Literature DB >> 16446589

Computer-assisted total knee arthroplasty using patient-specific templating.

M A Hafez1, K L Chelule, B B Seedhom, K P Sherman.   

Abstract

Current techniques used for total knee arthroplasty rely on conventional instrumentation that violates the intramedullary canals. Accuracy of the instrumentation is questionable, and assembly and disposal of the numerous pieces is time consuming. Navigation techniques are more accurate, but their broad application is limited by cost and complexity. We aimed to prove a new concept of computer-assisted preoperative planning to provide patient-specific templates that can replace conventional instruments. Computed tomography-based planning was used to design two virtual templates. Using rapid prototyping technology, virtual templates were transferred into physical templates (cutting blocks) with surfaces that matched the distal femur and proximal tibia. We performed 45 total knee arthroplasties on 16 cadaveric and 29 plastic knees, including a comparative trial against conventional instrumentations. All operations were performed using patient-specific templates with no conventional instrumentations, intramedullary perforation, tracking, or registration. The mean time for bone cutting was 9 minutes with a surgical assistant and 11 minutes without an assistant. Computer-assisted analyses of six random computed tomography scans showed mean errors for alignment and bone resection within 1.7 degrees and 0.8 mm (maximum, 2.3 degrees and 1.2 mm, respectively). Patient-specific templates are a practical alternative to conventional instrumentations, but additional clinical validation is required before clinical use.

Entities:  

Mesh:

Year:  2006        PMID: 16446589     DOI: 10.1097/01.blo.0000201148.06454.ef

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  50 in total

1.  Letter to the editor: Do "premium" joint implants add value?: analysis of high cost joint implants in a community registry.

Authors:  Michael McClure; Peter Heeckt
Journal:  Clin Orthop Relat Res       Date:  2011-01       Impact factor: 4.176

2.  Assessment of patient-specific instrumentation precision through bone resection measurements.

Authors:  F Zambianchi; A Colombelli; V Digennaro; A Marcovigi; R Mugnai; F Fiacchi; D Sandoni; A Belluati; F Catani
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-12-24       Impact factor: 4.342

3.  Improved accuracy of component alignment with the implementation of image-free navigation in total knee arthroplasty.

Authors:  Ralf E Rosenberger; Christian Hoser; Sebastian Quirbach; Rene Attal; Alfred Hennerbichler; Christian Fink
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2007-12-22       Impact factor: 4.342

4.  Clinical, functional, and radiographic outcomes following total knee arthroplasty with patient-specific instrumentation, computer-assisted surgery, and manual instrumentation: a short-term follow-up study.

Authors:  Mark Yaffe; Michael Luo; Nitin Goyal; Philip Chan; Anay Patel; Max Cayo; S David Stulberg
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-12-13       Impact factor: 2.924

5.  A laboratory comparison of computer navigation and individualized guides for distal radius osteotomy.

Authors:  Burton Ma; Manuela Kunz; Braden Gammon; Randy E Ellis; David R Pichora
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-12-10       Impact factor: 2.924

6.  Tailor-made Surgical Guide Reduces Incidence of Outliers of Cup Placement.

Authors:  Takehito Hananouchi; Masanobu Saito; Tsuyoshi Koyama; Nobuhiko Sugano; Hideki Yoshikawa
Journal:  Clin Orthop Relat Res       Date:  2009-07-24       Impact factor: 4.176

7.  Small improvements in mechanical axis alignment achieved with MRI versus CT-based patient-specific instruments in TKA: a randomized clinical trial.

Authors:  Tilman Pfitzner; Matthew P Abdel; Philipp von Roth; Carsten Perka; Hagen Hommel
Journal:  Clin Orthop Relat Res       Date:  2014-07-15       Impact factor: 4.176

8.  Accuracy of Individualized Custom Tibial Cutting Guides in UKA.

Authors:  Thomas J Heyse; Joseph D Lipman; Carl W Imhauser; Scott M Tucker; Yogesh Rajak; Geoffrey H Westrich
Journal:  HSS J       Date:  2014-09-09

9.  Design improvement in patient-specific instrumentation for total knee arthroplasty improved the accuracy of the tibial prosthetic alignment in the coronal and axial planes.

Authors:  Kazumasa Yamamura; Yukihide Minoda; Ryo Sugama; Yoichi Ohta; Suguru Nakamura; Hideki Ueyama; Hiroaki Nakamura
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2019-06-25       Impact factor: 4.342

10.  [Total knee arthroplasty with the use of patient specific instruments. The VISIONAIRE system].

Authors:  C O Tibesku
Journal:  Orthopade       Date:  2016-04       Impact factor: 1.087

View more