Severe ostial saphenous vein graft disease leading to acute coronary syndromes following proximal aorto-saphenous anastomoses with the symmetry bypass connector device: is it a suture device or a "stent"?

Jeffrey J Cavendish1, William F Penny, Michael M Madani, Shahin Keramati, Ori Ben-Yehuda, Daniel G Blanchard, Ehtisham Mahmud, Anthony Perricone, Sotirios Tsimikas.   

Abstract

The Symmetry Bypass Connector (St. Jude Medical, St. Paul, Minnesota) is a nitinol, star-shaped device that was designed to facilitate placement of sutureless aorto-saphenous anastomoses during off-pump coronary artery bypass graft surgery (CABG). Although the device is approved for clinical use in Europe and the U.S., its short- and long-term safety and efficacy are not established. We report on 5 of 121 patients undergoing CABG who presented with an acute coronary syndrome two to five months following placement of this device. In each patient, all saphenous vein grafts (SVGs) placed (n = 11) with the device were totally occluded (n = 6) or compromised by ostial stenoses (n = 5). Treatment consisted of repeat CABG in one patient and percutaneous coronary intervention (PCI) in four patients with cutting balloon atherotomy and stenting. Following PCI, two of four patients presented again within two months with near-occlusive ostial restenosis in all stents placed. Intracoronary ultrasound showed severe neointimal hyperplasia, but only at the proximal interface of the device and stent. One patient was treated with brachytherapy in two SVGs but had a recurrence four months later and was treated with drug-eluting stents in both restenotic segments. Recalcitrant neointimal hyperplasia is postulated to be involved in the pathogenesis of anastomotic device stenosis, possibly similar to in-stent restenosis. Prospective randomized clinical trials are needed to assess the clinical safety and efficacy of this device. Pending such studies, consideration should be given in limiting its use to cases of unacceptably high risk of stroke during aortic cross-clamping. Dual antiplatelet agents, evaluation for ischemia, and close follow-up are warranted in patients that have already received the device.

Entities:  

Mesh:

Year:  2004        PMID: 14715195     DOI: 10.1016/j.jacc.2003.09.028

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  5 in total

1.  How long will the heart still beat?

Authors:  F L Wellens
Journal:  Tex Heart Inst J       Date:  2005

2.  Study of coronary artery bypass using the PAS-Port device: assessment by multidetector computed tomography.

Authors:  Takeshiro Fujii; Yoshinori Watanabe; Noritsugu Shiono; Tsukasa Ozawa; Satoshi Hamada; Hiroshi Masuhara; Chikao Teramoto; Masanori Hara; Nobuya Koyama
Journal:  Gen Thorac Cardiovasc Surg       Date:  2009-02-12

3.  Combined aortic valve replacement and coronary artery bypass grafting for a calcified ascending aorta.

Authors:  Hironori Baba; Masayoshi Umesue; Kanzi Matsui
Journal:  Gen Thorac Cardiovasc Surg       Date:  2012-04

4.  Corrosion of stainless steel sternal wire after long-term implantation.

Authors:  Yasuko Tomizawa; Takao Hanawa; Daisuke Kuroda; Hiroshi Nishida; Masahiro Endo
Journal:  J Artif Organs       Date:  2006       Impact factor: 1.731

5.  Safe introduction and quality control of new methods in coronary surgery.

Authors:  Jacob Bergsland
Journal:  Acta Inform Med       Date:  2011-12
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.