Simulated bioprosthetic heart valve deformation under quasi-static loading.

Wei Sun1, Ajay Abad, Michael S Sacks.   

Abstract

For more than 40 years, the replacement of diseased natural heart valves with prosthetic devices has dramatically extended the quality and length of the lives of millions of patients worldwide. However, bioprosthetic heart valves (BHV) continue to fail due to structural failure resulting from poor tissue durability and faulty design. Clearly, an in-depth understanding of the biomechanical behavior of BHV at both the tissue and functional prosthesis levels is essential to improving BHV design and to reduce rates of failure. In this study, we simulated quasi-static BHV leaflet deformation under 40, 80, and 120 mm Hg quasi-static transvalvular pressures. A Fung-elastic material model was used that incorporated material parameters and axes derived from actual leaflet biaxial tests and measured leaflet collagen fiber structure. Rigorous experimental validation of predicted leaflet strain field was used to validate the model results. An overall maximum discrepancy of 2.36% strain between the finite element (FE) results and experiment measurements was obtained, indicating good agreement between computed and measured major principal strains. Parametric studies utilizing the material parameter set from one leaflet for all three leaflets resulted in substantial variations in leaflet stress and strain distributions. This result suggests that utilization of actual leaflet material properties is essential for accurate BHV FE simulations. The present study also underscores the need for rigorous experimentation and accurate constitutive models in simulating BHV function and design.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16438226     DOI: 10.1115/1.2049337

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  54 in total

1.  A novel fibre-ensemble level constitutive model for exogenous cross-linked collagenous tissues.

Authors:  Michael S Sacks; Will Zhang; Silvia Wognum
Journal:  Interface Focus       Date:  2016-02-06       Impact factor: 3.906

2.  An anisotropic constitutive model for immersogeometric fluid-structure interaction analysis of bioprosthetic heart valves.

Authors:  Michael C H Wu; Rana Zakerzadeh; David Kamensky; Josef Kiendl; Michael S Sacks; Ming-Chen Hsu
Journal:  J Biomech       Date:  2018-04-12       Impact factor: 2.712

3.  Flow interactions with cells and tissues: cardiovascular flows and fluid-structure interactions. Sixth International Bio-Fluid Mechanics Symposium and Workshop, March 28-30, 2008, Pasadena, California.

Authors:  Morton H Friedman; Rob Krams; Krishnan B Chandran
Journal:  Ann Biomed Eng       Date:  2010-03       Impact factor: 3.934

4.  A metric for the stiffness of calcified aortic valves using a combined computational and experimental approach.

Authors:  Hoda Maleki; Shahrokh Shahriari; Louis G Durand; Michel R Labrosse; Lyes Kadem
Journal:  Med Biol Eng Comput       Date:  2013-09-14       Impact factor: 2.602

Review 5.  Computational modeling of cardiac valve function and intervention.

Authors:  Wei Sun; Caitlin Martin; Thuy Pham
Journal:  Annu Rev Biomed Eng       Date:  2014-04-16       Impact factor: 9.590

6.  Simulation of long-term fatigue damage in bioprosthetic heart valves: effects of leaflet and stent elastic properties.

Authors:  Caitlin Martin; Wei Sun
Journal:  Biomech Model Mechanobiol       Date:  2013-10-04

7.  Stent and leaflet stresses across generations of balloon-expandable transcatheter aortic valves.

Authors:  Yue Xuan; Danny Dvir; Zhongjie Wang; Jian Ye; Julius M Guccione; Liang Ge; Elaine E Tseng
Journal:  Interact Cardiovasc Thorac Surg       Date:  2020-06-01

8.  Planar biaxial mechanical behavior of bioartificial tissues possessing prescribed fiber alignment.

Authors:  Choon-Sik Jhun; Michael C Evans; Victor H Barocas; Robert T Tranquillo
Journal:  J Biomech Eng       Date:  2009-08       Impact factor: 2.097

Review 9.  Polymeric trileaflet prosthetic heart valves: evolution and path to clinical reality.

Authors:  Thomas E Claiborne; Marvin J Slepian; Syed Hossainy; Danny Bluestein
Journal:  Expert Rev Med Devices       Date:  2012-11       Impact factor: 3.166

10.  In vivo biomechanical assessment of triglycidylamine crosslinked pericardium.

Authors:  Michael S Sacks; Hirotsugu Hamamoto; Jeanne M Connolly; Robert C Gorman; Joseph H Gorman; Robert J Levy
Journal:  Biomaterials       Date:  2007-09-05       Impact factor: 12.479