Literature DB >> 23934540

Applicability of standard parameters in diagnostics of primary open-angle glaucoma.

Barbara Polaczek-Krupa1, Iwona Grabska-Liberek.   

Abstract

BACKGROUND: The aim of this study was to evaluate the sensitivity, specificity, and likelihood ratio of standard diagnostic parameters C/D, NFI, and MD and their applicability in diagnostics of primary open-angle glaucoma.
MATERIAL AND METHODS: This study included 194 subjects (371 eyes), ages 30 to 65 years old, who underwent full ophthalmologic examination and HRT, GDx, and FDT examinations. The values of C/D, NFI, and MD diagnostic parameters were determined. The data were statistically evaluated to determine their sensitivity, specificity, and likelihood ratio.
RESULTS: Values of the positive and negative likelihood ratios were C/D (11.471 and 0.159), NFI (3.739 and 0.152), and MD (6.323 and 0.309), respectively.
CONCLUSIONS: The C/D parameter showed the highest sensitivity and specificity, as well as high positive likelihood ratio and near-zero negative likelihood ratio. The NFI and MD parameters showed lower likelihood ratios and their applicability for the diagnosis of primary open-angle glaucoma is limited.

Entities:  

Mesh:

Year:  2013        PMID: 23934540      PMCID: PMC3743461          DOI: 10.12659/MSM.883994

Source DB:  PubMed          Journal:  Med Sci Monit        ISSN: 1234-1010


Background

Contemporary diagnosis of primary open-angle glaucoma is based on evaluation of the optic disc, the neuroretinal rim, and the retinal nerve fiber layers with indirect stereoscopic examination of a dilated pupil using a slit-lamp and Volk lens. Currently, methods such as HRT laser scanning ophthalmoscopy, GDxVCC laser scanning polarimetry, and FDT perimetry [1] are among the more commonly used. The HRT method uses a confocal laser scanning microscope that allows obtaining a 3D morphological image and determining detailed diagnostic parameters of the optic disc [2], including C/D ratio. The GDx laser scanning polarimetry is based on a measurement of the delay of polarized light components and allows the evaluation of retinal nerve fiber layer thickness in the peripapillary region and determining the NFI [2]. Evaluating the field of vision with the use of FDT enables assessment of retinal sensitivity and determination of its mean deviation [3]. Purpose: The aim of the study was to evaluate the applicability of C/D, NFI, and MD standard diagnostic parameters in diagnosis of primary open-angle glaucoma through determination of sensitivity, specificity, and likelihood ratios.

