OBJECTIVES: To describe our preclinical experience with Cone Beam CT (CBCT) in image-guided surgery of the temporal bone. STUDY DESIGN AND SETTINGS: A mobile isocentric C-arm (PowerMobil, Siemens Medical Systems, Erlangen, Germany) modified to include a flat-panel detector (Varian Imaging Products, Palo Alto, CA) and a motorized orbit was developed to acquire multiple projections in rotation about a subject. Initial experiments imaging steel wire in air were used to investigate the system's spatial resolution in 3D image reconstruction. Subsequently temporal bone dissection was performed on five cadaver heads using the modified C-arm as an image guidance system.
RESULTS: We obtained a spatial resolution of 0.85 mm. The image acquisition time was 120 seconds and the radiation dose approximately one-tenth of a conventional CT scan.
CONCLUSION: CBCT provided submillimeter accuracy at high speed with low radiation dosage to offer utility as an intraoperative imaging system. SIGNIFICANCE: CBCT offers technology that approximates "near-real-time" image guidance. EBM RATING: C-4.

Entities:  

Mesh:

Year:  2006        PMID: 16647538     DOI: 10.1016/j.otohns.2005.12.007

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   3.497


  19 in total

1.  Device setting modifications for 3D flatpanel imaging in skull base surgery.

Authors:  Frederike Hassepass; Wolfgang Maier; Antje Aschendorff; Stefan Bulla; Werner Vach; Roland Laszig; Tanja D Grauvogel
Journal:  Eur Arch Otorhinolaryngol       Date:  2012-04-06       Impact factor: 2.503

2.  Cone beam CT: a current overview of devices.

Authors:  A Nemtoi; C Czink; D Haba; A Gahleitner
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

3.  Automatic image-to-world registration based on x-ray projections in cone-beam CT-guided interventions.

Authors:  N M Hamming; M J Daly; J C Irish; J H Siewerdsen
Journal:  Med Phys       Date:  2009-05       Impact factor: 4.071

4.  Geometric calibration of a mobile C-arm for intraoperative cone-beam CT.

Authors:  M J Daly; J H Siewerdsen; Y B Cho; D A Jaffray; J C Irish
Journal:  Med Phys       Date:  2008-05       Impact factor: 4.071

Review 5.  [Function-retaining reconstruction after orbital trauma].

Authors:  M C Metzger; W A Lagrèze; R Schön
Journal:  Ophthalmologe       Date:  2011-06       Impact factor: 1.059

6.  Comparison of Intraoperative Portable CT Scanners in Skull Base and Endoscopic Sinus Surgery: Single Center Case Series.

Authors:  David B Conley; Bruce Tan; Bernard R Bendok; H Hunt Batjer; Rakesh Chandra; Douglas Sidle; Rudy J Rahme; Joseph G Adel; Andrew J Fishman
Journal:  Skull Base       Date:  2011-07

7.  Soft-tissue imaging with C-arm cone-beam CT using statistical reconstruction.

Authors:  Adam S Wang; J Webster Stayman; Yoshito Otake; Gerhard Kleinszig; Sebastian Vogt; Gary L Gallia; A Jay Khanna; Jeffrey H Siewerdsen
Journal:  Phys Med Biol       Date:  2014-02-07       Impact factor: 3.609

8.  Atlas-Based Segmentation of Temporal Bone Anatomy.

Authors:  Kimerly A Powell; Tong Liang; Brad Hittle; Don Stredney; Thomas Kerwin; Gregory J Wiet
Journal:  Int J Comput Assist Radiol Surg       Date:  2017-08-29       Impact factor: 2.924

9.  Demons deformable registration for CBCT-guided procedures in the head and neck: convergence and accuracy.

Authors:  S Nithiananthan; K K Brock; M J Daly; H Chan; J C Irish; J H Siewerdsen
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

10.  Intraoperative use of cone-beam computed tomography in a cadaveric ossified cochlea model.

Authors:  Emma Barker; Keith Trimble; Harley Chan; James Ramsden; Sajendra Nithiananthan; Adrian James; Gideon Bachar; Mike Daly; Jonathan Irish; Jeff Siewerdsen
Journal:  Otolaryngol Head Neck Surg       Date:  2009-05       Impact factor: 3.497

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.