccardo Toninato; Stefania Fortini; Giorgio Querzoli
Journal:  Biomed Eng Online       Date:  2017-02-16       Impact factor: 2.819

4.  Low-dose anticoagulation after isolated mechanical aortic valve replacement with Liva Nova Bicarbon prosthesis: A post hoc analysis of LOWERING-IT Trial.

Authors:  Michele Torella; Iolanda Aquila; Paolo Chiodini; Cristiano Amarelli; Gianpaolo Romano; Ester Elena Della Ratta; Marisa De Feo; Alessandro Della Corte; Ciro Indolfi; Daniele Torella; Luca Salvatore De Santo
Journal:  Sci Rep       Date:  2018-05-30       Impact factor: 4.379

5.  Clinical mid-term outcomes of the Chinese-made CL-V bileaflet mechanical heart valve in Chinese patients.

Authors:  Liang Qi; Shidong Liu; Yan Qiang; Honglin Zhao; Ruisheng Liu; Dianwei Cheng; Bing Song; Quanlin Guan
Journal:  J Thorac Dis       Date:  2021-01       Impact factor: 2.895

  5 in total