Abstract
OBJECTIVE: To evaluate the effect of thin overlapping reconstruction on the attenuation of small (< or =3 cm) renal cysts in the nephrographic phase of multidetector CT (MDCT).Entities:
Mesh:
Substances:
Year: 2006 PMID: 17143031 PMCID: PMC2667614 DOI: 10.3348/kjr.2006.7.4.275
Source DB: PubMed Journal: Korean J Radiol ISSN: 1229-6929 Impact factor: 3.500
Fig. 1Representative cases showing the importance of falsely increased attenuation in small renal cysts in the nephrographic phase of MDCT. All scans were performed with the 16-channel MDCT scanner used in the phantom study. Scans in the nephrographic phase were obtained 150 seconds after intravenous injection of contrast material (same material used in the phantom study). Slice thickness / interval was 5 mm/5 mm.
Fig. 2CT images of the phantom.
Attenuation Increase According to Slice Thickness/Interval and Sphere Size on 4- and 16-Channel MDCT (mean±SD [HU])
Note.-4 ch = 4-channel MDCT, 16 ch = 16-channel MDCT
*p < 0.05 versus @; **p < 0.05 versus ***; **, ***, and #p < 0.05 versus the values with T/I = 5/5 in the same sphere