Literature DB >> 16286410

Shapability, memory, and luminal changes in microcatheters after steam shaping: a comparison of 11 different microcatheters.

Hiro Kiyosue1, Yuzo Hori, Shunro Matsumoto, Mika Okahara, Syuichi Tanoue, Yoshiko Sagara, Hiromu Mori.   

Abstract

PURPOSE: The purpose of this study was to compare the characteristics of shaped microcatheters, including shapability, durability, and luminal changes.
MATERIALS AND METHODS: Eleven brands of steam-shaped microcatheters and one brand of preshaped microcatheter were evaluated. There were 2 nonreinforced and 10 reinforced devices supported by coils. For evaluation of shapability, the tip angle of 6 samples of each brand were measured after steam-shaping for 20 seconds with a shaping mandrel bent at a 90 degrees or 150 degrees angle. The ability to maintain the shaped angle after guidewire insertion stress (durability) was compared by calculation of the change in the tip angle by using 3 samples of each brand. Luminal change after steam shaping was evaluated by calculation of narrowing rate of the smallest diameter and observation of the surface morphology of the mold of each catheter lumen by using a silicone polymer by means of a fluorescent projection method.
RESULTS: The nonreinforced microcatheters and the fiber-braided microcatheter showed higher shapability than the others. The degree of distal microcatheter straightening with the microguidewire insertion was less pronounced in the preshaped microcatheter and the fiber-braided microcatheter. Spontaneous recovery to the initial tip angle 5 minutes after the guidewire procedure was observed in 10 brands to various degrees (87%-98%). Irregular luminal surface morphology at the angled portion was found in 6 reinforced brands. One nonreinforced catheter and the fiber-braided catheter showed high narrowing rates >6%.
CONCLUSION: There are differences in shapabilty, durability, and luminal changes of steam shaping in 12 brands of microcatheters. These characteristics could be important factors in catheter choice for endovascular procedures.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 16286410      PMCID: PMC7976220     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  4 in total

1.  Marker alignment for Guglielmi detachable coil embolization: practical considerations.

Authors:  O-Ki Kwon; Moon Hee Han
Journal:  AJNR Am J Neuroradiol       Date:  2002-09       Impact factor: 3.825

2.  Turn-back technique with use of a shaped microcatheter for superselective catheterization of arteries originating at acute angles.

Authors:  Hiro Kiyosue; Shunro Matsumoto; Yuzo Hori; Mika Okahara; Yoshiko Sagara; Hiromu Mori
Journal:  J Vasc Interv Radiol       Date:  2004-06       Impact factor: 3.464

3.  Endosaccular aneurysm occlusion with Guglielmi detachable coils for obstructive hydrocephalus caused by a large basilar tip aneurysm. Case report.

Authors:  A Watanabe; K Imamura; R Ishii
Journal:  J Neurosurg       Date:  1999-10       Impact factor: 5.115

4.  Distal-tip shape-consistency testing of steam-shaped microcatheters suitable for cerebral aneurysm coil placement.

Authors:  Toshi Abe; Masaru Hirohata; Norimitsu Tanaka; Yusuke Uchiyama; Kiminori Fujimoto; Naoko Fujimura; Alexander M Norbash; Naofumi Hayabuchi
Journal:  AJNR Am J Neuroradiol       Date:  2004 Jun-Jul       Impact factor: 3.825

  4 in total
  7 in total

1.  Complex curve microcatheters for berry aneurysm endovascular therapy.

Authors:  R S Pakbaz; C W Kerber
Journal:  AJNR Am J Neuroradiol       Date:  2007-01       Impact factor: 3.825

2.  Structural limitations of currently available microcatheters and coils for endovascular coiling of very small aneurysms.

Authors:  Yong Cheol Lim; Byung Moon Kim; Yong Sam Shin; Sun Yong Kim; Joonho Chung
Journal:  Neuroradiology       Date:  2008-05       Impact factor: 2.804

3.  Aneurysm coil embolization using a 1.5-fr distal outer diameter microcatheter.

Authors:  David A Stidd; Demetrius K Lopes; Michael Chen
Journal:  Neurointervention       Date:  2014-02-28

4.  Relationship between the morphology of A-1 segment of anterior cerebral artery and anterior communicating artery aneurysms.

Authors:  Wenfeng Feng; Long Zhang; Weiguang Li; Guozhong Zhang; Xiaoyan He; Gang Wang; Mingzhou Li; Songtao Qi
Journal:  Afr Health Sci       Date:  2014-03       Impact factor: 0.927

5.  A New Method of Microcatheter Heat-Forming for Cerebral Aneurysmal Coiling Using Stereolithography Three-Dimensional Printed Hollow Vessel Models.

Authors:  Sadao Nakajima; Makoto Sakamoto; Hiroki Yoshioka; Tetsuji Uno; Masamichi Kurosaki
Journal:  Yonago Acta Med       Date:  2021-02-01       Impact factor: 1.641

6.  Z-shaped microcatheter tip shaping for embolization of aneurysms at the proximal A1 segment of the anterior cerebral artery: a technical note.

Authors:  Ha Young Lee; Jae Sung Ahn; Dae Chul Suh; Deok Hee Lee
Journal:  Neurointervention       Date:  2011-08-31

7.  A Novel Technique for Microcatheter Additional Shaping During Intracranial Aneurysmal Coil Embolization: Microcatheter Shaping Cast.

Authors:  Tomotaka Ohshima; Reo Kawaguchi; Ryuya Maejima; Naoki Matsuo; Shigeru Miyachi
Journal:  Asian J Neurosurg       Date:  2019 Jul-Sep
  7 in total