Literature DB >> 15990609

Comparison of dynamic contour tonometry with goldman applanation tonometry over a wide range of central corneal thickness.

Aoife Doyle1, Yves Lachkar.   

Abstract

PURPOSE: To compare dynamic contour tonometry with Goldmann applanation tonometry in structurally normal corneas over a wide range of central corneal thickness (CCT). PATIENTS AND METHODS: Twenty-five patients each with normal CCT (group A), thin corneas (group B), and thick corneas (group C) had IOP measured with the Goldmann (GAT) and dynamic contour tonometer (DCT).
RESULTS: In group A (mean CCT = 552 +/- 16 microm) the mean GAT was 15.9 +/- 3.1 mm Hg and mean DCT was 16 +/- 3.3 mm Hg (P = 0.91). In group B (mean CCT = 491 +/- 19 microm) the mean GAT was 13.2 +/- 3.5 mm Hg and the mean DCT was 15.9 +/- 3.5 mm Hg (P = 0.009). For group C (mean CCT = 615 +/- 22 microm), the mean GAT was 17.4 +/- 3.8 mm Hg and the mean DCT was 17.4 +/- 3.5 mm Hg (P = 0.95). The 95% agreement limits for DCT were -3.1 mm Hg to 2.9 mm Hg. The mean GAT-DCT difference was -2.6 mm Hg in thin corneas and -0.06 mm Hg in thick corneas. Below 520 microm reduction of 10 microm in CCT appears to result in a significant underestimation of the GAT IOP by 0.7 mm Hg (P < .001) and above 580 microm a non-significant overestimation of 0.2 mm Hg per 10 microm increase in CCT (P = 0.27).
CONCLUSION: Dynamic contour tonometer agrees well on average with GAT but the agreement limits are wide. In structurally normal thin corneas DCT may give a more accurate assessment of the true IOP but it does not appear to have any benefit over GAT in thick corneas.

Entities:  

Mesh:

Year:  2005        PMID: 15990609     DOI: 10.1097/01.ijg.0000169393.40298.05

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  26 in total

Review 1.  [Dynamic contour tonometry].

Authors:  C Kniestedt; H E Kanngiesser
Journal:  Ophthalmologe       Date:  2006-08       Impact factor: 1.059

2.  Intraocular pressure measured by dynamic contour tonometer and ocular response analyzer in normal tension glaucoma.

Authors:  Tetsuya Morita; Nobuyuki Shoji; Kazutaka Kamiya; Mana Hagishima; Fusako Fujimura; Kimiya Shimizu
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-08-20       Impact factor: 3.117

3.  Intraocular pressure after Descemet's stripping and non-Descemet's stripping automated endothelial keratoplasty.

Authors:  Yoshiro Mawatari; Akira Kobayashi; Hideaki Yokogawa; Kazuhisa Sugiyama
Journal:  Jpn J Ophthalmol       Date:  2011-03-13       Impact factor: 2.447

4.  Impact of corneal parameters on intraocular pressure measurements in different tonometry methods.

Authors:  Aleksandra Zakrzewska; Marta P Wiącek; Anna Machalińska
Journal:  Int J Ophthalmol       Date:  2019-12-18       Impact factor: 1.779

5.  [Evaluation of dynamic contour tonometry in penetrating keratoplasties].

Authors:  An Viestenz; A Langenbucher; B Seitz; Ar Viestenz
Journal:  Ophthalmologe       Date:  2006-09       Impact factor: 1.059

6.  Physiological diurnal variability and characteristics of the ocular pulse amplitude (OPA) with the dynamic contour tonometer (DCT-Pascal).

Authors:  Sayeh Pourjavan; Pierre-Yves Boëlle; Michéle Detry-Morel; Patrick De Potter
Journal:  Int Ophthalmol       Date:  2007-10-23       Impact factor: 2.031

7.  [Evaluation of dynamic contour tonometry in keratoconus].

Authors:  N Schädle; J D Unterlauft; T Klink; G Geerling
Journal:  Ophthalmologe       Date:  2009-06       Impact factor: 1.059

8.  Comparison of Goldmann applanation and dynamic contour tonometry in a population of Mexican open-angle glaucoma patients.

Authors:  J Jimenez-Roman; F Gil-Carrasco; A Martinez; O Albis-Donado; J D De la Torre-Tovar
Journal:  Int Ophthalmol       Date:  2012-11-16       Impact factor: 2.031

9.  Comparison between dynamic contour tonometry and Goldmann applanation tonometry.

Authors:  Jooeun Lee; Chang Hwan Lee; Jaewan Choi; Sam Young Yoon; Kyung Rim Sung; Seong Bae Park; Michael S Kook
Journal:  Korean J Ophthalmol       Date:  2009-03-09

10.  Intraocular pressure profile during the modified diurnal tension curve using Goldman applanation tonometry and dynamic contour tonometry.

Authors:  Carlos Gustavo Vasconcelos de Moraes; Alexandre Soares Castro Reis; Milena Eimi Sano; Alan Kardec Barreira; Roberto Murad Vessani; Remo Jr Susanna
Journal:  J Ocul Biol Dis Infor       Date:  2009-02-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.