Corrosion of stainless steel sternal wire after long-term implantation.

Yasuko Tomizawa1, Takao Hanawa, Daisuke Kuroda, Hiroshi Nishida, Masahiro Endo.   

Abstract

A variety of metallic components have been used in medical devices where lifelong durability and physical strength are demanded. To investigate the in vivo changes of implanted metallic medical devices in humans, stainless steel sternal wires removed from patients were evaluated. Stainless steel (316L) sternal wires removed from four patients after 10, 13, 22, and 30 years of implantation were evaluated using scanning electron microscopy (SEM) and energy dispersive X-ray spectroscopy (EDS). Macroscopically, the removed specimens maintained their metallic luster and color. Under SEM, small holes were observed sporadically at 10 years and they tended to connect in the drawing direction. The longer the implanted duration, the more numerous and deeper were the crevices observed. By EDS, sulfur, phosphorus, and calcium were identified in all areas at 10 years, in addition to the component elements of stainless steel, comprising iron, chromium, nickel, and manganese. Corrosion products observed at 30 years were identified as calcium phosphate. In conclusion, stainless steel sternal wires develop corroded pores that grow larger and deeper with time after implantation; however, the pores remain shallow even after decades of implantation and they may not be a cause of mechanical failure. An amount of metal ions equivalent to the corroded volume must have been released into the human body, but the effect of these metal ions on the body is not apparent.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16614803     DOI: 10.1007/s10047-005-0321-0

Source DB:  PubMed          Journal:  J Artif Organs        ISSN: 1434-7229            Impact factor:   1.731


  21 in total

1.  Allergic contact dermatitis and osteomyelitis due to sternal stainless steel wire.

Authors:  K Hayashi; H Kaneko; S Kawachi; T Saida
Journal:  Contact Dermatitis       Date:  1999-08       Impact factor: 6.600

2.  Pacemaker dermatitis from titanium.

Authors:  R Yamauchi; A Morita; T Tsuji
Journal:  Contact Dermatitis       Date:  2000-01       Impact factor: 6.600

3.  Taxi driver's occupational allergic contact dermatitis from nickel in euro coins.

Authors:  Javier Sánchez-Pérez; Diana Ruíz-Genao; Irene García Del Río; Amaro García Diez
Journal:  Contact Dermatitis       Date:  2003-06       Impact factor: 6.600

4.  Delayed chest wall pain due to sternal wire sutures.

Authors:  C E Eastridge; S S Mahfood; W A Walker; F H Cole
Journal:  Ann Thorac Surg       Date:  1991-01       Impact factor: 4.330

5.  Pacemaker contact sensitivity.

Authors:  M S Peters; A L Schroeter; H M van Hale; J C Broadbent
Journal:  Contact Dermatitis       Date:  1984-10       Impact factor: 6.600

6.  Evaluation of metal allergies in patients with coronary stents.

Authors:  U Hillen; M Haude; R Erbel; M Goos
Journal:  Contact Dermatitis       Date:  2002-12       Impact factor: 6.600

7.  Pyogenic granuloma caused by exposed sternotomy wires.

Authors:  Marian Elizabeth Northington; Conway C Huang
Journal:  Dermatol Surg       Date:  2004-12       Impact factor: 3.398

8.  Sternal wire-induced persistent chest pain: a possible hypersensitivity reaction.

Authors:  P G Fine; S V Karwande
Journal:  Ann Thorac Surg       Date:  1990-01       Impact factor: 4.330

9.  Nickel sensitivity in the general population.

Authors:  L Peltonen
Journal:  Contact Dermatitis       Date:  1979-01       Impact factor: 6.600

10.  Severe ostial saphenous vein graft disease leading to acute coronary syndromes following proximal aorto-saphenous anastomoses with the symmetry bypass connector device: is it a suture device or a "stent"?

Authors:  Jeffrey J Cavendish; William F Penny; Michael M Madani; Shahin Keramati; Ori Ben-Yehuda; Daniel G Blanchard; Ehtisham Mahmud; Anthony Perricone; Sotirios Tsimikas
Journal:  J Am Coll Cardiol       Date:  2004-01-07       Impact factor: 24.094

View more
  7 in total

1.  Recovery of motor deficit, cerebellar serotonin and lipid peroxidation levels in the cortex of injured rats.

Authors:  Antonio Bueno-Nava; Rigoberto Gonzalez-Pina; Alfonso Alfaro-Rodriguez; Vladimir Nekrassov-Protasova; Alfredo Durand-Rivera; Sergio Montes; Fructuoso Ayala-Guerrero
Journal:  Neurochem Res       Date:  2010-06-10       Impact factor: 3.996

Review 2.  Atrial septum defect closure device in a beating heart, from the perspective of a researcher in artificial organs.

Authors:  Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2012-06-23       Impact factor: 1.731

3.  Journal of Artificial Organs 2006: the year in review.

Authors:  Y Sawa; T Horiuchi; A Kishida; T Masuzawa; M Nishimura; E Tatsumi; Y Tomizawa; H Watanabe
Journal:  J Artif Organs       Date:  2007-06-20       Impact factor: 1.731

Review 4.  Materials for metallic stents.

Authors:  Takao Hanawa
Journal:  J Artif Organs       Date:  2009-06-18       Impact factor: 1.731

5.  In vivo evaluation of Zr-based bulk metallic glass alloy intramedullary nails in rat femora.

Authors:  Kazuhiro Imai; Sachiko Hiromoto
Journal:  J Mater Sci Mater Med       Date:  2013-11-27       Impact factor: 3.896

Review 6.  Intravascular foreign bodies: danger of unretrieved fragmented medical devices.

Authors:  Minori Tateishi; Yasuko Tomizawa
Journal:  J Artif Organs       Date:  2009-06-18       Impact factor: 1.731

Review 7.  Biocompatibility of titanium from the viewpoint of its surface.

Authors:  Takao Hanawa
Journal:  Sci Technol Adv Mater       Date:  2022-08-15       Impact factor: 7.821

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.