Literature DB >> 16011313

Imaging performance of A-PET: a small animal PET camera.

Suleman Surti1, Joel S Karp, Amy E Perkins, Chris A Cardi, Margaret E Daube-Witherspoon, Austin Kuhn, Gerd Muehllehner.   

Abstract

The evolution of positron emission tomography (PET) imaging for small animals has led to the development of dedicated PET scanner designs with high resolution and sensitivity. The animal PET scanner achieves these goals for imaging small animals such as mice and rats. The scanner uses a pixelated Anger-logic detector for discriminating 2 x 2 x 10 mm3 crystals with 19-mm-diameter photomultiplier tubes. With a 19.7-cm ring diameter, the scanner has an axial length of 11.9 cm and operates exclusively in three-dimensional imaging mode, leading to very high sensitivity. Measurements show that the scanner design achieves a spatial resolution of 1.9 mm at the center of the field-of-view. Initially designed with gadolinium orthosilicate but changed to lutetium- yttrium orthosilicate, the scanner now achieves a sensitivity of 3.6% for a point source at the center of the field-of-view with an energy window of 250-665 keV. Iterative image reconstruction, together with accurate data corrections for scatter, random, and attenuation, are incorporated to achieve high-quality images and quantitative data. These results are demonstrated through our contrast recovery measurements as well as sample animal studies.

Entities:  

Mesh:

Year:  2005        PMID: 16011313     DOI: 10.1109/tmi.2005.844078

Source DB:  PubMed          Journal:  IEEE Trans Med Imaging        ISSN: 0278-0062            Impact factor:   10.048


  31 in total

1.  Performance evaluation of the Philips MOSAIC small animal PET scanner.

Authors:  Marc C Huisman; Sybille Reder; Axel W Weber; Sibylle I Ziegler; Markus Schwaiger
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-11-22       Impact factor: 9.236

Review 2.  Imaging stem cells implanted in infarcted myocardium.

Authors:  Rong Zhou; Paul D Acton; Victor A Ferrari
Journal:  J Am Coll Cardiol       Date:  2006-11-01       Impact factor: 24.094

Review 3.  Instrumentation for molecular imaging in preclinical research: Micro-PET and Micro-SPECT.

Authors:  Arion F Chatziioannou
Journal:  Proc Am Thorac Soc       Date:  2005

4.  Imaging studies for evaluating impact of position sampling techniques in PET scanners.

Authors:  Suleman Surti; Matthew E Werner; Joel S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2010-10-01       Impact factor: 1.679

5.  Model-Based Normalization of a Fractional-Crystal Collimator for Small-Animal PET Imaging.

Authors:  Yusheng Li; Samuel Matej; Joel S Karp; Scott D Metzler
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2017-03-15

6.  Reduction of nanoparticle avidity enhances the selectivity of vascular targeting and PET detection of pulmonary inflammation.

Authors:  Blaine J Zern; Ann-Marie Chacko; Jin Liu; Colin F Greineder; Eric R Blankemeyer; Ravi Radhakrishnan; Vladimir Muzykantov
Journal:  ACS Nano       Date:  2013-02-08       Impact factor: 15.881

7.  Development of 124I immuno-PET targeting tumor vascular TEM1/endosialin.

Authors:  Ann-Marie Chacko; Chunsheng Li; Madhura Nayak; John L Mikitsh; Jia Hu; Catherine Hou; Luigi Grasso; Nicholas C Nicolaides; Vladimir R Muzykantov; Chaitanya R Divgi; George Coukos
Journal:  J Nucl Med       Date:  2014-02-13       Impact factor: 10.057

8.  Resolution Enhancement in PET Reconstruction Using Collimation.

Authors:  Scott D Metzler; Samuel Matej; Joel S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2013-02       Impact factor: 1.679

9.  Improved spatial resolution in PET scanners using sampling techniques.

Authors:  Suleman Surti; Ryan Scheuermann; Matthew E Werner; Joel S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2009-06-01       Impact factor: 1.679

10.  5-[18F]Fluoroalkyl pyrimidine nucleosides: probes for positron emission tomography imaging of herpes simplex virus type 1 thymidine kinase gene expression.

Authors:  Ann-Marie Chacko; Eric Blankemeyer; Brian P Lieberman; Wenchao Qu; Hank F Kung
Journal:  Nucl Med Biol       Date:  2009-01       Impact factor: 2.408

View more