Relationship of fasting and hourly blood glucose levels to HbA1c values: safety, accuracy, and improvements in glucose profiles obtained using a 7-day continuous glucose sensor.

Satish Garg1, Lois Jovanovic.   

Abstract

OBJECTIVE: In this study, we evaluated the safety and efficacy of 7-day transcutaneous, real-time, continuous glucose monitoring (CGM) in subjects with insulin-requiring diabetes. RESEARCH DESIGN AND METHODS: Eighty-six subjects were enrolled at five U.S. centers. Subjects wore a sensor inserted under the skin of the abdomen for 7 days during each of three consecutive periods. Data were blinded during period 1 and unblinded during periods 2 and 3.
RESULTS: Of the 6,811 matched self-monitoring of blood glucose to sensor values prospectively analyzed, 97.2% fell in the Clarke error grid zones A and B, and median absolute relative difference was 11.4%. After unblinding, subjects reduced time spent at <55 mg/dl by 0.3 h/day, reduced time spent at >240 mg/dl by 1.5 h/day, and increased time in the target zone (81-140 mg/dl) by 1.4 h/day (P < 0.05 for all three comparisons). Improvements were seen in both types 1 and 2 diabetes and with use of both multiple daily injections and continuous subcutaneous insulin infusion. Modal day graphs were generated in six groups of subjects based on HbA1c (A1C) (<or=6, 6-7, 7-8, 8-9, 9-10, and >10%). Mean glucose levels from midnight to 7:00 a.m. (fasting and dawn phenomenon periods) were only normal for subjects with A1C <or=6%. All other groups were hyperglycemic during this and all periods. Reductions in overall mean glucose were achieved for the four highest A1C groupings with unblinded device use.
CONCLUSIONS: This is the first report of a real-time, transcutaneous glucose sensor that functioned for 7 days. The use of CGM in the unblinded phase resulted in improvements in target-range glycemia across all A1C values.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 17130198     DOI: 10.2337/dc06-1361

Source DB:  PubMed          Journal:  Diabetes Care        ISSN: 0149-5992            Impact factor:   19.112


  39 in total

1.  Continuous glucose monitoring to assess the ecologic validity of dietary glycemic index and glycemic load.

Authors:  Anthony N Fabricatore; Cara B Ebbeling; Thomas A Wadden; David S Ludwig
Journal:  Am J Clin Nutr       Date:  2011-11-09       Impact factor: 7.045

2.  Real-time glucose estimation algorithm for continuous glucose monitoring using autoregressive models.

Authors:  Yenny Leal; Winston Garcia-Gabin; Jorge Bondia; Eduardo Esteve; Wifredo Ricart; Jose-Manuel Fernández-Real; Josep Vehí
Journal:  J Diabetes Sci Technol       Date:  2010-03-01

3.  Effect of short-term use of a continuous glucose monitoring system with a real-time glucose display and a low glucose alarm on incidence and duration of hypoglycemia in a home setting in type 1 diabetes mellitus.

Authors:  Raymond J Davey; Timothy W Jones; Paul A Fournier
Journal:  J Diabetes Sci Technol       Date:  2010-11-01

4.  Are Systematic Reviews and Meta-Analyses Appropriate Tools for Assessing Evolving Medical Device Technologies?

Authors:  David Price; Claudia Graham; Christopher G Parkin; Thomas A Peyser
Journal:  J Diabetes Sci Technol       Date:  2015-09-29

5.  The Future of Glucose Monitoring.

Authors:  Satish K Garg
Journal:  Diabetes Technol Ther       Date:  2016-02       Impact factor: 6.118

Review 6.  A tale of two compartments: interstitial versus blood glucose monitoring.

Authors:  Eda Cengiz; William V Tamborlane
Journal:  Diabetes Technol Ther       Date:  2009-06       Impact factor: 6.118

7.  Improved glycemic control in intensively treated adult subjects with type 1 diabetes using insulin guidance software.

Authors:  Satish K Garg; Tevin R Bookout; Kim K McFann; William C Kelly; Christie Beatson; Samuel L Ellis; Raymond S Gutin; Peter A Gottlieb
Journal:  Diabetes Technol Ther       Date:  2008-10       Impact factor: 6.118

8.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

9.  The correlation of hemoglobin A1c to blood glucose.

Authors:  Ken Sikaris
Journal:  J Diabetes Sci Technol       Date:  2009-05-01

10.  An analysis of the SEVEN system: have we reached the summit of needle-type sensor accuracy?

Authors:  Iris M E Wentholt; J Hans DeVries
Journal:  J Diabetes Sci Technol       Date:  2009-09-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.