Comparison of dynamic contour tonometry and goldmann applanation tonometry in African American subjects.

Felipe A Medeiros1, Pamela A Sample, Robert N Weinreb.   

Abstract

PURPOSE: To evaluate the relationship between intraocular pressure (IOP) measurements obtained by dynamic contour tonometry (DCT) and Goldmann applanation tonometry (GAT) in African Americans and to assess whether these measures were influenced by ocular parameters including corneal thickness, corneal curvature, and axial length.
DESIGN: Observational clinical study. PARTICIPANTS: The study included 176 eyes of 94 African Americans.
METHODS: All participants underwent IOP evaluation with DCT and GAT, as well as measurements of central corneal thickness (CCT), corneal curvature, corneal astigmatism, spherical equivalent, and axial length. MAIN OUTCOME MEASURES: Univariate and multiple regression analyses were used to evaluate the associations between IOP (as measured with DCT and GAT) and CCT, corneal curvature, corneal astigmatism, spherical equivalent, axial length, and age. Bland-Altman plots were used to evaluate the agreement between IOP measurements obtained by DCT and GAT.
RESULTS: In multiple regression analysis, GAT IOP measurements were significantly associated with CCT (P<0.001) and age (P = 0.003), whereas DCT IOP measurements were associated only with age (P = 0.027). The difference between DCT and GAT IOP measurements was significantly influenced by corneal thickness (P<0.001) and ocular pulse amplitude (OPA; P = 0.004). Patients with thick corneas tended to have higher GAT IOP measurements compared with DCT, whereas in patients with thin corneas, GAT IOP measurements tended to be lower than DCT measurements. Larger values of OPA were associated with a decrease in the difference between DCT and GAT measurements, whereas smaller values of OPA were associated with an increase in the difference between DCT and GAT measurements.
CONCLUSIONS: Dynamic contour tonometry measurements in African Americans seem to provide an estimate of IOP that is less influenced by corneal properties than those provided by GAT.

Entities:  

Mesh:

Year:  2006        PMID: 17141320     DOI: 10.1016/j.ophtha.2006.06.065

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  12 in total

1.  Intraocular pressure after Descemet's stripping and non-Descemet's stripping automated endothelial keratoplasty.

Authors:  Yoshiro Mawatari; Akira Kobayashi; Hideaki Yokogawa; Kazuhisa Sugiyama
Journal:  Jpn J Ophthalmol       Date:  2011-03-13       Impact factor: 2.447

2.  Dynamic contour tonometry over silicone hydrogel contact lens.

Authors:  Andrew K C Lam; Jimmy S H Tse
Journal:  J Optom       Date:  2013-08-28

3.  Goldmann applanation tonometry and dynamic contour tonometry in eyes with elevated intraocular pressure (IOP): comparison in the same eyes after subsequent medical normalization of IOP.

Authors:  Chungkwon Yoo; Young Sub Eom; Yong Yeon Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-07-22       Impact factor: 3.117

4.  The African Descent and Glaucoma Evaluation Study (ADAGES): design and baseline data.

Authors:  Pamela A Sample; Christopher A Girkin; Linda M Zangwill; Sonia Jain; Lyne Racette; Lida M Becerra; Robert N Weinreb; Felipe A Medeiros; M Roy Wilson; Julio De León-Ortega; Celso Tello; Christopher Bowd; Jeffrey M Liebmann
Journal:  Arch Ophthalmol       Date:  2009-09

5.  Correlation of intraocular pressure measured with goldmann and dynamic contour tonometry in normal and glaucomatous eyes.

Authors:  Tony Realini; Robert N Weinreb; Gerald Hobbs
Journal:  J Glaucoma       Date:  2009-02       Impact factor: 2.503

6.  The effect of contact lens-induced corneal edema on Goldmann applanation tonometry and dynamic contour tonometry.

Authors:  Jong-Hyun Oh; Chungkwon Yoo; Yong Yeon Kim; Hyo-Myung Kim; Jong-Suk Song
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-10-09       Impact factor: 3.117

7.  A comparison between Goldmann applanation tonometry and dynamic contour tonometry after photorefractive keratectomy.

Authors:  Afshin Lotfi Sadigh; Rohollah F Fouladi; Hassan Hashemi; Amir Houshang Beheshtnejad
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-09-02       Impact factor: 3.117

8.  Comparison of the quality score of intraocular pressure and ocular pulse amplitude values measured by the Pascal dynamic contour tonometer.

Authors:  S Pourjavan; P Boghossian; M Detry-Morel
Journal:  Int Ophthalmol       Date:  2008-11-20       Impact factor: 2.031

9.  Measuring accurate IOPs: Does correction factor help or hurt?

Authors:  Pinakin Gunvant; Robert D Newcomb; Elliot M Kirstein; Victor E Malinovsky; Richard J Madonna; Richard E Meetz
Journal:  Clin Ophthalmol       Date:  2010-07-21

10.  Interpretation of uniocular and binocular trials of glaucoma medications: an observational case series.

Authors:  Christopher T Leffler; Lina Amini
Journal:  BMC Ophthalmol       Date:  2007-10-05       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.