Material and Method

Our study included 194 patients (371 eyes) ages 30–65 years old who reported to the outpatients department of the Clinic of Ophthalmology at the Medical Center of Postgraduate Education in years 2007–2008 for prophylactic examination, glasses choice, or diagnosis towards primary open-angle glaucoma. These patients were divided into 3 groups. Group A consisted of 50 patients (95 eyes, average age of 43.7±9.0 years old) with diagnosed primary open-angle glaucoma, confirmed by additional examinations. Group B consisted of 67 primary open-angle patients with suspected glaucoma (128 eyes, average age of 43.2±11.0 years old) without distinctive glaucoma lesions in the field of vision. Group C was the control group and included healthy patients with normal results of examination for primary open-angle glaucoma (77 patients, 148 eyes, average age of 46.3±10.9 years old). Detailed characteristics of the examined population have been shown in previous publications [4,5]. All the patients underwent full ophthalmologic examination and then were examined with HRT laser scanning ophthalmoscopy, GDxVCC laser scanning polarimetry, and FDT field of vision examination on Humphrey Matrix perimeter. Standard diagnostic parameters were determined. C/D ratio was determined as the square root of the surface ratio of cup to disc [6]. NFI was determined as the probability of glaucomatous damage to the retinal nerve fiber layer [7]. MD value was determined as the difference between retinal sensitivity in a healthy eye (taking age into consideration) and retinal sensitivity values measured in the examined patient in all examined points [8]. All the basic ophthalmological examinations were performed in the Clinic of Ophthalmology at the Medical Center of Postgraduate Education in Warsaw. GDx, HRT, and FDT examinations were performed in the Institute of Glaucoma and Eye Diseases in Warsaw. The study was approved by the Bioethics Commission of the Medical Center of Postgraduate Education in Warsaw (resolution of 2 March 2005 and of 30 January 2008). Although no invasive examination methods were used, patients were informed of the aim and course of the study and gave consent to participate in it. Criteria for qualification of patients were specified in accordance with the guidelines of the European Glaucoma Society [1], described in detail in previous publications [4,5]. Exclusion criteria were: eye refraction error greater than 2 Dsph and media opacity, which lowers the quality of HRT and GDx results, and patients with very small or very large surface of the optic disc (less than 1.69 mm2 or more than 2.82 mm2) evaluated in HRT examination. Computer-stored measurement data matrix (MS Excel spread sheet) was used for statistical analysis [9]. The significance of differences between mean values of measured parameters in all groups was measured with the t test. Sensitivity and specificity of these parameters and likelihood ratio (LR) in relation to glaucoma patients were determined. Mean values of these parameters determined in patients with suspected glaucoma (group B) were accepted as cut-off values for calculations. Sensitivity of the method was designated as the proportion of glaucoma patients correctly detected as a result of the analysis performed, and specificity as the proportion of healthy patients correctly detected as a result of the analysis. Likelihood ratio of the method was calculated from the value of its sensitivity and specificity. If positive LR is greater than 10 and negative LR is less than 0.1, the method may be used to confirm a diagnosis, but when positive LR values are from 5 to 10 and negative LR values are from 0.1 to 0.2, the method can still be considered as a useful diagnostic device.

Results

Results of the measurements of the nerve fiber layer thickness index (NFI), cup-to-disc ratio (C/D), and mean deviation (MD) of retinal sensitivity in patients with primary open-angle glaucoma (group A), patients with suspected primary open-angle glaucoma (group B), and in healthy patients (group C) are shown in Table 1.
Table 1

Results of measurements of NFI, C/D and MD of retinal sensitivity in groups A, B and C.

NFIC/DMD
Group A (glaucoma patients, 95 eyes)
Mean value37.00.65−3.00
Standard deviation22.70.115.07
Group B (glaucoma suspected, 128 eyes)
Mean value18.50.58−0.77
Standard deviation5.60.112.49
Group C (healthy, 148 eyes)
Mean value15.10.43−0.29
Standard deviation4.80.111.94
Calculated values of sensitivity and specificity of individual diagnostic parameters and their likelihood ratio values are shown in Table 2.
Table 2

Juxtaposition of sensitivity values, specificity parameters, C/D, NFI and MD parameters and value of their likelihood ratio.

Result of analysisA GroupC GroupTotal
C/D
Positive811192
Negative14137151
Total95148243
Sensitivity0.853
Specificity0.926
Likelihood ratio(+)11.472
(−)0.159
NFI
Positive8435119
Negative11113124
Total95148243
Sensitivity0.884
Specificity0.764
Likelihood ratio(+)3.739
(−)0.152
MD
Positive691786
Negative26131157
Total95148243
Sensitivity0.726
Specificity0.885
Likelihood ratio(+)6.323
(−)0.309

