The use of megavoltage CT (MVCT) images for dose recomputations.

K M Langen1, S L Meeks, D O Poole, T H Wagner, T R Willoughby, P A Kupelian, K J Ruchala, J Haimerl, G H Olivera.   

Abstract

Megavoltage CT (MVCT) images of patients are acquired daily on a helical tomotherapy unit (TomoTherapy, Inc., Madison, WI). While these images are used primarily for patient alignment, they can also be used to recalculate the treatment plan for the patient anatomy of the day. The use of MVCT images for dose computations requires a reliable CT number to electron density calibration curve. In this work, we tested the stability of the MVCT numbers by determining the variation of this calibration with spatial arrangement of the phantom, time and MVCT acquisition parameters. The two calibration curves that represent the largest variations were applied to six clinical MVCT images for recalculations to test for dosimetric uncertainties. Among the six cases tested, the largest difference in any of the dosimetric endpoints was 3.1% but more typically the dosimetric endpoints varied by less than 2%. Using an average CT to electron density calibration and a thorax phantom, a series of end-to-end tests were run. Using a rigid phantom, recalculated dose volume histograms (DVHs) were compared with plan DVHs. Using a deformed phantom, recalculated point dose variations were compared with measurements. The MVCT field of view is limited and the image space outside this field of view can be filled in with information from the planning kVCT. This merging technique was tested for a rigid phantom. Finally, the influence of the MVCT slice thickness on the dose recalculation was investigated. The dosimetric differences observed in all phantom tests were within the range of dosimetric uncertainties observed due to variations in the calibration curve. The use of MVCT images allows the assessment of daily dose distributions with an accuracy that is similar to that of the initial kVCT dose calculation.

Entities:  

Mesh:

Year:  2005        PMID: 16148392     DOI: 10.1088/0031-9155/50/18/002

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  43 in total

1.  Image-based modeling of tumor shrinkage in head and neck radiation therapy.

Authors:  Ming Chao; Yaoqin Xie; Eduardo G Moros; Quynh-Thu Le; Lei Xing
Journal:  Med Phys       Date:  2010-05       Impact factor: 4.071

2.  On the radiobiological impact of metal artifacts in head-and-neck IMRT in terms of tumor control probability (TCP) and normal tissue complication probability (NTCP).

Authors:  Yusung Kim; Wolfgang A Tomé
Journal:  Med Biol Eng Comput       Date:  2007-06-16       Impact factor: 2.602

3.  Monte Carlo investigations of megavoltage cone-beam CT using thick, segmented scintillating detectors for soft tissue visualization.

Authors:  Yi Wang; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Amit Sawant; Hong Du
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

4.  The use of EPID-measured leaf sequence files for IMRT dose reconstruction in adaptive radiation therapy.

Authors:  Louis Lee; Weihua Mao; Lei Xing
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

5.  Intrafractional motion of the prostate during hypofractionated radiotherapy.

Authors:  Yaoqin Xie; David Djajaputra; Christopher R King; Sabbir Hossain; Lijun Ma; Lei Xing
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-09-01       Impact factor: 7.038

6.  Deformable registration of abdominal kilovoltage treatment planning CT and tomotherapy daily megavoltage CT for treatment adaptation.

Authors:  Deshan Yang; Summer R Chaudhari; S Murty Goddu; David Pratt; Divya Khullar; Joseph O Deasy; Issam El Naqa
Journal:  Med Phys       Date:  2009-02       Impact factor: 4.071

7.  High-DQE EPIDs based on thick, segmented BGO and CsI:Tl scintillators: performance evaluation at extremely low dose.

Authors:  Yi Wang; Larry E Antonuk; Qihua Zhao; Youcef El-Mohri; Louis Perna
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

8.  Monte Carlo investigations of the effect of beam divergence on thick, segmented crystalline scintillators for radiotherapy imaging.

Authors:  Yi Wang; Youcef El-Mohri; Larry E Antonuk; Qihua Zhao
Journal:  Phys Med Biol       Date:  2010-06-04       Impact factor: 3.609

9.  Optimization of the design of thick, segmented scintillators for megavoltage cone-beam CT using a novel, hybrid modeling technique.

Authors:  Langechuan Liu; Larry E Antonuk; Youcef El-Mohri; Qihua Zhao; Hao Jiang
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

10.  On voxel-by-voxel accumulated dose for prostate radiation therapy using deformable image registration.

Authors:  Jialu Yu; Nicholas Hardcastle; Kyoungkeun Jeong; Edward T Bender; Mark A Ritter; Wolfgang A Tomé
Journal:  Technol Cancer Res Treat       Date:  2014-11-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.