Literature DB >> 16292656

GDx-VCC performance in discriminating normal from glaucomatous eyes with early visual field loss.

Stefano Da Pozzo1, Mirko Fuser, Odilla Vattovani, Giuseppe Di Stefano, Giuseppe Ravalico.   

Abstract

BACKGROUND: To evaluate the ability of scanning laser polarimetry with variable corneal compensation (GDx-VCC) in separating healthy from glaucomatous patients with early visual field (VF) loss.
METHODS: Sixty-two healthy and 48 glaucomatous age-matched patients with early glaucoma [mean deviation (MD): -1.74dB +/-1.69] underwent complete ophthalmological evaluation, automated achromatic perimetry (AAP) and retinal nerve fiber layer (RNFL) measurement with GDx-VCC. One randomly selected eye from each subject was considered. Glaucomatous VF defects had either Glaucoma Hemifield Test (GHT) outside normal limits or pattern standard deviation (PSD) outside 95% confidence limits. Mean (+/-SD) MD, PSD and GDx-VCC parameters in the two groups were compared by t-test. For each GDx-VCC parameter, area under receiver operating characteristics (AUROC) curve and sensitivity at predetermined specificity > or =80% and > or =95% were calculated. Moreover, the parameter with largest AUROC was evaluated by likelihood ratios (LRs).
RESULTS: Mean values for MD, PSD and ten of 14 GDx-VCC parameters were significantly different between the two groups (P<0.001). The three parameters with largest AUROCs were the nerve fiber indicator (NFI) (0.870), superior average (0.817) and normalized superior area (0.816) (P=0.08 for differences between AUROCs). NFI displayed sensitivity values of 80.2% and 60.4% for specificity > or =80% and > or =95%, respectively. At NFI cutoff value of 30, positive LR was 34.9 (95% CI: 4.9-247.6) and negative LR was 0.45 (95% CI: 0.32-0.61). Interval LRs showed large effect on post-test probability for NFI values < or =18 or > or =31.
CONCLUSIONS: In our sample of eyes with early VF loss, GDx-VCC showed moderate-to-good discriminating ability. Among the best performing parameters, NFI had the largest AUROC, but several glaucomatous eyes (21, 43.8%) had NFI <30. This suggests that algorithm for NFI calculation requires some refinement when eyes with early VF loss are evaluated.

Entities:  

Mesh:

Year:  2005        PMID: 16292656     DOI: 10.1007/s00417-005-0144-y

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  27 in total

1.  Evidence-based emergency medicine/editorial. The problem with sensitivity and specificity...

Authors:  E John Gallagher
Journal:  Ann Emerg Med       Date:  2003-08       Impact factor: 5.721

2.  Comparison of algorithms for detection of localised nerve fibre layer defects using scanning laser polarimetry.

Authors:  F A Medeiros; R Susanna
Journal:  Br J Ophthalmol       Date:  2003-04       Impact factor: 4.638

3.  A method of comparing the areas under receiver operating characteristic curves derived from the same cases.

Authors:  J A Hanley; B J McNeil
Journal:  Radiology       Date:  1983-09       Impact factor: 11.105

4.  Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry.

Authors:  D S Greenfield; R W Knighton; X R Huang
Journal:  Am J Ophthalmol       Date:  2000-06       Impact factor: 5.258

5.  Scanning laser polarimetry of the retinal nerve fiber layer in perimetrically unaffected eyes of glaucoma patients.

Authors:  Nicolaas J Reus; Hans G Lemij
Journal:  Ophthalmology       Date:  2004-12       Impact factor: 12.079

6.  Association between scanning laser polarimetry measurements using variable corneal polarization compensation and visual field sensitivity in glaucomatous eyes.

Authors:  Christopher Bowd; Linda M Zangwill; Robert N Weinreb
Journal:  Arch Ophthalmol       Date:  2003-07

7.  Comparison of the GDx VCC scanning laser polarimeter, HRT II confocal scanning laser ophthalmoscope, and stratus OCT optical coherence tomograph for the detection of glaucoma.

