Comparison of 4 MV photon surface dose among Varian, Siemens, and Elekta linear accelerators for tangential breast treatment: a phantom study.

Satoshi Yamaguchi1, Kumiko Karasawa, Tomohisa Furuya, Takayuki Fujita, Yutaka Tutumi, Kohei Miura, Takahiro Takada, Kana Ito, Shuichi Ozawa.   

Abstract

PURPOSE: We have compared the differences in a 4-MV photon surface dose among Varian, Siemens, and Elekta linear accelerators (linacs) with wedges for tangential breast treatment.
MATERIALS AND METHODS: The wedge factor and the surface dose were measured using a solid water phantom and an ion chamber for each linear accelerator with various field sizes and wedge angles. A tangential treatment plan was applied to an elliptical hollow cylinder water phantom with a radiochromic film placed thereon. A dose was delivered to a simulated target in the phantom, and the resulting dose distribution was analyzed using a film scanner.
RESULTS: Varian's wedges resulted in the highest wedge factors, ranging from 0.37 to 0.75 depending on the wedge angles. Varian's wedges led to the highest normalized skin doses, ranging between 0.40 and 0.73 depending on the wedge angles and field sizes. In the cylinder phantom test with two tangential beams, the Varian linac provided a nearly 20% higher maximum dose than the Siemens and Elekta linacs.
CONCLUSION: The Varian linac resulted in the highest surface doses, and the Elekta linac led to the lowest for nearly all the measurement conditions we employed, including open beams.

Entities:  

Mesh:

Year:  2007        PMID: 17225047     DOI: 10.1007/s11604-006-0094-2

Source DB:  PubMed          Journal:  Radiat Med        ISSN: 0288-2043


  9 in total

1.  Measurement of radiotherapy x-ray skin dose on a chest wall phantom.

Authors:  K Y Quach; J Morales; M J Butson; A B Rosenfeld; P E Metcalfe
Journal:  Med Phys       Date:  2000-07       Impact factor: 4.071

2.  Photon beam skin dose analyses for different clinical setups.

Authors:  S Kim; C R Liu; T C Zhu; J R Palta
Journal:  Med Phys       Date:  1998-06       Impact factor: 4.071

3.  Comparison of build-up dose between Elekta and Varian linear accelerators for high-energy photon beams using radiochromic film and clinical implications for IMRT head and neck treatments.

Authors:  L Paelinck; C De Wagter; A Van Esch; W Duthoy; T Depuydt; W De Neve
Journal:  Phys Med Biol       Date:  2005-02-07       Impact factor: 3.609

Review 4.  Pathophysiology of irradiated skin and breast.

Authors:  J O Archambeau; R Pezner; T Wasserman
Journal:  Int J Radiat Oncol Biol Phys       Date:  1995-03-30       Impact factor: 7.038

5.  Impact of radiation therapy on acute toxicity in breast conservation therapy for early breast cancer.

Authors:  M Back; M Guerrieri; C Wratten; A Steigler
Journal:  Clin Oncol (R Coll Radiol)       Date:  2004-02       Impact factor: 4.126

6.  Outcome of breast-conserving therapy in the Tokyo Women's Medical University Breast Cancer Society experience.

Authors:  Kumiko Karasawa; Takao Obara; Tadao Shimizu; Shunsuke Haga; Takahiro Okamoto; Yukio Ito; Takako Kamio; Tsunehito Kimura; Shingo Kameoka; Norio Mitsuhashi; Toshio Nishikawa; Motohiko Aiba; Mazaru Izuo
Journal:  Breast Cancer       Date:  2003       Impact factor: 4.239

7.  Intensity modulated radiation therapy (IMRT) decreases acute skin toxicity for women receiving radiation for breast cancer.

Authors:  Gary M Freedman; Penny R Anderson; Jinsheng Li; Debra F Eisenberg; Alexandra L Hanlon; Lu Wang; Nicos Nicolaou
Journal:  Am J Clin Oncol       Date:  2006-02       Impact factor: 2.339

8.  Determination of field size-dependent wedge factors from a few selected measurements.

Authors:  Richard A Popple; Ivan A Brezovich; Jun Duan; Sui Shen; Prem N Pareek; Sung-Joon Ye
Journal:  J Appl Clin Med Phys       Date:  2005-01-12       Impact factor: 2.102

9.  Surface and buildup dose characteristics for 6, 10, and 18 MV photons from an Elekta Precise linear accelerator.

Authors:  Eric E Klein; Jacqueline Esthappan; Zuofeng Li
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

  9 in total
  1 in total

1.  Surface and build-up dose comparison between Elekta 6 MV flattening filter and flattening-filter-free beams using an advanced Markus ionization chamber and a solid water-equivalent phantom.

Authors:  Toshikazu Imae; Shigeharu Takenaka; Yuichi Watanabe; Atsushi Aoki; Kanako Matsuda; Katsutake Sasaki; Shigeki Saegusa; Kanabu Nawa; Keiichi Nakagawa; Osamu Abe
Journal:  J Appl Clin Med Phys       Date:  2020-11-12       Impact factor: 2.102

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.