Literature DB >> 33358883

Development and Testing of a Novel Locking Pin Cap to Create a Fixed-Angle K-Wire Plate Construct.

David M Brogan1, Simon Tang2, Kaitlyn Broz3, Christopher J Dy4.   

Abstract

PURPOSE: To test the effectiveness of a novel locking pin cap to attach a K-wire rigidly to a volar locking plate and resist fracture displacement compared with commercially available alternatives.
METHODS: Two different methods of fracture fixation were tested on a total of 12 Sawbones models with volar shear distal radius fracture (6/group). The fragments were fixed with either 2 commercially available pin plates (industry standard) or a volar plate with 2 locking screws fixing the scaphoid facet and 2 pins locked to the plate with a novel locking pin cap in the lunate facet. Axial load conditioning was performed followed by sinusoidal loading to 250 N at 50 mm/s. A motion capture system was used to assess the relative movement of the fracture fragments relative to the intact shaft. The strength of the fixation construct was quantified by (1) the force required to achieve a 2-mm gap between the shaft and fracture fragments and (2) ultimate load to failure.
RESULTS: One industry standard pin plate demonstrated disassociation of the pin from the plate after fatigue conditioning. This did not occur in the locking pin cap group. The locking pin cap construct group was able to sustain a significantly higher load compared with the industry standard when the construct was displaced to the 2-mm gap. The locking pin cap also significantly increased the ultimate load to failure compared with the industry standard.
CONCLUSIONS: The novel locking pin cap creates a fixed-angle attachment of a K-wire to an existing locking screw hole in a plate. This fixed-angle K-wire is significantly stronger in preventing gap formation and resisting ultimate failure than commercially available plates that use bent K-wires. CLINICAL RELEVANCE: The development of novel techniques to secure small articular fragments may ultimately improve clinical outcomes.
Copyright © 2021 American Society for Surgery of the Hand. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Biomechanical testing; locking pin cap

Mesh:

Year:  2020        PMID: 33358883      PMCID: PMC8693392          DOI: 10.1016/j.jhsa.2020.11.001

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  17 in total

1.  The effects of screw length on stability of simulated osteoporotic distal radius fractures fixed with volar locking plates.

Authors:  Lindley B Wall; Michael D Brodt; Matthew J Silva; Martin I Boyer; Ryan P Calfee
Journal:  J Hand Surg Am       Date:  2012-02-02       Impact factor: 2.230

2.  The operative treatment of fractures of the distal radius is increasing: results from a nationwide Swedish study.

Authors:  C Mellstrand-Navarro; H J Pettersson; H Tornqvist; S Ponzer
Journal:  Bone Joint J       Date:  2014-07       Impact factor: 5.082

3.  Volar fixation for dorsally displaced fractures of the distal radius: a preliminary report.

Authors:  Jorge L Orbay; Diego L Fernandez
Journal:  J Hand Surg Am       Date:  2002-03       Impact factor: 2.230

4.  Fracture of the distal radius: risk factors for complications after locked volar plate fixation.

Authors:  Maximillian Soong; Roderick van Leerdam; Thierry G Guitton; Christopher Got; Julia Katarincic; David Ring
Journal:  J Hand Surg Am       Date:  2011-01       Impact factor: 2.230

5.  Volar Marginal Rim Fracture Fixation With Volar Fragment-Specific Hook Plate Fixation.

Authors:  Maureen A O'Shaughnessy; Alexander Y Shin; Sanjeev Kakar
Journal:  J Hand Surg Am       Date:  2015-06-10       Impact factor: 2.230

6.  Intrafocal (Kapandji) pinning of distal radius fractures with and without external fixation.

Authors:  T E Trumble; W Wagner; D P Hanel; N B Vedder; M Gilbert
Journal:  J Hand Surg Am       Date:  1998-05       Impact factor: 2.230

7.  Complications following internal fixation of unstable distal radius fracture with a palmar locking-plate.

Authors:  Rohit Arora; Martin Lutz; Alfred Hennerbichler; Dietmar Krappinger; David Espen; Markus Gabl
Journal:  J Orthop Trauma       Date:  2007-05       Impact factor: 2.512

8.  Loss of fixation of the volar lunate facet fragment in fractures of the distal part of the radius.

Authors:  Neil G Harness; Jesse B Jupiter; Jorge L Orbay; Keith B Raskin; Diego L Fernandez
Journal:  J Bone Joint Surg Am       Date:  2004-09       Impact factor: 5.284

9.  Bridging external fixation and supplementary Kirschner-wire fixation versus volar locked plating for unstable fractures of the distal radius: a randomised, prospective trial.

Authors:  K Egol; M Walsh; N Tejwani; T McLaurin; C Wynn; N Paksima
Journal:  J Bone Joint Surg Br       Date:  2008-09

10.  Distal radius fractures and the volar lunate facet fragment: Kirschner wire fixation in addition to volar-locked plating.

Authors:  Amy M Moore; David G Dennison
Journal:  Hand (N Y)       Date:  2014-06
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.