Accuracy study of computer-assisted drilling: the effect of bone density, drill bit characteristics, and use of a mechanical guide.

T Hüfner1, J Geerling, G Oldag, M Richter, M Kfuri, T Pohlemann, C Krettek.   

Abstract

OBJECTIVE: This study was designed to determine the clinical relevant accuracy of CT-based navigation for drilling.
DESIGN: Experimental model.
SETTING: Laboratory.
METHODS: Twelve drills of varying lengths and diameters were tested with 2 different set-ups. Group 1 used free-hand navigated drilling technique with foam blocks equipped with titanium target points. Group 2 (control) used a newly developed 3-dimensional measurement device equipped with titanium target points with a fixed entry for the navigated drill to minimize bending forces. One examiner performed 690 navigated drillings using solely the monitor screen for control in both groups. The difference between the planned and the actual starting and target point (up to 150 mm distance) was measured (mm). STATISTICS: Levene test and a nonpaired t test. Significance level was set as P < 0.05.
RESULTS: The core accuracy of the navigation system measured with the 3-dimensional device was 0.5 mm. The mean distance from planned to actual entry points in group 1 was 1.3 (range, 0.6-3.4 mm). The mean distance between planned and actual target point was 3.4 (range, 1.7-5.8 mm). Free-hand navigated drilling showed an increased difference with increased length of the drill bits as well as with increased drilling channel for drill bits 2.5 and 3.2 mm and not for 3.5 and 4.5 mm (P < 0.05).
CONCLUSIONS: The core accuracy of the navigation system is high. Compared with the navigated free-hand technique, the results suggest that drill bit deflection interferes directly with the precision. The precision is decreased when using small diameter and longer drill bits.

Entities:  

Mesh:

Year:  2005        PMID: 15891540

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  9 in total

1.  Computer Navigation for Pediatric Femoral ACL Tunnel Placement.

Authors:  Charles A Popkin; Charles M Chan; Jared A Nowell; Stephen G Crowley; Margaret Wright; Christopher S Ahmad
Journal:  Iowa Orthop J       Date:  2019

2.  [Electromagnetic navigation of transpedicular punctures: more precise than the standard?].

Authors:  J A K Ohnsorge; K Kowalski; P Bruners; M Weisskopf; M Lühmann; T Penzkofer; U Maus
Journal:  Orthopade       Date:  2013-03       Impact factor: 1.087

3.  Anterior and distal tunnel orientation for anatomic reconstruction of the medial patellofemoral ligament is safer in patients with open growth plates.

Authors:  Sebastián Irarrázaval; Pablo Besa; Francisco Fernández; Tomás Fernández; Maria Tuca; Maria J Lira; Mario Orrego
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2020-08-18       Impact factor: 4.342

4.  [Navigated retrograde drilling in Osteochondrosis dissecans (OCD) of the talus].

Authors:  M Richter; S Zech
Journal:  Oper Orthop Traumatol       Date:  2011-12       Impact factor: 1.154

5.  [A new parallel drill guide for navigating femoral neck screw placement. Development and evaluation].

Authors:  D Kendoff; T Hüfner; M Citak; C Maier; F Wesemeier; A Pearle; C Krettek
Journal:  Unfallchirurg       Date:  2006-10       Impact factor: 1.000

Review 6.  Drilling of bone: A comprehensive review.

Authors:  Rupesh Kumar Pandey; S S Panda
Journal:  J Clin Orthop Trauma       Date:  2013-01-18

Review 7.  Thermal Osteonecrosis Caused by Bone Drilling in Orthopedic Surgery: A Literature Review.

Authors:  Charles Timon; Conor Keady
Journal:  Cureus       Date:  2019-07-24

8.  Fiber-reinforced composite post removal using guided endodontics: a case report.

Authors:  Changgi Cho; Hyo Jin Jo; Jung-Hong Ha
Journal:  Restor Dent Endod       Date:  2021-09-23

9.  Comparative Analysis of Optoelectronic Accuracy in the Laboratory Setting Versus Clinical Operative Environment: A Systematic Review.

Authors:  Bryan W Cunningham; Daina M Brooks
Journal:  Global Spine J       Date:  2022-04
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.