Abstract

BACKGROUND: Total hip arthroplasty with use of metal-on-metal bearings has been reintroduced as an alternative to the use of metal-on-polyethylene bearings because of theoretical advantages such as reduced wear and a lower prevalence of osteolysis. However, we observed early osteolysis in a cohort of patients who had been managed with second-generation metal-on-metal hip replacements and investigated the possible etiologic role of metal hypersensitivity.
METHODS: We retrospectively analyzed 165 patients (169 hips) who had undergone primary cementless total hip replacement with a contemporary metal-on-metal total hip design between 2000 and 2002. After a minimum duration of follow-up of twenty-four months, nine patients (ten hips) had an osteolytic lesion localized to the greater trochanter. Skin-patch tests for hypersensitivity to metals were performed on the nine patients and on nine randomly selected patients with total hip replacements who did not have osteolytic changes and who were matched to the study cohort for age and gender. Microbiological cultures, histopathologic examinations, and immunohistochemical analysis were performed on samples of periprosthetic tissue that were collected during revision arthroplasty on two hips with early osteolysis.
RESULTS: The patients with early osteolysis had a significantly higher rate of hypersensitivity reaction to cobalt compared with controls (p = 0.031). The retrieved periprosthetic tissues showed no evidence of metallic staining, but histologic analysis revealed a perivascular accumulation of CD3-positive T-cells and CD68-positive macrophages and an absence of both particle-laden macrophages and polymorphonuclear cells. Immunohistochemical analysis demonstrated that bone-resorbing cytokines such as IL-1beta and TNF-alpha were produced mainly by infiltrating lymphocytes and activated macrophages.
CONCLUSIONS: These findings raise the possibility that early osteolysis in patients with this second-generation metal-on-metal hip replacement is associated with abnormalities consistent with delayed-type hypersensitivity to metal. A prospective study in which a large group of patients is evaluated with multiple diagnostic methods is needed in order to establish whether there is a causal relationship between metal hypersensitivity and osteolysis.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15995119     DOI: 10.2106/JBJS.D.02641

Source DB:  PubMed          Journal:  J Bone Joint Surg Am        ISSN: 0021-9355            Impact factor:   5.284


  74 in total

1.  Low incidence of groin pain and early failure with large metal articulation total hip arthroplasty.

Authors:  John B Meding; Lindsey K Meding; E Michael Keating; Michael E Berend
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

2.  The John Charnley Award: an accurate and extremely sensitive method to separate, display, and characterize wear debris: part 2: metal and ceramic particles.

Authors:  Fabrizio Billi; Paul Benya; Aaron Kavanaugh; John Adams; Harry McKellop; Edward Ebramzadeh
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

3.  Simple isolation method for the bulk isolation of wear particles from metal on metal bearing surfaces generated in a hip simulator test.

Authors:  Fang Lu; Matt Royle; Ferdinand V Lali; Alister J Hart; Simon Collins; Jonathan Housden; Julia C Shelton
Journal:  J Mater Sci Mater Med       Date:  2012-03-06       Impact factor: 3.896

4.  Management of failed metal-on-metal total hip arthroplasty.

Authors:  Justin W Griffin; Michele D'Apuzzo; James A Browne
Journal:  World J Orthop       Date:  2012-06-18

5.  Metallosis following a dual coat porous hydroxyapatite shoulder hemiarthroplasty.

Authors:  Humza Khan; Mark Hurworth; Alan Kop
Journal:  J Orthop       Date:  2015-03-12

6.  [Allergic reactions as differential diagnosis for periprosthetic infection].

Authors:  H Meyer; A Krüger; A Roessner; C H Lohmann
Journal:  Orthopade       Date:  2012-01       Impact factor: 1.087

7.  Perivascular lymphocytic infiltration is not limited to metal-on-metal bearings.

Authors:  Vincent Y Ng; Adolph V Lombardi; Keith R Berend; Michael D Skeels; Joanne B Adams
Journal:  Clin Orthop Relat Res       Date:  2011-02       Impact factor: 4.176

8.  The influence of the centre of rotation on implant survival using a modular stem hip prosthesis.

Authors:  Francesco Traina; Marcello De Fine; Federico Biondi; Enrico Tassinari; Andrea Galvani; Aldo Toni
Journal:  Int Orthop       Date:  2008-12-20       Impact factor: 3.075

9.  Nickel induces secretion of IFN-gamma by splenic natural killer cells.

Authors:  Ji Yeon Kim; Kyungmin Huh; Ki Young Lee; Jun Mo Yang; Tae Jin Kim
Journal:  Exp Mol Med       Date:  2009-04-30       Impact factor: 8.718

10.  2008 John Charnley award: metal ion levels after metal-on-metal total hip arthroplasty: a randomized trial.

Authors:  C Anderson Engh; Steven J MacDonald; Supatra Sritulanondha; Abigail Thompson; Douglas Naudie; Charles A Engh
Journal:  Clin Orthop Relat Res       Date:  2008-10-15       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.