Response of 10 pulse oximeters to an in vitro test system.

K J Reynolds1, J T Moyle, M K Sykes, C E Hahn.   

Abstract

Pulse oximeters are often used in situations in which severe hypoxaemia may occur. We have developed an in vitro system to test the accuracy of pulse oximeter calibration. The probe of 10 different oximeters was attached to a model finger in an in vitro blood circuit, and pulse oximeter readings (SpO2) were compared with multi-wavelength in vitro oximeter readings (SO2) over a range of SO2 values from 50 to 100%. The oximeters tested varied widely in their accuracy and linearity. We conclude that the system can test the accuracy, reproducibility and linearity of response of pulse oximeter readings at low oxyhaemoglobin saturations.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1642914     DOI: 10.1093/bja/68.4.365

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  6 in total

1.  In vitro performance test system for pulse oximeters.

Authors:  K J Reynolds; J T Moyle; L B Gale; M K Sykes; C E Hahn
Journal:  Med Biol Eng Comput       Date:  1992-11       Impact factor: 2.602

2.  Evaluation of a new pulse oximeter testing device.

Authors:  J A Fisher; T Martire; G A Volgyesi
Journal:  Can J Anaesth       Date:  1996-02       Impact factor: 5.063

3.  The effect of dyshemoglobins on pulse oximetry: Part I, Theoretical approach and Part II, Experimental results using an in vitro test system.

Authors:  K J Reynolds; E Palayiwa; J T Moyle; M K Sykes; C E Hahn
Journal:  J Clin Monit       Date:  1993-04

4.  Accuracy and utility of a new reflectance pulse oximeter for fetal monitoring during labor.

Authors:  A Izumi; H Minakami; I Sato
Journal:  J Clin Monit       Date:  1997-03

5.  A prototype device for standardized calibration of pulse oximeters.

Authors:  C Hornberger; P Knoop; W Nahm; H Matz; E Konecny; H Gehring; R Bonk; H Frankenberger; G Meyfroidt; P Wouters; J Gil-Rodriguez; L Ponz; K Benekos; J Valais; J Avgerinos; A Karoutis; A Ikiades; S Weininger
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

6.  Multi-analyte calibration and verification of a multi-parameter laser-based pulse oximeter.

Authors:  Jonas A Pologe; Donald H Arnold; Theodore P Delianides
Journal:  J Clin Monit Comput       Date:  2021-04-19       Impact factor: 2.502