NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes.

Brad J Kemp1, Chang Kim, John J Williams, Alexander Ganin, Val J Lowe.   

Abstract

UNLABELLED: The National Electrical Manufacturers Association (NEMA) NU 2-2001 performance measurements were conducted on the Discovery RX, a whole-body PET/CT system under development by GE Healthcare. The PET scanner uses 4.2 x 6.3 x 30 mm lutetium yttrium orthosilicate (LYSO) crystals grouped in 9 x 6 blocks. There are 24 rings with 630 crystals per ring and the ring diameter is 88.6 cm. The transaxial and axial fields of view are 70.0 and 15.7 cm, respectively. The scanner has retractable septa and can operate in both 2-dimensional (2D) and 3-dimensional (3D) modes. 2D acquisitions use ring differences of +/-4 for direct and +/-5 for cross slices; 3D acquisitions use a ring difference of 23. The coincident window width is 6.5 ns and the energy window is 425-650 keV. Other than the detectors, the system uses the same hardware and software as a Discovery ST. The CT scanner is a 16-slice LightSpeed; the performance characteristics of the CT component are not included herein.
METHODS: Performance measurements of sensitivity, spatial resolution, image quality, scatter fraction and counting rate performance, and image quality were obtained using NEMA methodology.
RESULTS: The system sensitivity in 2D and 3D was measured as 1.7 cps/kBq and 7.3 cps/kBq, respectively. The transaxial resolution for 2D (3D) was 5.1 mm full width at half maximum (FWHM) (5.0 mm) at 1 cm from gantry center and the radial and tangential resolutions were 5.9 mm (5.9 mm) and 5.1 mm (5.2 mm) at 10 cm, respectively. The axial resolution for 2D (3D) was 4.8 mm FWHM (5.8 mm) and 6.3 mm (6.5 mm) at 1 cm and 10 cm from gantry center, respectively. The scatter fraction was 13.1% and 31.8% in 2D and 3D. The peak noise equivalent count rate (NECR) was 155 kcps at 92.1 kBq/mL in 2D and 117.7 kcps at 21.7 kBq/mL in 3D for a noise-free estimation of randoms. The contrast of the 22, 17, 13, and 10 mm hot spheres in the image quality phantom in 2D (3D) were 74.6% (72.4%), 56.7% (59.5%), 46.2% (44.6%), and 17.9% (18.0%), respectively.
CONCLUSION: The Discovery RX is a scanner that possesses high NECR, low scatter fraction, and good spatial resolution characteristics.

Entities:  

Mesh:

Year:  2006        PMID: 17138738

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  34 in total

Review 1.  Positron emission tomography imaging in neurological disorders.

Authors:  Marios Politis; Paola Piccini
Journal:  J Neurol       Date:  2012-09       Impact factor: 4.849

2.  Noise propagation in resolution modeled PET imaging and its impact on detectability.

Authors:  Arman Rahmim; Jing Tang
Journal:  Phys Med Biol       Date:  2013-09-13       Impact factor: 3.609

3.  Performance of the new generation of whole-body PET/CT scanners: Discovery STE and Discovery VCT.

Authors:  M Teräs; T Tolvanen; J J Johansson; J J Williams; J Knuuti
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-07-28       Impact factor: 9.236

4.  Comparison between two super-resolution implementations in PET imaging.

Authors:  Guoping Chang; Tinsu Pan; Feng Qiao; John W Clark; Osama R Mawlawi
Journal:  Med Phys       Date:  2009-04       Impact factor: 4.071

5.  A practical, automated quality assurance method for measuring spatial resolution in PET.

Authors:  Martin A Lodge; Arman Rahmim; Richard L Wahl
Journal:  J Nucl Med       Date:  2009-07-17       Impact factor: 10.057

6.  Washout of ⁸²Rb as a marker of impaired tissue integrity, obtained by list-mode cardiac PET/CT: relationship with perfusion/metabolism patterns of myocardial viability.

Authors:  David T Chien; Paco Bravo; Takahiro Higuchi; Jennifer Merrill; Frank M Bengel
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-04-29       Impact factor: 9.236

7.  Reliability of predicting image signal-to-noise ratio using noise equivalent count rate in PET imaging.

Authors:  Tingting Chang; Guoping Chang; John W Clark; Rami H Diab; Eric Rohren; Osama R Mawlawi
Journal:  Med Phys       Date:  2012-10       Impact factor: 4.071

8.  Implementation of an automated respiratory amplitude gating technique for PET/CT: clinical evaluation.

Authors:  Guoping Chang; Tingting Chang; Tinsu Pan; John W Clark; Osama R Mawlawi
Journal:  J Nucl Med       Date:  2009-12-15       Impact factor: 10.057

9.  Bayesian PET image reconstruction incorporating anato-functional joint entropy.

Authors:  Jing Tang; Arman Rahmim
Journal:  Phys Med Biol       Date:  2009-11-11       Impact factor: 3.609

10.  Rubidium-82 PET-CT for quantitative assessment of myocardial blood flow: validation in a canine model of coronary artery stenosis.

Authors:  Riikka Lautamäki; Richard T George; Kakuya Kitagawa; Takahiro Higuchi; Jennifer Merrill; Corina Voicu; Anthony DiPaula; Stephan G Nekolla; João A C Lima; Albert C Lardo; Frank M Bengel
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-11-05       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.