Literature DB >> 15205148

Distal-tip shape-consistency testing of steam-shaped microcatheters suitable for cerebral aneurysm coil placement.

Toshi Abe1, Masaru Hirohata, Norimitsu Tanaka, Yusuke Uchiyama, Kiminori Fujimoto, Naoko Fujimura, Alexander M Norbash, Naofumi Hayabuchi.   

Abstract

BACKGROUND AND
PURPOSE: Although the selection of microcatheter for endovascular aneurysmal treatment is one important factor in patient outcome, the use of steam shaping for achieving safe entry and stability during coil placement has not, to our knowledge, been systematically evaluated. The goal of this study was to compare the durability of distal microcatheter steam shaping in five different catheters with typical intraprocedural stresses that are similar to those encountered during aneurysm coil placement.
METHODS: Distal tips of microcatheters were shaped into a 90 degrees turn with distal straight-segment lengths of 3, 5, or 7 mm by using steam, performed according to the instructions for use included with each catheter. In a water bath kept at body temperature, the changes in catheter tip angle were recorded and measured following microcatheter insertion into a guiding catheter, microguidewire insertion through the microcatheter, and Guglielmi detachable coil (GDC) placement through the microcatheter.
RESULTS: The degree of distal microcatheter straightening with typical intraprocedural manipulations was more pronounced on braided microcatheters and on microcatheters with 3- or 5-mm distal-shaped segments. The degree of spontaneous recovery of the initially steamed shape was more pronounced with nonbraided catheters. The most significant single variable contributing to straightening of a steam-shaped catheter tip was the effect of microguidewire insertion. The catheter-tip straightening effect encountered with inserting GDCs was less than that encountered with microguidewire insertion. We demonstrated that the decreased catheter-tip angle encountered with a large-magnitude straightening stress spontaneously recovered once the stress was removed or when it was reduced to a smaller magnitude stress.
CONCLUSION: Our study shows that, although braided microcatheters are suitable for maintaining durable configurations when long distal-tip lengths are permissible, nonbraided microcatheters demonstrate the most durable distal-tip configurations when short distal-tip lengths are called for. This may be one of significant factors in catheter choice for endovascular treatment of aneurysm.

Entities:  

Mesh:

Year:  2004        PMID: 15205148      PMCID: PMC7975654     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  3 in total

1.  Marker alignment for Guglielmi detachable coil embolization: practical considerations.

Authors:  O-Ki Kwon; Moon Hee Han
Journal:  AJNR Am J Neuroradiol       Date:  2002-09       Impact factor: 3.825

2.  Electrothrombosis of saccular aneurysms via endovascular approach. Part 2: Preliminary clinical experience.

Authors:  G Guglielmi; F Viñuela; J Dion; G Duckwiler
Journal:  J Neurosurg       Date:  1991-07       Impact factor: 5.115

3.  Electrothrombosis of saccular aneurysms via endovascular approach. Part 1: Electrochemical basis, technique, and experimental results.

Authors:  G Guglielmi; F Viñuela; I Sepetka; V Macellari
Journal:  J Neurosurg       Date:  1991-07       Impact factor: 5.115

  3 in total
  6 in total

1.  Shapability, memory, and luminal changes in microcatheters after steam shaping: a comparison of 11 different microcatheters.

Authors:  Hiro Kiyosue; Yuzo Hori; Shunro Matsumoto; Mika Okahara; Syuichi Tanoue; Yoshiko Sagara; Hiromu Mori
Journal:  AJNR Am J Neuroradiol       Date:  2005 Nov-Dec       Impact factor: 3.825

2.  The feasibility and efficacy of endovascular treatment for very small or tiny ruptured paraclinoid aneurysms.

Authors:  Jing Xu; Xian-Yi Chen; Ding-Yao Jiang; Wen Li; Jian-Min Zhang
Journal:  Neuroradiology       Date:  2012-08-09       Impact factor: 2.804

3.  Efficacy and safety of guidewireless catheterization with a steerable microcatheter in patients scheduled for yttrium-90 radioembolization: a prospective multicenter trial.

Authors:  Zbigniew Serafin; Olivier Dudeck; Maciej Powerski; Florian Wolf; Ralph Drewes; Maciej Pech
Journal:  Wideochir Inne Tech Maloinwazyjne       Date:  2020-01-21       Impact factor: 1.195

4.  Structural limitations of currently available microcatheters and coils for endovascular coiling of very small aneurysms.

Authors:  Yong Cheol Lim; Byung Moon Kim; Yong Sam Shin; Sun Yong Kim; Joonho Chung
Journal:  Neuroradiology       Date:  2008-05       Impact factor: 2.804

5.  A New Method of Microcatheter Heat-Forming for Cerebral Aneurysmal Coiling Using Stereolithography Three-Dimensional Printed Hollow Vessel Models.

Authors:  Sadao Nakajima; Makoto Sakamoto; Hiroki Yoshioka; Tetsuji Uno; Masamichi Kurosaki
Journal:  Yonago Acta Med       Date:  2021-02-01       Impact factor: 1.641

6.  A Novel Technique for Microcatheter Additional Shaping During Intracranial Aneurysmal Coil Embolization: Microcatheter Shaping Cast.

Authors:  Tomotaka Ohshima; Reo Kawaguchi; Ryuya Maejima; Naoki Matsuo; Shigeru Miyachi
Journal:  Asian J Neurosurg       Date:  2019 Jul-Sep
  6 in total