Pivot design in bileaflet valves.

F Vallana1, S Rinaldi, P M Galletti, A Nguyen, A Piwnica.   

Abstract

The design criteria leading to the development of a new bileaflet valve (Sorin Bicarbon) were derived from the analysis of functional requirements, the performance of existing prostheses, and the availability of an advanced carbon coating technology (Carbofilm). The hinge is the critical element affecting fluid dynamics, durability, and thrombus formation in bileaflet valves. A comparative study of three existing models led to a new hinge design that was based on coupling two spheric surfaces with different radii of curvature (leaflet pivot and hinge recess) and obtained by electroerosion into a Carbofilm-coated metallic housing. In this valve, the point of contact moves continuously by rolling, not sliding. This minimizes friction and wear and allows uninterrupted washing of the blood exposed surfaces even during diastole (a finding established in patients using transesophageal echocardiography). Tricuspid implantation without anticoagulation in 33 sheep did not lead to thrombotic events (follow-up, 40-400 days). In the first 36 clinical implants observed for 15 months (mitral position, size 29; two unrelated deaths), the mean diastolic gradient by echo Doppler was 4 +/- 1.25 mmHg; the functional area was 3.2 +/- 0.6 cm2. No leaflet fracture and no thrombotic or embolic complications were observed clinically using a standard anticoagulant regimen.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1457931     DOI: 10.1097/00002480-199207000-00107

Source DB:  PubMed          Journal:  ASAIO J        ISSN: 1058-2916            Impact factor:   2.872


  2 in total

1.  Clinical experience with the Bicarbon heart valve prosthesis.

Authors:  Yoshio Misawa; Tsutomu Saito; Hiroaki Konishi; Shin-Ichi Oki; Yuichiro Kaminishi; Yasuhito Sakano; Hideki Morita; Kei Aizawa
Journal:  J Cardiothorac Surg       Date:  2007-01-25       Impact factor: 1.637

2.  Long-term experience with the Sorin Bicarbon and Edwards Mira mechanical valve prostheses in the mitral position.

Authors:  Toshihiro Fukui; Toshihiko Shibata; Yasuyuki Sasaki; Hidekazu Hirai; Manabu Motoki; Yosuke Takahashi; Shigefumi Suehiro
Journal:  Gen Thorac Cardiovasc Surg       Date:  2007-03
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.