Measurement error of lumbar total disc replacement range of motion.

Moe R Lim1, Randall T Loder, Russel C Huang, Stephen Lyman, Kai Zhang, Andrew Sama, Elias C Papadopoulos, Kristin Warner, Federico P Girardi, Frank P Cammisa.   

Abstract

STUDY
DESIGN: A retrospective review of lumbar total disc replacement (TDR) radiographs.
OBJECTIVE: To determine the error and variability in measuring TDR radiographic range of motion (ROM). SUMMARY OF BACKGROUND DATA: Motion preservation is the driving force behind lumbar TDR technology. In the recent literature, sagittal radiographic TDR ROM as low as 2 degrees has been reported. In these studies, ROM was determined by using the Cobb method to measure TDR sagittal alignment angles in flexion-extension lateral radiographs. However, previous studies in the spinal deformity literature have shown that the Cobb method is very susceptible to measurement error.
METHODS: There were 5 observers, including 2 attending orthopedic spine surgeons, 1 spine fellow, 1 fifth-year resident, and 1 fourth-year resident, who measured the ROM of 50 ProDisc II (Synthes Spine Solutions, New York, NY) TDRs on standard flexion-extension lumbar spine radiograph sets. Repeated measurements were made on 2 occasions using the Cobb method. Measurement variability was calculated using 3 statistical methods.
RESULTS: The 3 statistical methods resulted in extremely similar values for TDR ROM observer variability. Overall, the intraobserver variability of TDR ROM measurement was +/-4.6 degrees, and interobserver variability was +/-5.2 degrees .
CONCLUSIONS: To be 95% certain that an implanted TDR prosthesis has any sagittal motion, a ROM of at least 4.6 degrees must be observed, which is the upper limit of intraobserver measurement variability for a TDR with a true ROM of 0 degrees. To be 95% certain that a change in TDR ROM has occurred between 2 measurements by the same observer, a change in ROM of at least 9.6 degrees must be observed (the entire range of +/-4.6 degrees intraobserver variability). ROM measurement variability should be considered when evaluating the success or failure of motion preservation in lumbar TDR.

Entities:  

Mesh:

Year:  2006        PMID: 16648735     DOI: 10.1097/01.brs.0000216452.54421.ea

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  13 in total

1.  Total lumbar disc replacement in athletes: clinical results, return to sport and athletic performance.

Authors:  Christoph J Siepe; Karsten Wiechert; Mohamed F Khattab; Andreas Korge; H Michael Mayer
Journal:  Eur Spine J       Date:  2007-01-05       Impact factor: 3.134

Review 2.  Design concepts in lumbar total disc arthroplasty.

Authors:  Fabio Galbusera; Chiara M Bellini; Thomas Zweig; Stephen Ferguson; Manuela T Raimondi; Claudio Lamartina; Marco Brayda-Bruno; Maurizio Fornari
Journal:  Eur Spine J       Date:  2008-10-23       Impact factor: 3.134

3.  Radiographic total disc replacement angle measurement accuracy using the Oxford Cobbometer: precision and bias.

Authors:  Victor Kosmopoulos; Kosmas Stafylas; John McManus; Constantin Schizas
Journal:  Eur Spine J       Date:  2008-05-22       Impact factor: 3.134

4.  Consequences of patient position in the radiographic measurement of artificial disc replacement angles.

Authors:  Victor Kosmopoulos; John McManus; Constantin Schizas
Journal:  Eur Spine J       Date:  2007-09-11       Impact factor: 3.134

Review 5.  Do in vivo kinematic studies provide insight into adjacent segment degeneration? A qualitative systematic literature review.

Authors:  Masoud Malakoutian; David Volkheimer; John Street; Marcel F Dvorak; Hans-Joachim Wilke; Thomas R Oxland
Journal:  Eur Spine J       Date:  2015-06-09       Impact factor: 3.134

Review 6.  [Adjacent segment movement after monosegmental total disc replacement and monosegmental fusion of segments L4/5].

Authors:  M Däxle; T Kocak; F Lattig; H Reichel; B Cakir
Journal:  Orthopade       Date:  2013-02       Impact factor: 1.087

7.  Radiographic measurement of the sagittal plane deformity in patients with osteoporotic spinal fractures evaluation of intrinsic error.

Authors:  Ahmet Alanay; Murat Pekmezci; Oguz Karaeminogullari; Emre Acaroglu; Muharrem Yazici; Akin Cil; Bas Pijnenburg; Yasemin Genç; Fethullah C Oner
Journal:  Eur Spine J       Date:  2007-10-03       Impact factor: 3.134

8.  We Need to Talk about Lumbar Total Disc Replacement.

Authors:  Stephen Beatty
Journal:  Int J Spine Surg       Date:  2018-08-03

9.  The LP-ESP(®) lumbar disc prosthesis with 6 degrees of freedom: development and 7 years of clinical experience.

Authors:  Jean-Yves Lazennec; Alain Aaron; Adrien Brusson; Jean-Patrick Rakover; Marc-Antoine Rousseau
Journal:  Eur J Orthop Surg Traumatol       Date:  2013-01-11

10.  Validation of Pre-operative Templating for Total Disc Replacement Surgery.

Authors:  Justin F M Hollenbeck; Jill A Fattor; Vikas Patel; Evalina Burger; Paul J Rullkoetter; Christopher M J Cain
Journal:  Int J Spine Surg       Date:  2019-02-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.