Literature DB >> 16516730

Biomechanical stability of a fixed-angle volar plate versus fragment-specific fixation system: cyclic testing in a C2-type distal radius cadaver fracture model.

Kenneth F Taylor1, Brent G Parks, Keith A Segalman.   

Abstract

PURPOSE: To compare the biomechanical stability of 2 recently introduced fixation systems in an intra-articular, dorsal comminution distal radius fracture model.
METHODS: AO/ASIF type C2 fractures were simulated in 10 matched pairs of fresh-frozen cadaveric arms randomized between fixed-angle volar plate and fragment-specific fixation systems. Specimens were loaded in extension cyclically for 2,000 repetitions followed by a single cycle to failure. Initial, intermediate, and final stiffness values and failure load values were obtained and compared.
RESULTS: Both systems were able to sustain physiologic cyclic loading. The fragment-specific system was significantly stiffer than the fixed-angle volar plate system for the ulnar segment in both the precycle and postcycle values. No other comparisons were significant with respect to stiffness. No significant difference in load to failure was found between the systems with respect to ulnar, radial, or overall fragment displacement.
CONCLUSIONS: Both fixed-angle volar plate and fragment-specific fixation systems performed comparably in a simulated early postoperative motion protocol. Fragment-specific fixation had improved stiffness characteristics only with respect to the smaller ulnar-sided fragment.

Entities:  

Mesh:

Year:  2006        PMID: 16516730     DOI: 10.1016/j.jhsa.2005.12.017

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  10 in total

1.  Efficacy of radial styloid targeting screws in volar plate fixation of intra-articular distal radial fractures: a biomechanical study in a cadaver fracture model.

Authors:  Kousuke Iba; Yasuhiro Ozasa; Takuro Wada; Tomoaki Kamiya; Toshihiko Yamashita; Mitsuhiro Aoki
Journal:  J Orthop Surg Res       Date:  2010-12-02       Impact factor: 2.359

2.  Radial Plate Fixation: A Novel Technique for Distal Radius Fractures.

Authors:  Jeffrey D Hoffmann; Jeremy Stewart; Nicholas Kusnezov; John Dunn; Miguel Pirela-Cruz
Journal:  Hand (N Y)       Date:  2016-10-22

3.  Efficiency of fragment specific fixation plates in the treatment of comminuted distal radial fractures.

Authors:  Elhadi Musaab Elhadi Babikir; Eyad Daood Al-Maqdassy; Khalid Hasan; Shamsi Abdul Hameed; Abduljabbar Alhammoud; Mohammed Mubarak Al-Ateeq Al-Dosari
Journal:  Int Orthop       Date:  2017-06-27       Impact factor: 3.075

4.  Effect of Metacarpal Fixation Configuration on Strength and Fracture Gap Displacements in a Cadaveric Model of Externally Fixed Distal Radius Fractures.

Authors:  Kenneth F Taylor; Jyoti Sharma; Neena A Davisson; Evan P Roush; Christine M Jones; Gregory S Lewis
Journal:  Hand (N Y)       Date:  2019-06-10

5.  Wound Complications with Multiple Wrist Incisions in Distal Radius Reconstruction.

Authors:  Joshua J Meaike; Christian Athens; Nicole Sgromolo; Alexander Y Shin; Peter C Rhee
Journal:  J Wrist Surg       Date:  2021-10-31

6.  A comparative study of fragment-specific versus volar plate fixation of distal radius fractures.

Authors:  Douglas M Sammer; Douglas S Fuller; Hyungjin Myra Kim; Kevin C Chung
Journal:  Plast Reconstr Surg       Date:  2008-11       Impact factor: 4.730

7.  Development and Testing of a Novel Locking Pin Cap to Create a Fixed-Angle K-Wire Plate Construct.

Authors:  David M Brogan; Simon Tang; Kaitlyn Broz; Christopher J Dy
Journal:  J Hand Surg Am       Date:  2020-12-25       Impact factor: 2.230

8.  SPRING PLATES IN DISTAL RADIO FRACTURES: "IN VITRO" MECHANICAL PROPERTIES.

Authors:  Ana Lécia Carneiro Leão de Araújo Lima; Alex Eduardo Calderon Irusta; Alexandre Martins Portelinha; Lauro Tofollo; Antonio Carlos Shimano; Amanda Favaro Cagnolati; Nilton Mazzer; Claudio Henrique Barbieri
Journal:  Acta Ortop Bras       Date:  2018       Impact factor: 0.513

9.  Evaluation of the mechanical properties and clinical application of nickel-titanium shape memory alloy scaphoid arc nail.

Authors:  Muguo Song; Yongyue Su; Chuan Li; Yongqing Xu
Journal:  Eng Life Sci       Date:  2021-03-09       Impact factor: 2.678

10.  DISTAL RADIAL FRACTURES IN PATIENTS OVER 60 YEARS OLD: ORTHOGONAL PLATES VERSUS VOLAR PLATE.

Authors:  Edgard Novaes de França Bisneto; Emygdio José Leomil de Paula; Marcelo Rosa de Resende; Rames Mattar Júnior; Arnaldo Valdir Zumiotti
Journal:  Rev Bras Ortop       Date:  2015-11-16
  10 in total