Literature DB >> 21122156

Efficacy of radial styloid targeting screws in volar plate fixation of intra-articular distal radial fractures: a biomechanical study in a cadaver fracture model.

Kousuke Iba1, Yasuhiro Ozasa, Takuro Wada, Tomoaki Kamiya, Toshihiko Yamashita, Mitsuhiro Aoki.   

Abstract

BACKGROUND: The locking screws target the radial styloid, theoretically provide greater stability against radial styloid fragment. However, it is unknown whether the radial styloid locking screws increased the stability of the volar plating system fixation along the entire distal radius or not. In this study, we evaluated the stability of the volar plating system fixation with or without the radial styloid screws using a biomechanical study in a cadaver fracture model.
METHODS: Six matched pairs of fresh-frozen human cadaver wrists complete from the proximal forearm to the metacarpal bones were prepared to simulate standardized 3-part intra-articular and severe comminuted fractures. Specimens were fixed using the volar plating system with or without 2 radial styloid screws. Each specimen was loaded at a constant rate of 20 mm/min to failure. Load data was recorded and, ultimate strength and change in gap between distal and proximal fragments were measured. Data for ultimate strength and screw failure after failure loading were compared between the 2 groups.
RESULTS: The average ultimate strength at failure of the volar plate fixation with radial styloid screws (913.5 ± 157.1 N) was significantly higher than that without them (682.2 ± 118.6 N). After failure loading, the average change in gap between the ulnar and proximal fragment was greater than that between the radial and proximal fragment. The number of bent or broken screws in ulnar fragment was higher than that in radial fragment. The number of specimens with bent or broken screws in cases with radial styloid screws was fewer than that in the fixation without radial styloid screws group.
CONCLUSION: The ulnar fragment is more intensively stressed than the radial fragment under axial loading of distal radius at full wrist extension. The radial styloid screws were effective in stable volar plate fixation of distal radial fractures.

Entities:  

Year:  2010        PMID: 21122156      PMCID: PMC3014894          DOI: 10.1186/1749-799X-5-90

Source DB:  PubMed          Journal:  J Orthop Surg Res        ISSN: 1749-799X            Impact factor:   2.359


  12 in total

1.  Volar fixed-angle plating of distal radius extension fractures: influence of plate position on secondary loss of reduction--a biomechanic study in a cadaveric model.

Authors:  Herwig Drobetz; Adam L Bryant; Tom Pokorny; Ralf Spitaler; Martin Leixnering; Jesse B Jupiter
Journal:  J Hand Surg Am       Date:  2006-04       Impact factor: 2.230

2.  Force distribution across wrist joint: application of pressure-sensitive conductive rubber.

Authors:  T Hara; E Horii; K N An; W P Cooney; R L Linscheid; E Y Chao
Journal:  J Hand Surg Am       Date:  1992-03       Impact factor: 2.230

3.  Treatment of unstable distal radial fractures with the volar locking plating system.

Authors:  Kevin C Chung; Andrew J Watt; Sandra V Kotsis; Zvi Margaliot; Steven C Haase; H Myra Kim
Journal:  J Bone Joint Surg Am       Date:  2006-12       Impact factor: 5.284

4.  Palmar plating for dorsally displaced fractures of the distal radius.

Authors:  Masayuki Kamano; Yoshinobu Honda; Kenichi Kazuki; Masataka Yasuda
Journal:  Clin Orthop Relat Res       Date:  2002-04       Impact factor: 4.176

5.  Palmar locking plate for treatment of unstable dorsal dislocated distal radius fractures.

Authors:  R Arora; M Lutz; D Fritz; R Zimmermann; J Oberladstätter; M Gabl
Journal:  Arch Orthop Trauma Surg       Date:  2005-05-13       Impact factor: 3.067

Review 6.  Biomechanics and biology of plate fixation of distal radius fractures.

Authors:  Alan E Freeland; Kurre T Luber
Journal:  Hand Clin       Date:  2005-08       Impact factor: 1.907

7.  Biomechanics in uniaxial compression of three distal radius volar plates.

Authors:  Denju Osada; Satoshi Fujita; Kazuya Tamai; Akira Iwamoto; Kazuo Tomizawa; Koichi Saotome
Journal:  J Hand Surg Am       Date:  2004-05       Impact factor: 2.230

8.  Prospective study of distal radius fractures treated with a volar locking plate system.

Authors:  Denju Osada; Shuzo Kamei; Koichiro Masuzaki; Morimitsu Takai; Masahiro Kameda; Kazuya Tamai
Journal:  J Hand Surg Am       Date:  2008 May-Jun       Impact factor: 2.230

9.  Biomechanical stability of a fixed-angle volar plate versus fragment-specific fixation system: cyclic testing in a C2-type distal radius cadaver fracture model.

Authors:  Kenneth F Taylor; Brent G Parks; Keith A Segalman
Journal:  J Hand Surg Am       Date:  2006-03       Impact factor: 2.230

10.  Operative management of distal radial fractures with 2.4-millimeter locking plates. A multicenter prospective case series.

Authors:  Jesse B Jupiter; M Marent-Huber
Journal:  J Bone Joint Surg Am       Date:  2009-01       Impact factor: 5.284

View more
  2 in total

1.  The effects of screw length on stability of simulated osteoporotic distal radius fractures fixed with volar locking plates.

Authors:  Lindley B Wall; Michael D Brodt; Matthew J Silva; Martin I Boyer; Ryan P Calfee
Journal:  J Hand Surg Am       Date:  2012-02-02       Impact factor: 2.230

2.  The 70° Supinated Oblique View: A Cadaveric Analysis to Determine Ideal Radial Styloid Screw Position in Locked Volar Plating of Distal Radius Fractures.

Authors:  Kirsten Sumner; Louis C Grandizio; Samantha Schultz; Joel C Klena
Journal:  Hand (N Y)       Date:  2019-03-29
  2 in total