Literature DB >> 16306839

Biomechanical evaluation of the pullout strength of cervical screws.

Bryan P Conrad1, Andrew G Cordista, MaryBeth Horodyski, Glenn R Rechtine.   

Abstract

OBJECTIVE: In the process of anterior cervical fusion, little is known about the biomechanics of anterior cervical screw pullout. In this study, three different aspects of cervical screw fixation were evaluated: self-tapping (ST) versus self-drilling (SD) screws, the effect of screw geometry (length, diameter, thread pitch), and the use of rescue screws.
METHODS: Nine screws consisting of different diameters, lengths, and thread pitch (cancellous and cortical) were tested in peak pullout force in an artificial bone model using an MTS 858 Mini Bionix test system. Rescue screws (4.5 mm) were then inserted in the failed holes of 4.0-mm screws and extracted to determine their holding strength.
RESULTS: Length of screws and thread pitch both had a significant effect on the pullout force. Each 1 mm of increased screw length translates to 16 N of increased force to pullout in the foam bone model. Pullout strength did not vary significantly according to screw diameter or between SD and ST screws. However, the SD screw has an advantage because it can decrease the length of surgery. A decrease in pullout force of between 43% and 70% was found when using rescue screws.
CONCLUSIONS: In situations in which the use of rescue/salvage screws is required, the surgeon should anticipate a significant decrease in the holding force compared with the original screw. Future directions for research include an evaluation of pullout force for screw and plate constructs.

Mesh:

Year:  2005        PMID: 16306839     DOI: 10.1097/01.bsd.0000140196.99995.65

Source DB:  PubMed          Journal:  J Spinal Disord Tech        ISSN: 1536-0652


  14 in total

1.  Experimental in vivo acute and chronic biomechanical and histomorphometrical comparison of self-drilling and self-tapping anterior cervical screws.

Authors:  Maximiliano Aguiar Porto; Patrícia Silva; Rodrigo Rosa; José Batista Volpon; Antônio Carlos Shimano; Francisco José Albuquerque de Paula; Helton Luiz Aparecido Defino
Journal:  Eur Spine J       Date:  2011-12-16       Impact factor: 3.134

2.  Cervical anterior transpedicular screw fixation (ATPS)--Part II. Accuracy of manual insertion and pull-out strength of ATPS.

Authors:  Heiko Koller; Frank Acosta; Mark Tauber; Michael Fox; Hudelmaier Martin; Rosmarie Forstner; Peter Augat; Rainer Penzkofer; Christian Pirich; H Kässmann; Herbert Resch; Wolfgang Hitzl
Journal:  Eur Spine J       Date:  2008-01-26       Impact factor: 3.134

3.  Effect of pilot hole on biomechanical and in vivo pedicle screw-bone interface.

Authors:  Patrícia Silva; Rodrigo César Rosa; Antonio Carlos Shimano; Helton L A Defino
Journal:  Eur Spine J       Date:  2013-05-08       Impact factor: 3.134

4.  Biomechanics of Cervical "Skip" Corpectomy Versus Standard Multilevel Corpectomy.

Authors:  Murat Yilmaz; Kasim Zafer Yüksel; Seungwon Baek; Anna G U S Newcomb; Sedat Dalbayrak; Volker K H Sonntag; Neil R Crawford
Journal:  Clin Spine Surg       Date:  2017-04       Impact factor: 1.876

5.  Can high-friction intraannular material increase screw pullout strength in osteoporotic bone?

Authors:  Daniel Bronsnick; Ryan E Harold; Ari Youderian; Giovanni Solitro; Farid Amirouche; Benjamin Goldberg
Journal:  Clin Orthop Relat Res       Date:  2014-10-02       Impact factor: 4.176

6.  Cement-augmented screws in a cervical two-level corpectomy with anterior titanium mesh cage reconstruction: a biomechanical study.

Authors:  Sebastian Hartmann; Claudius Thomé; Anja Tschugg; Johannes Paesold; Pujan Kavakebi; Werner Schmölz
Journal:  Eur Spine J       Date:  2017-01-21       Impact factor: 3.134

7.  Serious complication of cement augmentation for damaged pilot hole.

Authors:  Moon Young Jung; Dong Ah Shin; In Bo Hahn; Tae Gon Kim; Ryoong Huh; Sang Sup Chung
Journal:  Yonsei Med J       Date:  2010-05       Impact factor: 2.759

Review 8.  4- and 5-level anterior fusions of the cervical spine: review of literature and clinical results.

Authors:  Heiko Koller; Axel Hempfing; Luis Ferraris; Oliver Maier; Wolfgang Hitzl; Peter Metz-Stavenhagen
Journal:  Eur Spine J       Date:  2007-06-29       Impact factor: 3.134

9.  Pullout of a lumbar plate with varying screw lengths.

Authors:  Daniel Kyle Palmer; David Rios; Wyzscx Merfil Patacxil; Paul A Williams; Wayne K Cheng; Serkan İnceoğlu
Journal:  Int J Spine Surg       Date:  2012-12-01

10.  A Surgical Method for Determining Proper Screw Length in ACDF.

Authors:  Hae-Gi Park; Moo-Sung Kang; Kyung-Hyun Kim; Jeong-Yoon Park; Keun-Su Kim; Sung-Uk Kuh
Journal:  Korean J Spine       Date:  2014-09-30
View more