Literature DB >> 17156149

Computer-assisted colonoscopy (the NeoGuide Endoscopy System): results of the first human clinical trial ("PACE study").

Axel Eickhoff1, Jacques van Dam, Ralf Jakobs, Valerie Kudis, Dirk Hartmann, Ulrich Damian, Uwe Weickert, Dieter Schilling, Jürgen F Riemann.   

Abstract

OBJECTIVES: Unsedated colonoscopy is an uncomfortable procedure for most patients. Discomfort during colonoscopy is largely related to looping of the colonoscope, which displaces the colon from its native configuration and stretches attachments to the mesentery. A novel computer-assisted colonoscope utilizes a fully articulated, computer-controlled insertion tube. On manual insertion of the colonoscope, the position and angle of the scope's tip are encoded into a computer algorithm. As the colonoscope is advanced, the computer directs each successive segment to take the same shape that the tip had at a given insertion depth. The insertion tube thus changes its shape at different insertion depths in a "follow-the-leader" manner.
METHODS: This initial clinical trial with this novel colonoscopy system was designed as a prospective, nonrandomized, unblinded, feasibility study. Three physicians of varying levels of experience participated in the study.
RESULTS: Eleven consecutive patients (seven men, four women, age range 19-80) meeting inclusion criteria for screening or diagnostic colonoscopy were enrolled in the study. The cecum was reached in 10 consecutive patients (100%). Findings included diverticular disease in two cases and multiple colonic polyps in two cases. Postprocedure assessment at discharge, 48 h, and 30 days revealed no complications or adverse effects. Physician satisfaction and patient acceptance of this new technique were high.
CONCLUSIONS: In this limited, first of its kind feasibility study, the computer-assisted colonoscope was shown to perform colonoscopy safely and effectively. The colonoscope's unique design limited loop formation during colonoscopy. Large-scale clinical trials are indicated.

Entities:  

Mesh:

Year:  2006        PMID: 17156149     DOI: 10.1111/j.1572-0241.2006.01002.x

Source DB:  PubMed          Journal:  Am J Gastroenterol        ISSN: 0002-9270            Impact factor:   10.864


  35 in total

1.  Magnetic air capsule robotic system: proof of concept of a novel approach for painless colonoscopy.

Authors:  P Valdastri; G Ciuti; A Verbeni; A Menciassi; P Dario; A Arezzo; M Morino
Journal:  Surg Endosc       Date:  2011-12-17       Impact factor: 4.584

2.  Determining scope position during colonoscopy without use of ionizing radiation or magnetic imaging: the enhanced mapping ability of the NeoGuide Endoscopy System.

Authors:  Johannes Striegel; Ralf Jakobs; Jacques Van Dam; Uwe Weickert; Jürgen F Riemann; Axel Eickhoff
Journal:  Surg Endosc       Date:  2010-08-21       Impact factor: 4.584

3.  New endoscopy devices to improve population adherence to colorectal cancer prevention programs.

Authors:  Asimina Gaglia; Ioannis S Papanikolaou; Wilfried Veltzke-Schlieker
Journal:  World J Gastrointest Endosc       Date:  2010-07-16

4.  Sedation-risk-free colonoscopy for minimizing the burden of colorectal cancer screening.

Authors:  Felix W Leung; Abdulrahman M Aljebreen; Emilio Brocchi; Eugene B Chang; Wei-Chih Liao; Takeshi Mizukami; Melvin Schapiro; Konstantinos Triantafyllou
Journal:  World J Gastrointest Endosc       Date:  2010-03-16

Review 5.  Application of robotics in gastrointestinal endoscopy: A review.

Authors:  Baldwin Po Man Yeung; Philip Wai Yan Chiu
Journal:  World J Gastroenterol       Date:  2016-02-07       Impact factor: 5.742

Review 6.  Mechanical analysis of insertion problems and pain during colonoscopy: why highly skill-dependent colonoscopy routines are necessary in the first place... and how they may be avoided.

Authors:  Arjo J Loeve; Paul Fockens; Paul Breedveld
Journal:  Can J Gastroenterol       Date:  2013       Impact factor: 3.522

Review 7.  Methods of reducing discomfort during colonoscopy.

Authors:  Felix W Leung
Journal:  Dig Dis Sci       Date:  2008-06       Impact factor: 3.199

8.  Robotics in Colonoscopy.

Authors:  Dan Cater; Arpita Vyas; Dinesh Vyas
Journal:  Am J Robot Surg       Date:  2014-06

Review 9.  Advanced endoscopic technologies for colorectal cancer screening.

Authors:  Keith L Obstein; Pietro Valdastri
Journal:  World J Gastroenterol       Date:  2013-01-28       Impact factor: 5.742

10.  Pilot feasibility study of the method of water infusion without air insufflation in sedated colonoscopy.

Authors:  Joseph W Leung; Rodolei Salera; Lee Toomsen; Surinder Mann; Felix W Leung
Journal:  Dig Dis Sci       Date:  2008-12-05       Impact factor: 3.199

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.