Literature DB >> 16517772

Problems encountered during and after stent-graft treatment of aortic dissection.

Jong Yun Won1, Sang-Hyun Suh, Heung-Kyu Ko, Kwang Hoon Lee, Won Heum Shim, Byung-Chul Chang, Dong Hoon Choi, Sang Joon Park, Do Yun Lee.   

Abstract

PURPOSE: The authors report their experiences with the problems encountered during and after stent-graft placement in patients with aortic dissections.
MATERIALS AND METHODS: Between July 1994 and December 2003, stent-graft placement was performed in 60 patients with aortic dissections, 9 patients with Stanford type A dissection and 51 patients with type B. Ten patients had acute dissection and 50 patients had chronic dissection. Each patient was followed for 12 to 107 months (mean, 33 mo). All cases were retrospectively reviewed for any problem that occurred during and after the procedure.
RESULTS: During the procedure, there were two cases (3%) of stent-graft migration, one case (2%) of stent-graft torsion, two cases (3%) of stent-graft folding due to oversizing, three cases (5%) of persistent thoracic false lumen flow by the appearance of a hidden intimal tear, one case (2%) of new intimal tear that resulted in retrograde type A dissection, and 12 cases (20%) of type I endoleaks on aortograms taken immediately after the procedure. During the follow-up period, one case (2%) of transient cerebral ischemia, six cases (10%) of persistent type I endoleaks, two cases (3%) of type II endoleaks through the intercostal artery, four cases (7%) with progressive dilatation of abdominal false lumen, six cases (10%) of new intimal tears which resulted in saccular aneurysms (n=4) or new dissections (n=2) on either or both ends of the stent-graft, and two cases (3%) of mechanical failure were observed. Overall, five patients (8%) required surgical conversion, and there were no cases of procedure-related mortality.
CONCLUSION: Various problems can occur during and after stent-graft placement in aortic dissection. The thorough evaluation of preoperative imaging and close follow-up are mandatory to optimize the management of such problems.

Entities:  

Mesh:

Substances:

Year:  2006        PMID: 16517772     DOI: 10.1097/01.RVI.0000195141.98163.30

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  9 in total

1.  Unusual complications of endovascular repair of the thoracic aorta: MDCT findings.

Authors:  T Valente; G Rossi; F Lassandro; G Rea; M Marino; G Dialetto; R Muto; M Scaglione
Journal:  Radiol Med       Date:  2012-01-07       Impact factor: 3.469

2.  A computational study on the biomechanical factors related to stent-graft models in the thoracic aorta.

Authors:  S K Lam; George S K Fung; Stephen W K Cheng; K W Chow
Journal:  Med Biol Eng Comput       Date:  2008-07-11       Impact factor: 2.602

3.  Risk factors for distal stent graft-induced new entry following endovascular repair of type B aortic dissection.

Authors:  Qing Li; Long-Fei Wang; Wei-Guo Ma; Shang-Dong Xu; Jun Zheng; Xiao-Yan Xing; Lian-Jun Huang; Li-Zhong Sun
Journal:  J Thorac Dis       Date:  2015-11       Impact factor: 2.895

Review 4.  Acute complicated and uncomplicated type III aortic dissection: an endovascular perspective.

Authors:  Castigliano M Bhamidipati; Gorav Ailawadi
Journal:  Semin Thorac Cardiovasc Surg       Date:  2009

5.  Endovascular management for ruptured Stanford B acute aortic dissection.

Authors:  Atsushi Aoki; Takanori Suezawa; Kenji Sangawa; Mamoru Tago
Journal:  Gen Thorac Cardiovasc Surg       Date:  2011-02-10

6.  Biomechanical Material Characterization of Stanford Type-B Dissected Porcine Aortas.

Authors:  Aashish Ahuja; Jillian N Noblet; Tony Trudnowski; Bhavesh Patel; Joshua F Krieger; Sean Chambers; Ghassan S Kassab
Journal:  Front Physiol       Date:  2018-09-26       Impact factor: 4.566

7.  Ulcer-like projections into the dilated false lumen after stent-graft placement for aortic dissection: A case report.

Authors:  Xi He; Eijun Sueyoshi; Shun Nakaji; Masataka Uetani
Journal:  Medicine (Baltimore)       Date:  2022-01-07       Impact factor: 1.889

8.  Infolding of fenestrated endovascular stent graft.

Authors:  Jason G E Zelt; Prasad Jetty; Adnan Hadziomerovic; Sudhir Nagpal
Journal:  J Vasc Surg Cases Innov Tech       Date:  2017-07-20

9.  Validated Computational Model to Compute Re-apposition Pressures for Treating Type-B Aortic Dissections.

Authors:  Aashish Ahuja; Xiaomei Guo; Jillian N Noblet; Joshua F Krieger; Blayne Roeder; Stephan Haulon; Sean Chambers; Ghassan S Kassab
Journal:  Front Physiol       Date:  2018-05-09       Impact factor: 4.566

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.