OBJECTIVE: The aim of the work was to establish optimum ambient light conditions for viewing radiologic images of the wrist on liquid crystal display monitors.
MATERIALS AND METHODS: Five ambient light levels were investigated: 480, 100, 40, 25, and 7 lux. Seventy-nine experienced radiologists were asked to examine 30 posteroanterior wrist images and decide whether a fracture was present. All images were displayed on liquid crystal display monitors. Receiver operating characteristic analysis was performed, and the numbers of false-positive and false-negative findings were recorded.
RESULTS: For all the radiologists, greater area under the receiver operating characteristic curve and lower numbers of false-positive and false-negative findings were recorded at 40 and 25 lux compared with 480 and 100 lux. At 7 lux, the results were generally similar to those at 480 and 100 lux. The experience and knowledge of radiologists specializing in imaging of musculoskeletal trauma appeared to compensate in part for inappropriate lighting levels.
CONCLUSION: Typical office lighting and current recommendations on ambient lighting can reduce diagnostic efficacy compared with lower levels of ambient lighting. If, however, no light other than that of the monitor is used, results are similar to those with excessive levels of lighting. Careful control of ambient lighting is therefore required to ensure that diagnostic accuracy is maximized, particularly for clinicians not expert in interpreting posteroanterior wrist images.

Entities:  

Mesh:

Year:  2007        PMID: 17242225     DOI: 10.2214/AJR.05.2048

Source DB:  PubMed          Journal:  AJR Am J Roentgenol        ISSN: 0361-803X            Impact factor:   3.959


  29 in total

1.  The effects of ambient lighting in chest radiology reading rooms.

Authors:  Benjamin J Pollard; Ehsan Samei; Amarpreet S Chawla; Craig Beam; Laura E Heyneman; Lynne M Hurwitz Koweek; Santiago Martinez-Jimenez; Lacey Washington; Noriyuki Hashimoto; H Page McAdams
Journal:  J Digit Imaging       Date:  2012-08       Impact factor: 4.056

2.  Effect of ambient light level at the monitor surface on digital radiographic evaluation of approximal carious lesions: an in vitro study.

Authors:  K Hellén-Halme; A Lith
Journal:  Dentomaxillofac Radiol       Date:  2012-03       Impact factor: 2.419

3.  The impact of pediatric-specific dose modulation curves on radiation dose and image quality in head computed tomography.

Authors:  Joana Santos; Shane Foley; Graciano Paulo; Mark F McEntee; Louise Rainford
Journal:  Pediatr Radiol       Date:  2015-08-05

4.  Grey-scale inversion improves detection of lung nodules.

Authors:  J W Robinson; J T Ryan; M F McEntee; S J Lewis; M G Evanoff; L A Rainford; P C Brennan
Journal:  Br J Radiol       Date:  2012-05-09       Impact factor: 3.039

5.  Mammographic density measurements are not affected by mammography system.

Authors:  Christine N Damases; Patrick C Brennan; Mark F McEntee
Journal:  J Med Imaging (Bellingham)       Date:  2015-03-04

6.  DICOM part 14: GSDF-calibrated medical grade monitor vs a DICOM part 14: GSDF-calibrated "commercial off-the-shelf" (COTS) monitor for viewing 8-bit dental images.

Authors:  D J McIlgorm; J P McNulty
Journal:  Dentomaxillofac Radiol       Date:  2014-11-25       Impact factor: 2.419

7.  Intercountry analysis of breast density classification using visual grading.

Authors:  Christine N Damases; Peter Hogg; Mark F McEntee
Journal:  Br J Radiol       Date:  2017-06-14       Impact factor: 3.039

8.  Grey-scale inversion improves detection of lung nodules.

Authors:  J W Robinson; J T Ryan; M F McEntee; S J Lewis; M G Evanoff; L A Rainford; P C Brennan
Journal:  Br J Radiol       Date:  2013-01       Impact factor: 3.039

9.  3T MRI of the knee with optimised isotropic 3D sequences: Accurate delineation of intra-articular pathology without prolonged acquisition times.

Authors:  Osamah M Abdulaal; Louise Rainford; Peter MacMahon; Eoin Kavanagh; Marie Galligan; James Cashman; Allison McGee
Journal:  Eur Radiol       Date:  2017-04-21       Impact factor: 5.315

10.  A status report on free-response analysis.

Authors:  D P Chakraborty
Journal:  Radiat Prot Dosimetry       Date:  2010-01-18       Impact factor: 0.972

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.