Literature DB >> 16151344

Auditory localization abilities in bilateral cochlear implant recipients.

Carl A Verschuur1, Mark E Lutman, Richard Ramsden, Paula Greenham, Martin O'Driscoll.   

Abstract

OBJECTIVE: To quantify binaural advantage for auditory localization in the horizontal plane by bilateral cochlear implant (CI) recipients. Also, to determine whether the use of dual microphones with one implant improves localization.
METHODS: Twenty subjects from the UK multicenter trial of bilateral cochlear implantation with Nucleus 24 K/M device were recruited. Sound localization was assessed in an anechoic room with an 11-loudspeaker array under four test conditions: right CI, left CI, binaural CI, and dual microphone. Two runs were undertaken for each of five stimuli (speech, tones, noise, transients, and reverberant speech). Order of conditions was counterbalanced across subjects.
RESULTS: Mean localization error with bilateral implants was 24 degrees compared with 67 degrees for monaural implant and dual microphone conditions (chance performance is 65 degrees). Normal controls average 2 to 3 degrees in similar conditions. Binaural performance was significantly better than monaural performance for all subjects, for all stimulus types, and for different sound sources. Only small differences in performance with different stimuli were observed.
CONCLUSIONS: Bilateral cochlear implantation with the Nucleus 24 device provides marked improvement in horizontal plane localization abilities compared with unilateral CI use for a range of stimuli having different spectral and temporal characteristics. Benefit was obtained by all subjects, for all stimulus types, and for all sound directions. However, binaural performance was still worse than that obtained by normal hearing listeners and hearing aid users with the same methodology. Monaural localization performance was at chance. There is no benefit for localization with dual microphones.

Mesh:

Year:  2005        PMID: 16151344     DOI: 10.1097/01.mao.0000185073.81070.07

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  30 in total

1.  Current and planned cochlear implant research at New York University Laboratory for Translational Auditory Research.

Authors:  Mario A Svirsky; Matthew B Fitzgerald; Arlene Neuman; Elad Sagi; Chin-Tuan Tan; Darlene Ketten; Brett Martin
Journal:  J Am Acad Audiol       Date:  2012-06       Impact factor: 1.664

2.  A longitudinal study in adults with sequential bilateral cochlear implants: time course for individual ear and bilateral performance.

Authors:  Ruth M Reeder; Jill B Firszt; Laura K Holden; Michael J Strube
Journal:  J Speech Lang Hear Res       Date:  2014-06-01       Impact factor: 2.297

Review 3.  The development of the Nucleus Freedom Cochlear implant system.

Authors:  James F Patrick; Peter A Busby; Peter J Gibson
Journal:  Trends Amplif       Date:  2006-12

4.  Binaural masking level differences in actual and simulated bilateral cochlear implant listeners.

Authors:  Thomas Lu; Ruth Litovsky; Fan-Gang Zeng
Journal:  J Acoust Soc Am       Date:  2010-03       Impact factor: 1.840

5.  Comparison of speech recognition and localization performance in bilateral and unilateral cochlear implant users matched on duration of deafness and age at implantation.

Authors:  Camille C Dunn; Richard S Tyler; Sarah Oakley; Bruce J Gantz; William Noble
Journal:  Ear Hear       Date:  2008-06       Impact factor: 3.570

6.  [Processing of interaural time differences in normal-hearing subjects and cochlear implant users with FSP and HDCIS coding strategy].

Authors:  N Heidekrüger; T Rahne; L Wagner
Journal:  HNO       Date:  2019-11       Impact factor: 1.284

7.  Lateralization of Interaural Level Differences with Multiple Electrode Stimulation in Bilateral Cochlear-Implant Listeners.

Authors:  Olga A Stakhovskaya; Matthew J Goupell
Journal:  Ear Hear       Date:  2017 Jan/Feb       Impact factor: 3.570

8.  Cochlear implantation in nontraditional candidates: preliminary results in adolescents with asymmetric hearing loss.

Authors:  Jamie H Cadieux; Jill B Firszt; Ruth M Reeder
Journal:  Otol Neurotol       Date:  2013-04       Impact factor: 2.311

9.  An attempt to improve bilateral cochlear implants by increasing the distance between electrodes and providing complementary information to the two ears.

Authors:  Richard S Tyler; Shelley A Witt; Camille C Dunn; Ann Perreau; Aaron J Parkinson; Blake S Wilson
Journal:  J Am Acad Audiol       Date:  2010-01       Impact factor: 1.664

10.  Performance over time on adults with simultaneous bilateral cochlear implants.

Authors:  Son-A Chang; Richard S Tyler; Camille C Dunn; Haihong Ji; Shelley A Witt; Bruce Gantz; Marlan Hansen
Journal:  J Am Acad Audiol       Date:  2010-01       Impact factor: 1.664

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.