Compensation comparison method for assessment of retinal straylight.

Luuk Franssen1, Joris E Coppens, Thomas J T P van den Berg.   

Abstract

PURPOSE: Presently, no instrument or method exists that is generally accepted for routine clinical assessment of (functional) retinal straylight. Yet retinal straylight is the cause of major patient complaints, such as hindrance from glare and loss of contrast. It results from disturbances in the optical media that increase light-scattering over angles of 1 degrees to 90 degrees . Its assessment would help to decide whether to perform surgery for (early) cataract and would help in the evaluation of corneal or vitreal turbidity.
METHODS: The psychophysical technique of the "direct compensation" method was adapted to make it suitable for routine clinical assessment. In the new approach, called "compensation comparison, " the central test field is subdivided into two half fields: one with and one without counterphase compensation light. The subject's task is a forced-choice comparison between the two half fields, to decide which half flickers more strongly. A theoretical form for the respective psychometric function was defined and experimentally verified in a laboratory experiment involving seven subjects, with and without artificially increased light scattering. The method was applied in a separate multicenter study. Its reliability was additionally tested with a commercial implement (C-Quant; Oculus Optikgeräte, Wetzlar-Dutenhofen, Germany).
RESULTS: A repeated-measures SD of 0.07 log units was achieved, to be compared with differences in the young normal population of 0.4 log units and an increase with healthy aging by 0.5 log units at 80 years and by 1.0 or more log units with (early) cataract or corneal disturbances. Reliability was further found to be high when using the commercial version of the
CONCLUSIONS: The compensation comparison method for measuring retinal straylight is suited for clinical use to diagnose patients with complaints caused by large angle light scattering in the eye such as early cataract.

Entities:  

Mesh:

Year:  2006        PMID: 16431978     DOI: 10.1167/iovs.05-0690

Source DB:  PubMed          Journal:  Invest Ophthalmol Vis Sci        ISSN: 0146-0404            Impact factor:   4.799


  44 in total

1.  Analysis of optical side-effects of fs-laser therapy in human presbyopic lens simulated with modified contact lenses.

Authors:  Monika Peter; Robert Kammel; Roland Ackermann; Stefan Schramm; Bernd-Ulrich Seifert; Katharina Frey; Marcus Blum; Stefan Nolte; Kathleen S Kunert
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-02-29       Impact factor: 3.117

2.  Method for in vitro assessment of straylight from intraocular lenses.

Authors:  Grzegorz Łabuz; Fernando Vargas-Martín; Thomas J T P van den Berg; Norberto López-Gil
Journal:  Biomed Opt Express       Date:  2015-10-19       Impact factor: 3.732

3.  Comparison of pseudophakic retinal straylight in spherical/aspherical and hydrophobic/hydrophilic intraocular lens.

Authors:  Yong Tang; Hui Song; Jing Chen; Xin Tang
Journal:  Int J Ophthalmol       Date:  2015-12-18       Impact factor: 1.779

4.  Straylight, lens yellowing and aberrations of eyes in Type 1 diabetes.

Authors:  X Adnan; Marwan Suheimat; Ankit Mathur; Nathan Efron; David A Atchison
Journal:  Biomed Opt Express       Date:  2015-03-13       Impact factor: 3.732

5.  Retinal straylight and light distortion phenomena in normal and post-LASIK eyes.

Authors:  Alejandro Cerviño; Cesar Villa-Collar; Jose Manuel Gonzalez-Meijome; Teresa Ferrer-Blasco; Santiago García-Lázaro
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-01-18       Impact factor: 3.117

6.  Relevance of pupil size in the clinical determination of retinal straylight on young healthy human eyes.

Authors:  Santiago García-Lázaro; Teresa Ferrer-Blasco; Susana Ortí-Navarro; Alejandro Cerviño; Robert Montés-Micó
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-12-22       Impact factor: 3.117

7.  Performance of a differential contrast sensitivity method to measure intraocular scattering.

Authors:  Alexandros Pennos; Harilaos Ginis; Augusto Arias; Dimitrios Christaras; Pablo Artal
Journal:  Biomed Opt Express       Date:  2017-02-06       Impact factor: 3.732

8.  Validation of a spectral light scattering method to differentiate large from small particles in intraocular lenses.

Authors:  Grzegorz Łabuz; Eleni Papadatou; Fernando Vargas-Martín; Norberto López-Gil; Nicolaas J Reus; Thomas J T P van den Berg
Journal:  Biomed Opt Express       Date:  2017-02-27       Impact factor: 3.732

9.  Reproducibility of straylight measurement by C-Quant for assessment of retinal straylight using the compensation comparison method.

Authors:  Ivo Guber; Lucas M Bachmann; Josef Guber; Frank Bochmann; Alex P Lange; Michael A Thiel
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-05-13       Impact factor: 3.117

10.  System based on the contrast of Purkinje images to measure corneal and lens scattering.

Authors:  Pau Santos; Juan A Martínez-Roda; Juan C Ondategui; Fernando Díaz-Doutón; Jorge A Ortiz Cazal; Meritxell Vilaseca
Journal:  Biomed Opt Express       Date:  2018-09-18       Impact factor: 3.732

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.