Literature DB >> 16720451

Intracochlear misdirected implantation of a cochlear implant.

R A Tange1, W Grolman, A Maat.   

Abstract

This paper discusses a rare complication of an intracochlear misdirection of the electrode of a cochlear implant in a 55-year-old male. The patient received a cochlear implant using the mastoid-saving surgical approach. Intraoperative measurements showed impedance and NRT reactions suggesting a reasonable function of the inner ear and the implant. Postoperatively our patient suffered from passing vertigo. Postoperative CT scans revealed a misdirection implantation of the cochlear implant into the vestibular part of the inner ear. A deformed implant was removed and a reimplantation was successfully performed after enlarging and reshaping the cochleostomy. Our patient now enjoys all the benefits of an optimal functioning cochlear implant. Intracochlear misdirection of the electrode can occur even when intraoperative measurements seem to be normal. When a patient suffers from unexplained vertigo after cochlear implantation with a poor function of the implant a misdirection of the active electrode must be considered. New high resolution peroperative three-dimensional imaging techniques can probably help to avoid such insertion failures.

Entities:  

Mesh:

Year:  2006        PMID: 16720451     DOI: 10.1080/00016480500445206

Source DB:  PubMed          Journal:  Acta Otolaryngol        ISSN: 0001-6489            Impact factor:   1.494


  12 in total

Review 1.  [Intra- and postoperative electrophysiological diagnostics].

Authors:  T Wesarg; S Arndt; A Aschendorff; R Laszig; R Beck; L Jung; S Zirn
Journal:  HNO       Date:  2017-04       Impact factor: 1.284

2.  Relationship of cochlea with surrounding neurovascular structures and their implication in cochlear implantation.

Authors:  Daisy Sahni; Anjali Singla; Ashok Gupta; Tulika Gupta; Anjali Aggarwal
Journal:  Surg Radiol Anat       Date:  2015-02-08       Impact factor: 1.246

3.  Extra-Cochlear Insertion in Cochlear Implantation: A Potentially Disastrous Condition.

Authors:  Emine Deniz Gözen; Fırat Tevetoğlu; Haydar Murat Yener; Eyüp Kara; Ahmet Ataş; Sabri Şirolu; Osman Kızılkılıç; Harun Cansız; Emin Karaman
Journal:  J Int Adv Otol       Date:  2019-12       Impact factor: 1.017

Review 4.  [Intraoperative audiological-technical diagnostics during cochlear implant surgery].

Authors:  T Wesarg; S Arndt; A Aschendorff; R Laszig; S Zirn
Journal:  HNO       Date:  2014-10       Impact factor: 1.284

5.  Complications of Paediatric Cochlear Implantation in the Population of Uttarakhand.

Authors:  V P Singh; Chetan Bansal; Tanvi Kanwar
Journal:  Indian J Otolaryngol Head Neck Surg       Date:  2019-01-21

6.  Electrophysiological detection of electrode fold-over in perimodiolar cochlear implant electrode arrays: a multi-center study case series.

Authors:  P Mittmann; G Lauer; A Ernst; S Mutze; F Hassepass; S Arndt; D Arweiler-Harbeck; F Christov
Journal:  Eur Arch Otorhinolaryngol       Date:  2019-09-24       Impact factor: 2.503

7.  Cochlear implant electrode misplacement: incidence, evaluation, and management.

Authors:  Yu-Lan Mary Ying; Jerry W Lin; John S Oghalai; Robert A Williamson
Journal:  Laryngoscope       Date:  2013-01-08       Impact factor: 3.325

8.  Incomplete and false tract insertions in cochlear implantation: retrospective review of surgical and auditory outcomes.

Authors:  Ashish Vashishth; Andrea Fulcheri; Maurizio Guida; Antonio Caruso; Mario Sanna
Journal:  Eur Arch Otorhinolaryngol       Date:  2018-03-05       Impact factor: 2.503

9.  Evaluation of 4 multisection CT systems in postoperative imaging of a cochlear implant: a human cadaver and phantom study.

Authors:  B M Verbist; R M S Joemai; W M Teeuwisse; W J H Veldkamp; J Geleijns; J H M Frijns
Journal:  AJNR Am J Neuroradiol       Date:  2008-05-15       Impact factor: 3.825

10.  Optical coherence tomography of the cochlea in the porcine model.

Authors:  Ali Sepehr; Hamid R Djalilian; Janice E Chang; Zhongping Chen; Brian J F Wong
Journal:  Laryngoscope       Date:  2008-08       Impact factor: 3.325

View more