Authors:  Felipe A Medeiros; Linda M Zangwill; Christopher Bowd; Robert N Weinreb
Journal:  Arch Ophthalmol       Date:  2004-06

8.  The mode of progressive disc cupping in ocular hypertension and glaucoma.

Authors:  J E Pederson; D R Anderson
Journal:  Arch Ophthalmol       Date:  1980-03

9.  Clinically detectable nerve fiber atrophy precedes the onset of glaucomatous field loss.

Authors:  A Sommer; J Katz; H A Quigley; N R Miller; A L Robin; R C Richter; K A Witt
Journal:  Arch Ophthalmol       Date:  1991-01

10.  Scanning laser polarimetry with variable corneal compensation and optical coherence tomography in normal and glaucomatous eyes.

Authors:  Harmohina Bagga; David S Greenfield; William Feuer; Robert W Knighton
Journal:  Am J Ophthalmol       Date:  2003-04       Impact factor: 5.258

View more
  10 in total

1.  Scanning laser polarimetry with variable and enhanced corneal compensation in normal and glaucomatous eyes.

Authors:  Mitra Sehi; Delia C Guaqueta; William J Feuer; David S Greenfield
Journal:  Am J Ophthalmol       Date:  2006-10-25       Impact factor: 5.258

2.  Imaging of the optic nerve and retinal nerve fiber layer: an essential part of glaucoma diagnosis and monitoring.

Authors:  Jacek Kotowski; Gadi Wollstein; Hiroshi Ishikawa; Joel S Schuman
Journal:  Surv Ophthalmol       Date:  2013-10-16       Impact factor: 6.048

3.  Glaucoma classification model based on GDx VCC measured parameters by decision tree.

Authors:  Mei-Ling Huang; Hsin-Yi Chen
Journal:  J Med Syst       Date:  2009-07-04       Impact factor: 4.460

Review 4.  Optic nerve head and fibre layer imaging for diagnosing glaucoma.

Authors:  Manuele Michelessi; Ersilia Lucenteforte; Francesco Oddone; Miriam Brazzelli; Mariacristina Parravano; Sara Franchi; Sueko M Ng; Gianni Virgili
Journal:  Cochrane Database Syst Rev       Date:  2015-11-30

5.  Retinal vessel diameter in bilateral glaucoma suspects: comparison between the eye converted to glaucoma and the contralateral non-converted eye.

Authors:  Eunjoo Yoo; Chungkwon Yoo; Tae-Eun Lee; Yong Yeon Kim
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2016-06-02       Impact factor: 3.117

6.  Evaluation of the significance of some diagnostic parameters in making an early diagnose of primary open-angle glaucoma.

Authors:  Barbara Polaczek-Krupa; Iwona Grabska-Liberek
Journal:  Med Sci Monit       Date:  2012-07

7.  Scanning laser polarimetry in glaucoma.

Authors:  Tanuj Dada; Reetika Sharma; Dewang Angmo; Gautam Sinha; Shibal Bhartiya; Sanjay K Mishra; Anita Panda; Ramanjit Sihota
Journal:  Indian J Ophthalmol       Date:  2014-11       Impact factor: 1.848

8.  Applicability of standard parameters in diagnostics of primary open-angle glaucoma.

Authors:  Barbara Polaczek-Krupa; Iwona Grabska-Liberek
Journal:  Med Sci Monit       Date:  2013-08-09

9.  Use of macular thickness parameters for the diagnosis of primary open-angle glaucoma.

Authors:  Barbara Polaczek-Krupa; Iwona Grabska-Liberek; Michał Kamiński
Journal:  Arch Med Sci       Date:  2014-02-23       Impact factor: 3.318

10.  Photopic negative response of full-field electroretinography in patients with different stages of glaucomatous optic neuropathy.

Authors:  Marta Kirkiewicz; Wojciech Lubiński; Krzysztof Penkala
Journal:  Doc Ophthalmol       Date:  2016-01-29       Impact factor: 2.379

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.