Literature DB >> 11156123

SMA circuits reduce platelet consumption and platelet factor release during cardiac surgery.

J O Defraigne1, J Pincemail, G Dekoster, R Larbuisson, M Dujardin, F Blaffart, J L David, R Limet.   

Abstract

BACKGROUND: Platelet count and function are particularly damaged by cardiopulmonary bypass (CPB). This study evaluated the effects of a novel CPB circuit in terms of platelet count and activation, and postoperative need for blood products.
METHODS: One hundred patients undergoing coronary grafting were randomized in two groups: control group (n = 50) and test group (n = 50, surface modifying additives circuit, SMA group). Blood samples were taken before, during, and after CPB. Postoperative blood loss, number of transfused blood products, and postoperative variables were recorded.
RESULTS: The platelet count decreased less in the SMA group compared to the control group (end of CPB: respectively, 165 +/- 9 x 10(3)/mm3 vs 137 +/- 8 x 10(3)/mm3; p < 0.01). This was paralleled by a reduction in beta-thromboglobulin plasma levels in the SMA group. There was a trend to decreased blood loss in the SMA group, but the difference was significant only in patients taking aspirin preoperatively (p < 0.05). In the SMA group nearly 50% less fresh frozen plasma and platelet units were administered (p < 0.01). No operative deaths were observed.
CONCLUSIONS: The use of circuits with surface additives is clinically safe, preserves platelet levels, and attenuates platelet activation. This may lead to a reduced need for blood products.

Entities:  

Mesh:

Substances:

Year:  2000        PMID: 11156123     DOI: 10.1016/s0003-4975(00)01838-5

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  3 in total

1.  In vivo comparison study of FDA-approved surface-modifying additives and poly-2-methoxyethylacrylate circuit surfaces coatings during cardiopulmonary bypass.

Authors:  Angela Ask; David Holt; Lynette Smith
Journal:  J Extra Corpor Technol       Date:  2006-03

2.  SMART tubing presents an increased risk of disconnection during extracorporeal circulation.

Authors:  Ross Newling; Richard Morris
Journal:  J Extra Corpor Technol       Date:  2005-12

3.  Toward an artificial endothelium: Development of blood-compatible surfaces for extracorporeal life support.

Authors:  Teryn R Roberts; Mark R S Garren; Hitesh Handa; Andriy I Batchinsky
Journal:  J Trauma Acute Care Surg       Date:  2020-08       Impact factor: 3.697

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.