Discussion

We found statistically significant differences between individual groups of the examined population in results of the C/D, NFI, and MD measurements (Table 1). The mean value of C/D ration in Group A (glaucoma patients) was 0.65±−0.11 vs. 0.43±−0.11 in Group C (healthy control patients). These results acquired are consistent with those of Jonas et al. [10] and confirm that linear C/D ratio is of great use in diagnosing primary open-angle glaucoma. The mean value of NFI in primary open-angle glaucoma patients was 37.0±22.7 vs. 15.1±−4.8 in healthy patients. This is consistent with the observations of da Pozzo et al. [11], who have shown that the NFI is lower than 18 in healthy patients and does not exceed 31 in glaucoma patients. The mean value of MD in primary open-angle glaucoma patients was −3.00±5.07 dB and −0.29±−1.94 dB in healthy controls. Bowd et al. [12] determined the MD value of −0.16 dB for an analogous group of healthy patients and −3.8 dB for primary open-angle glaucoma patients. The analysis of sensitivity and specificity values of the standard diagnostic parameters C/D, NFI and MD and their likelihood ratio value (Table 2) shows that the parameters with the greatest sensitivity and specificity are the C/D ratio (85.3% and 92.6%) and NFI (88.4% and 76.4%). Much lower values were obtained for the MD parameter (72.6 and 88.5). The positive likelihood ratio for C/D parameter was 11.5 and the negative likelihood ratio was 0.16. In the case of NFI and MD parameters, the positive likelihood ratios were 3.74 and 6.32, respectively, and the negative ratios 0.15 and 0.31, respectively. The very high positive likelihood ratio obtained for the C/D parameter (greater than 10) and near-zero likelihood ratio means that confirmation of glaucomatous damage to the optic disc can be obtained through measurement of this parameter. Although the measurement of NFI and MD can reveal certain useful information, it is of lesser diagnostic value.

Conclusions

The C/D parameter showed the highest sensitivity and specificity, as well as high positive likelihood ratio and near-zero negative likelihood ratio. The NFI and MD parameters showed lower likelihood ratios and their applicability for the diagnosis of primary open-angle glaucoma is limited.
  6 in total

1.  [Evaluation of retinal thickness in the macula region in a population of healthy subjects by using the retinal thickness analyzer RTA].

Authors:  Barbara Polaczek-Krupa; Iwona Grabska-Liberek
Journal:  Klin Oczna       Date:  2009

2.  GDx-VCC performance in discriminating normal from glaucomatous eyes with early visual field loss.

Authors:  Stefano Da Pozzo; Mirko Fuser; Odilla Vattovani; Giuseppe Di Stefano; Giuseppe Ravalico
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2005-11-16       Impact factor: 3.117

3.  [Changes of the retinal thickness in the macula region in primary open-angle glaucoma patients measured with RTA analyzer].

Authors:  Barbara Polaczek-Krupa; Iwona Grabska-Liberek
Journal:  Klin Oczna       Date:  2010

4.  Detecting early glaucoma by assessment of retinal nerve fiber layer thickness and visual function.

Authors:  C Bowd; L M Zangwill; C C Berry; E Z Blumenthal; C Vasile; C Sanchez-Galeana; C F Bosworth; P A Sample; R N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  2001-08       Impact factor: 4.799

5.  Visual function-specific perimetry for indirect comparison of different ganglion cell populations in glaucoma.

Authors:  P A Sample; C F Bosworth; E Z Blumenthal; C Girkin; R N Weinreb
Journal:  Invest Ophthalmol Vis Sci       Date:  2000-06       Impact factor: 4.799

6.  Ranking of optic disc variables for detection of glaucomatous optic nerve damage.

Authors:  J B Jonas; A Bergua; P Schmitz-Valckenberg; K I Papastathopoulos; W M Budde
Journal:  Invest Ophthalmol Vis Sci       Date:  2000-06       Impact factor: 4.799

  6 in total
  1 in total

1.  Predictive Values of Optical Coherence Tomography (OCT) Parameters in Assessment of Glaucoma progression.

Authors:  Sanja Sefic Kasumovic; Aida Kasumovic; Suzana Pavljasevic; Emir Cabric; Milka Mavija; Irena Sesar; Sabina Dacić- Lepara; Mirko Jankov
Journal:  Acta Inform Med       Date:  2014-08-21
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.