Literature DB >> 16806616

Migration and cyclic motion of a new short-stemmed hip prosthesis--a biomechanical in vitro study.

F M Westphal1, N Bishop, M Honl, E Hille, K Püschel, M M Morlock.   

Abstract

BACKGROUND: Uncemented, short-stemmed hip prostheses have been developed to reduce the risk of stress shielding and to preserve femural bone stock. The long-term success of these implants is yet uncertain. Prerequisite for osseointegration is sufficient primary stability. In this study the cyclic motion and migration patterns of a new short-stemmed hip implant were compared with those for two clinically successful shaft prostheses.
METHODS: The prostheses were implanted in paired fresh human femura and loaded dynamically (gait cycle) with increasing load (max 2,100 N) up to 15,000 cycles. Relative displacements between prosthesis and bone were recorded using a 3D-video analysis system.
FINDINGS: The short stem displayed a biphasic migration pattern with stabilisation at maximum load. Initial migration was predominantly into varus and was greater than that for the shaft prostheses. Failure occurred in cases of poor bone quality and malpositioning. Cyclic motion of the short prosthesis was less than that for the shaft prostheses. Surface finish showed no effect. System stiffness for the new stem was lower than for the shaft prostheses.
INTERPRETATION: The new stem tended to migrate initially more than the shaft prostheses, but stabilised when cortical contact was achieved or the cancellous bone was compacted sufficiently. Bone quality and correct positioning were important factors for the short stem. The lower cyclic motion of the new stem should be favourable for bony ingrowth. The lower system bending stiffness with the new implant indicated a more physiological loading of the bone and should thereby reduce the effects of stress shielding.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16806616     DOI: 10.1016/j.clinbiomech.2006.04.004

Source DB:  PubMed          Journal:  Clin Biomech (Bristol, Avon)        ISSN: 0268-0033            Impact factor:   2.063


  16 in total

1.  [Short-stemmed endoprostheses in total hip arthroplasty].

Authors:  J Gulow; R Scholz; G Freiherr von Salis-Soglio
Journal:  Orthopade       Date:  2007-04       Impact factor: 1.087

2.  CORR Insights®: Long-term results and bone remodeling after THA with a short, metaphyseal-fitting anatomic cementless stem.

Authors:  Michael G Zywiel
Journal:  Clin Orthop Relat Res       Date:  2013-11-27       Impact factor: 4.176

3.  Long-term results and bone remodeling after THA with a short, metaphyseal-fitting anatomic cementless stem.

Authors:  Young-Hoo Kim; Jang-Won Park; Jun-Shik Kim; Jun-Seok Kang
Journal:  Clin Orthop Relat Res       Date:  2013-10-26       Impact factor: 4.176

4.  The influence of resection height on proximal femoral strain patterns after Metha short stem hip arthroplasty: an experimental study on composite femora.

Authors:  Thilo Floerkemeier; Jens Gronewold; Sebastian Berner; Gavin Olender; Christof Hurschler; Henning Windhagen; Gabriela von Lewinski
Journal:  Int Orthop       Date:  2012-12-28       Impact factor: 3.075

5.  Periprosthetic DXA after total hip arthroplasty with short vs. ultra-short custom-made femoral stems: 37 patients followed for 3 years.

Authors:  Carlina V Albanese; Francesco S Santori; Laura Pavan; Ian D Learmonth; Roberto Passariello
Journal:  Acta Orthop       Date:  2009-06       Impact factor: 3.717

Review 6.  [Endoprosthetic treatment of osteoporosis-related coxarthrosis : aspects of safe patient treatment].

Authors:  S Kirschner; A Hartmann; K-P Günther; C Hamann
Journal:  Orthopade       Date:  2014-04       Impact factor: 1.087

7.  A prospective randomized radiographic and dual-energy X-ray absorptiometric study of migration and bone remodeling after implantation of two modern short-stemmed femoral prostheses.

Authors:  Volker Brinkmann; Florian Radetzki; Karl Stefan Delank; David Wohlrab; Alexander Zeh
Journal:  J Orthop Traumatol       Date:  2015-02-11

8.  Migration analysis of a metaphyseal anchored short-stem hip prosthesis.

Authors:  Florian Schmidutz; Thomas Graf; Farhad Mazoochian; Andreas Fottner; Andrea Bauer-Melnyk; Volkmar Jansson
Journal:  Acta Orthop       Date:  2012-08       Impact factor: 3.717

9.  A new short uncemented, proximally fixed anatomic femoral implant with a prominent lateral flare: design rationals and study design of an international clinical trial.

Authors:  Tobias Renkawitz; Francesco S Santori; Joachim Grifka; Carlos Valverde; Michael M Morlock; Ian D Learmonth
Journal:  BMC Musculoskelet Disord       Date:  2008-11-04       Impact factor: 2.362

10.  The short stem GHEs in total hip replacement - experience after 380 implantations.

Authors:  Mohamed Ghanem; M Farag; P Schneider; P Hitzler; J Gulow; G Freiherr von Salis-Soglio
Journal:  GMS Interdiscip Plast Reconstr Surg DGPW       Date:  2013-11-19
View more